In vivo porcine characterization of atrial lesion safety and efficacy utilizing a circular pulsed‐field ablation catheter including assessment of collateral damage to adjacent tissue in supratherapeutic ablation applications

INTRODUCTION: Pulsed‐field ablation (PFA), an ablative method that causes cell death by irreversible electroporation, has potential safety advantages over radiofrequency ablation and cryoablation. Pulmonary vein (PV) isolation was performed in a porcine model to characterize safety and performance o...

Descripción completa

Detalles Bibliográficos
Autores principales: Hsu, Jonathan C., Gibson, Douglas, Banker, Rajesh, Doshi, Shephal K., Gidney, Brett, Gomez, Tara, Berman, Dror, Datta, Keshava, Govari, Assaf, Natale, Andrea
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9545022/
https://www.ncbi.nlm.nih.gov/pubmed/35510408
http://dx.doi.org/10.1111/jce.15522
_version_ 1784804727690625024
author Hsu, Jonathan C.
Gibson, Douglas
Banker, Rajesh
Doshi, Shephal K.
Gidney, Brett
Gomez, Tara
Berman, Dror
Datta, Keshava
Govari, Assaf
Natale, Andrea
author_facet Hsu, Jonathan C.
Gibson, Douglas
Banker, Rajesh
Doshi, Shephal K.
Gidney, Brett
Gomez, Tara
Berman, Dror
Datta, Keshava
Govari, Assaf
Natale, Andrea
author_sort Hsu, Jonathan C.
collection PubMed
description INTRODUCTION: Pulsed‐field ablation (PFA), an ablative method that causes cell death by irreversible electroporation, has potential safety advantages over radiofrequency ablation and cryoablation. Pulmonary vein (PV) isolation was performed in a porcine model to characterize safety and performance of a novel, fully‐integrated biphasic PFA system comprising a multi‐channel generator, variable loop circular catheter, and integrated PFA mapping software module. METHODS: Eight healthy porcine subjects were included. To evaluate safety, multiple ablations were performed, including sites not generally targeted for therapeutic ablation, such as the right inferior PV lumen, right superior PV ostium, and adjacent to the esophagus and phrenic nerve. To evaluate the efficacy, animals were recovered, followed for 30(±3) days, then re‐mapped. Gross pathological and histopathological examinations assessed procedural injuries, chronic thrombosis, tissue ablation, penetration depth, healing, and inflammatory response. RESULTS: All eight animals survived follow‐up. PV narrowing was not observed acutely nor at follow‐up, even when ablation was performed deep to the PV ostium. No injury was seen grossly or histologically in adjacent structures. All PVs were durably isolated, confirmed by bidirectional block at re‐map procedure. Histological examination showed complete, transmural necrosis around the circumference of the ablated section of right PVs. CONCLUSION: This preclinical evaluation of a fully‐integrated PFA system demonstrated effective and durable ablation of cardiac tissue and PV isolation without collateral damage to adjacent structures, even when ablation was performed in more extreme settings than those used therapeutically. Histological staining confirmed complete transmural cell necrosis around the circumference of the PV ostium at 30 days.
format Online
Article
Text
id pubmed-9545022
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-95450222022-10-14 In vivo porcine characterization of atrial lesion safety and efficacy utilizing a circular pulsed‐field ablation catheter including assessment of collateral damage to adjacent tissue in supratherapeutic ablation applications Hsu, Jonathan C. Gibson, Douglas Banker, Rajesh Doshi, Shephal K. Gidney, Brett Gomez, Tara Berman, Dror Datta, Keshava Govari, Assaf Natale, Andrea J Cardiovasc Electrophysiol Original Articles INTRODUCTION: Pulsed‐field ablation (PFA), an ablative method that causes cell death by irreversible electroporation, has potential safety advantages over radiofrequency ablation and cryoablation. Pulmonary vein (PV) isolation was performed in a porcine model to characterize safety and performance of a novel, fully‐integrated biphasic PFA system comprising a multi‐channel generator, variable loop circular catheter, and integrated PFA mapping software module. METHODS: Eight healthy porcine subjects were included. To evaluate safety, multiple ablations were performed, including sites not generally targeted for therapeutic ablation, such as the right inferior PV lumen, right superior PV ostium, and adjacent to the esophagus and phrenic nerve. To evaluate the efficacy, animals were recovered, followed for 30(±3) days, then re‐mapped. Gross pathological and histopathological examinations assessed procedural injuries, chronic thrombosis, tissue ablation, penetration depth, healing, and inflammatory response. RESULTS: All eight animals survived follow‐up. PV narrowing was not observed acutely nor at follow‐up, even when ablation was performed deep to the PV ostium. No injury was seen grossly or histologically in adjacent structures. All PVs