Cargando…

The Effect of Tedlar Bags on the Composition of Exhaled Human Breath Samples

The PVF Tedlar is widely used for gas collection in clinical diagnostics and environmental research. However, sample collection is frequently associated with the degradation, adsorption, or transformation of sensitive chemicals. Here, we explore to what extent the Tedlar bag collection effects the c...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhao, Jiuyan, Zhu, Lanlan, Zhang, Wei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9546690/
https://www.ncbi.nlm.nih.gov/pubmed/36212946
http://dx.doi.org/10.1155/2022/5665921
Descripción
Sumario:The PVF Tedlar is widely used for gas collection in clinical diagnostics and environmental research. However, sample collection is frequently associated with the degradation, adsorption, or transformation of sensitive chemicals. Here, we explore to what extent the Tedlar bag collection effects the composition of expired breath samples. Collected breath samples were analyzed using the EESI-MS technique after the storage time of 30 min, 1 h, 2 h, 3 h, 4 h, 5 h, and 6 h, respectively. Our results demonstrated the gradual MS signal decay after 3 h storage. The decay rate of 3 h is about 45% and 6 h is about 88%. Therefore, the Tedlar bag is suggested as a reliable breath holder on the time scale of <3 h.