Clinical outcomes of cryoballoon ablation for pulmonary vein isolation: Impact of intraprocedural heart rhythm

BACKGROUND: The current study sought to assess the impact of the intraprocedural heart rhythm (sinus rhythm [SR] vs. atrial fibrillation [AF]) on acute procedural characteristics, durability of pulmonary vein isolation (PVI) and long-term clinical outcomes of cryoballoon (CB) ablation. METHODS: A to...

Descripción completa

Detalles Bibliográficos
Autores principales: Reissmann, Bruno, Heeger, Christian-H., Opitz, Karena, Schlüter, Michael, Wohlmuth, Peter, Rottner, Laura, Fink, Thomas, Gerds-Li, Jin-Hong, Mathew, Shibu, Lemes, Christine, Maurer, Tilman, Ouyang, Feifan, Kuck, Karl-Heinz, Rillig, Andreas, Schöppenthau, Doreen, Metzner, Andreas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Via Medica 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9550325/
https://www.ncbi.nlm.nih.gov/pubmed/33140384
http://dx.doi.org/10.5603/CJ.a2020.0147
Descripción
Sumario:BACKGROUND: The current study sought to assess the impact of the intraprocedural heart rhythm (sinus rhythm [SR] vs. atrial fibrillation [AF]) on acute procedural characteristics, durability of pulmonary vein isolation (PVI) and long-term clinical outcomes of cryoballoon (CB) ablation. METHODS: A total of 195 patients with symptomatic paroxysmal (n = 136) or persistent AF (n = 59) underwent CB-based PVI. Ablation procedures were either performed in SR (SR group; n = 147) or during AF (AF group; n = 48). Persistent AF was more frequent in the AF group than in the SR group (62% vs. 20%). All other patient baseline characteristics did not differ between the two groups. RESULTS: The nadir temperature during the CB applications was significantly lower in the AF group than in patients in the SR group (−49 [interquartile range, −44; −54]°C vs. −47 [−42; −52]°C, p = 0.002). Median procedure and fluoroscopy times as well as the rate of real-time recordings were not different between the two groups. Repeat ablation for the treatment of atrial arrhythmia recurrence was performed in 60 patients (SR: 44 [30%] patients; AF: 16 [33%] patients), with a trend towards a lower rate of pulmonary vein reconnections in the AF group (p = 0.07). There was no difference in 3-year arrhythmia-free survival (p = 0.8). CONCLUSIONS: Cryoballoon-based PVI during AF results in lower nadir balloon temperatures and a trend towards a higher durability of PVI as compared to procedures performed in SR. The rate of real-time PVI recordings was not affected by the intraprocedural heart rhythm.