were durably isolated, confirmed by bidirectional block at re‐map procedure. Histological examination showed complete, transmural necrosis around the circumference of the ablated section of right PVs. CONCLUSION: This preclinical evaluation of a fully‐integrated PFA system demonstrated effective and durable ablation of cardiac tissue and PV isolation without collateral damage to adjacent structures, even when ablation was performed in more extreme settings than those used therapeutically. Histological staining confirmed complete transmural cell necrosis around the circumference of the PV ostium at 30 days. John Wiley and Sons Inc. 2022-05-16 2022-07 /pmc/articles/PMC9545022/ /pubmed/35510408 http://dx.doi.org/10.1111/jce.15522 Text en © 2022 The Authors. Journal of Cardiovascular Electrophysiology published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Articles
Hsu, Jonathan C.
Gibson, Douglas
Banker, Rajesh
Doshi, Shephal K.
Gidney, Brett
Gomez, Tara
Berman, Dror
Datta, Keshava
Govari, Assaf
Natale, Andrea
In vivo porcine characterization of atrial lesion safety and efficacy utilizing a circular pulsed‐field ablation catheter including assessment of collateral damage to adjacent tissue in supratherapeutic ablation applications
title In vivo porcine characterization of atrial lesion safety and efficacy utilizing a circular pulsed‐field ablation catheter including assessment of collateral damage to adjacent tissue in supratherapeutic ablation applications
title_full In vivo porcine characterization of atrial lesion safety and efficacy utilizing a circular pulsed‐field ablation catheter including assessment of collateral damage to adjacent tissue in supratherapeutic ablation applications
title_fullStr In vivo porcine characterization of atrial lesion safety and efficacy utilizing a circular pulsed‐field ablation catheter including assessment of collateral damage to adjacent tissue in supratherapeutic ablation applications
title_full_unstemmed In vivo porcine characterization of atrial lesion safety and efficacy utilizing a circular pulsed‐field ablation catheter including assessment of collateral damage to adjacent tissue in supratherapeutic ablation applications
title_short In vivo porcine characterization of atrial lesion safety and efficacy utilizing a circular pulsed‐field ablation catheter including assessment of collateral damage to adjacent tissue in supratherapeutic ablation applications
title_sort in vivo porcine characterization of atrial lesion safety and efficacy utilizing a circular pulsed‐field ablation catheter including assessment of collateral damage to adjacent tissue in supratherapeutic ablation applications
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9545022/
https://www.ncbi.nlm.nih.gov/pubmed/35510408
http://dx.doi.org/10.1111/jce.15522
work_keys_str_mv AT hsujonathanc invivoporcinecharacterizationofatriallesionsafetyandefficacyutilizingacircularpulsedfieldablationcatheterincludingassessmentofcollateraldamagetoadjacenttissueinsupratherapeuticablationapplications
AT gibsondouglas invivoporcinecharacterizationofatriallesionsafetyandefficacyutilizingacircularpulsedfieldablationcatheterincludingassessmentofcollateraldamagetoadjacenttissueinsupratherapeuticablationapplications
AT bankerrajesh invivoporcinecharacterizationofatriallesionsafetyandefficacyutilizingacircularpulsedfieldablationcatheterincludingassessmentofcollateraldamagetoadjacenttissueinsupratherapeuticablationapplications
AT doshishephalk invivoporcinecharacterizationofatriallesionsafetyandefficacyutilizingacircularpulsedfieldablationcatheterincludingassessmentofcollateraldamagetoadjacenttissueinsupratherapeuticablationapplications
AT gidneybrett invivoporcinecharacterizationofatriallesionsafetyandefficacyutilizingacircularpulsedfieldablationcatheterincludingassessmentofcollateraldamagetoadjacenttissueinsupratherapeuticablationapplications
AT gomeztara invivoporcinecharacterizationofatriallesionsafetyandefficacyutilizingacircularpulsedfieldablationcatheterincludingassessmentofcollateraldamagetoadjacenttissueinsupratherapeuticablationapplications
AT bermandror invivoporcinecharacterizationofatriallesionsafetyandefficacyutilizingacircularpulsedfieldablationcatheterincludingassessmentofcollateraldamagetoadjacenttissueinsupratherapeuticablationapplications
AT dattakeshava invivoporcinecharacterizationofatriallesionsafetyandefficacyutilizingacircularpulsedfieldablationcatheterincludingassessmentofcollateraldamagetoadjacenttissueinsupratherapeuticablationapplications
AT govariassaf invivoporcinecharacterizationofatriallesionsafetyandefficacyutilizingacircularpulsedfieldablationcatheterincludingassessmentofcollateraldamagetoadjacenttissueinsupratherapeuticablationapplications
AT nataleandrea invivoporcinecharacterizationofatriallesionsafetyandefficacyutilizingacircularpulsedfieldablationcatheterincludingassessmentofcollateraldamagetoadjacenttissueinsupratherapeuticablationapplications