Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams
PURPOSE: Radiochromic films are versatile 2D dosimeters with high‐resolution and near tissue equivalence. To assure high precision and accuracy, a time‐consuming calibration process is required. To improve the time efficiency, a novel calibration method utilizing the ratio of the same dose profile m...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9543697/ https://www.ncbi.nlm.nih.gov/pubmed/35754376 http://dx.doi.org/10.1002/mp.15828 |
_version_ | 1784804435392724992 |
---|---|
author | Resch, Andreas F. Padilla Cabal, Fatima Regodic, Milovan Lechner, Wolfgang Heilemann, Gerd Kuess, Peter Georg, Dietmar Palmans, Hugo |
author_facet | Resch, Andreas F. Padilla Cabal, Fatima Regodic, Milovan Lechner, Wolfgang Heilemann, Gerd Kuess, Peter Georg, Dietmar Palmans, Hugo |
author_sort | Resch, Andreas F. |
collection | PubMed |
description | PURPOSE: Radiochromic films are versatile 2D dosimeters with high‐resolution and near tissue equivalence. To assure high precision and accuracy, a time‐consuming calibration process is required. To improve the time efficiency, a novel calibration method utilizing the ratio of the same dose profile measured at different monitor units (MUs) is introduced and tested in a proton and photon beam. METHODS: The calibration procedure employs the dose ratio of film measurements of the same relative profile for different absolute dose values. Hence, the ratio of the dose is constant at any point of the profile, but the ratio of the net optical densities is not constant. The key idea of the method is to optimize the calibration function until the ratio of the calculated doses is constant. The proposed method was tested in the dose range between 0.25–12 and 1–6 Gy in a proton and photon beam, respectively. A radial symmetric profile and a rectangular profile were created, both having a central plateau region of about 3 cm diameter and a dose falloff of about 1.5 cm at larger distances. The dose falloff region was used as input for the optimization method and the central plateau region served as dose reference points. Only the plateau region of the highest dose entered the optimization as an additional objective. The measured data were randomly split into differently sized training and test sets. The optimization was repeated 1000 times with random start value initialization using the same start values for the standard and the gradient method. Finally, a proton plan with four dose levels was created, which were separated spatially, to test the possibility of a full calibration within a single measurement. RESULTS: Parameter estimation was possible with as low as one dose ratio used for optimization in both the photon and the proton case, yet exhibiting a high sensitivity on the dose level. The root mean squared deviation (RMSD) of the dose was less than 1% when the dose ratio was in the order of 20, whereas the median RMSD of all optimizations was 1.7%. Using four dose levels for optimization resulted in a median RMSD of 1% when randomly selecting the dose levels. Having at least one dose ratio of about 20 included in the optimization considerably improved the RMSD of the calibration function. Using six or eight dose levels reduced the sensitivity on the dose level selection and the median RMSD was 0.8%. A full calibration was possible in a single measurement having four dose levels in one plan but spatially separated. CONCLUSIONS: The number of measurements required to obtain an EBT3 film calibration function could be reduced using the proposed dose ratio method while maintaining the same accuracy as with the standard method. |
format | Online Article Text |
id | pubmed-9543697 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-95436972022-10-14 Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams Resch, Andreas F. Padilla Cabal, Fatima Regodic, Milovan Lechner, Wolfgang Heilemann, Gerd Kuess, Peter Georg, Dietmar Palmans, Hugo Med Phys COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY PURPOSE: Radiochromic films are versatile 2D dosimeters with high‐resolution and near tissue equivalence. To assure high precision and accuracy, a time‐consuming calibration process is required. To improve the time efficiency, a novel calibration method utilizing the ratio of the same dose profile measured at different monitor units (MUs) is introduced and tested in a proton and photon beam. METHODS: The calibration procedure employs the dose ratio of film measurements of the same relative profile for different absolute dose values. Hence, the ratio of the dose is constant at any point of the profile, but the ratio of the net optical densities is not constant. The key idea of the method is to optimize the calibration function until the ratio of the calculated doses is constant. The proposed method was tested in the dose range between 0.25–12 and 1–6 Gy in a proton and photon beam, respectively. A radial symmetric profile and a rectangular profile were created, both having a central plateau region of about 3 cm diameter and a dose falloff of about 1.5 cm at larger distances. The dose falloff region was used as input for the optimization method and the central plateau region served as dose reference points. Only the plateau region of the highest dose entered the optimization as an additional objective. The measured data were randomly split into differently sized training and test sets. The optimization was repeated 1000 times with random start value initialization using the same start values for the standard and the gradient method. Finally, a proton plan with four dose levels was created, which were separated spatially, to test the possibility of a full calibration within a single measurement. RESULTS: Parameter estimation was possible with as low as one dose ratio used for optimization in both the photon and the proton case, yet exhibiting a high sensitivity on the dose level. The root mean squared deviation (RMSD) of the dose was less than 1% when the dose ratio was in the order of 20, whereas the median RMSD of all optimizations was 1.7%. Using four dose levels for optimization resulted in a median RMSD of 1% when randomly selecting the dose levels. Having at least one dose ratio of about 20 included in the optimization considerably improved the RMSD of the calibration function. Using six or eight dose levels reduced the sensitivity on the dose level selection and the median RMSD was 0.8%. A full calibration was possible in a single measurement having four dose levels in one plan but spatially separated. CONCLUSIONS: The number of measurements required to obtain an EBT3 film calibration function could be reduced using the proposed dose ratio method while maintaining the same accuracy as with the standard method. John Wiley and Sons Inc. 2022-07-14 2022-09 /pmc/articles/PMC9543697/ /pubmed/35754376 http://dx.doi.org/10.1002/mp.15828 Text en © 2022 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY Resch, Andreas F. Padilla Cabal, Fatima Regodic, Milovan Lechner, Wolfgang Heilemann, Gerd Kuess, Peter Georg, Dietmar Palmans, Hugo Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams |
title | Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams |
title_full | Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams |
title_fullStr | Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams |
title_full_unstemmed | Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams |
title_short | Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams |
title_sort | accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams |
topic | COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9543697/ https://www.ncbi.nlm.nih.gov/pubmed/35754376 http://dx.doi.org/10.1002/mp.15828 |
work_keys_str_mv | AT reschandreasf acceleratingandimprovingradiochromicfilmcalibrationbyutilizingthedoseratioinphotonandprotonbeams AT padillacabalfatima acceleratingandimprovingradiochromicfilmcalibrationbyutilizingthedoseratioinphotonandprotonbeams AT regodicmilovan acceleratingandimprovingradiochromicfilmcalibrationbyutilizingthedoseratioinphotonandprotonbeams AT lechnerwolfgang acceleratingandimprovingradiochromicfilmcalibrationbyutilizingthedoseratioinphotonandprotonbeams AT heilemanngerd acceleratingandimprovingradiochromicfilmcalibrationbyutilizingthedoseratioinphotonandprotonbeams AT kuesspeter acceleratingandimprovingradiochromicfilmcalibrationbyutilizingthedoseratioinphotonandprotonbeams AT georgdietmar acceleratingandimprovingradiochromicfilmcalibrationbyutilizingthedoseratioinphotonandprotonbeams AT palmanshugo acceleratingandimprovingradiochromicfilmcalibrationbyutilizingthedoseratioinphotonandprotonbeams |