Evaluating N95 Respirator Filtration, Seal, Qualitative and Quantitative Fit with Vaporous Hydrogen Peroxide Reprocessing

Background: The COVID-19 pandemic has created personal protective equipment (PPE) shortages, particularly of N95 respirators. Institutions have used decontamination strategies including vaporous hydrogen peroxide (VHP) to augment respirator supplies. VHP can be used to decontaminate nonporous surfac...

Descripción completa

Detalles Bibliográficos
Autores principales: Yen, Christina, Mehrotra, Preeti, Pepe, Dana, Wright, Sharon, Gordon, Patrick, Parameswaran, Lalitha
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cambridge University Press 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9551554/
http://dx.doi.org/10.1017/ash.2021.38
_version_ 1784806128999202816
author Yen, Christina
Mehrotra, Preeti
Pepe, Dana
Wright, Sharon
Gordon, Patrick
Parameswaran, Lalitha
author_facet Yen, Christina
Mehrotra, Preeti
Pepe, Dana
Wright, Sharon
Gordon, Patrick
Parameswaran, Lalitha
author_sort Yen, Christina
collection PubMed
description Background: The COVID-19 pandemic has created personal protective equipment (PPE) shortages, particularly of N95 respirators. Institutions have used decontamination strategies including vaporous hydrogen peroxide (VHP) to augment respirator supplies. VHP can be used to decontaminate nonporous surfaces without compromising material integrity. However, little is known about its impact on N95 respirator efficacy. We assessed whether repeated VHP reprocessing altered 4 key respirator efficacy qualities: quantitative fit, qualitative fit, seal check, and filtration rate. Methods: We conducted a prospective cohort study from June 15 to August 31, 2020. In total, 7 participants were fitted to a 3M 1860 small or regular N95 respirator based on qualitative and quantitative fit testing. Respirators underwent 25 disinfection cycles with the Bioquell BQ-50 VHP generator. After each cycle, participants donned and doffed respirators and performed a seal check. Participants were given 2 attempts to pass their seal check. Every 10 cycles, qualitative fit testing was done using an aerosolized Bitrex solution. Quantitative fit testing was conducted using a PortaCount Pro 8038 Fit Tester to generate a fit factor score. Appropriate fit is defined as a fit factor score of 100 or greater. Quantitative testing was done at cycles 1, 3, 5, 7, 10, 15, 20, and 25. Filtration efficiencies of particles ≥0.3 µm in diameter were measured using the TSI Optical Particle Sizer 3330 at cycles 1, 5, 10, 15, 20, and 25. The Fisher exact test was used to assess qualitative fit and seal check. The Kruskal-Wallis test was used to analyze quantitative fit and filtration rate. Results: We observed no seal-check or quantitative-fit test failures during the study window. All participants passed qualitative fit testing. Although there was a significant degree of variability in fit factor scores across disinfection cycles (mean score 163.5, p <0.05), there was no significant difference between participants (p = 0.6) (Figure 1). There was no statistically significant change in mean filtration rate from cycle 10 to 25 (P = .05), and the filtration rate remained >95% by cycle 25 (Figure 2). Conclusions: VHP reprocessing did not diminish the efficacy of N95 respirators based on the 4 metrics we assessed: filtration rate, seal check, qualitative fit, and quantitative fit. Of significance, the filtration rate remained well above the 95% standard filtration for N95 respirators—even through 25 cycles of reprocessing. VHP reprocessing is a safe, viable strategy to disinfect N95 respirators and extend their use, particularly during supply shortages. Funding: No Disclosures: None
format Online
Article
Text
id pubmed-9551554
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Cambridge University Press
record_format MEDLINE/PubMed
spelling pubmed-95515542022-10-12 Evaluating N95 Respirator Filtration, Seal, Qualitative and Quantitative Fit with Vaporous Hydrogen Peroxide Reprocessing Yen, Christina Mehrotra, Preeti Pepe, Dana Wright, Sharon Gordon, Patrick Parameswaran, Lalitha Antimicrob Steward Healthc Epidemiol Disinfection/Sterilization Background: The COVID-19 pandemic has created personal protective equipment (PPE) shortages, particularly of N95 respirators. Institutions have used decontamination strategies including vaporous hydrogen peroxide (VHP) to augment respirator supplies. VHP can be used to decontaminate nonporous surfaces without compromising material integrity. However, little is known about its impact on N95 respirator efficacy. We assessed whether repeated VHP reprocessing altered 4 key respirator efficacy qualities: quantitative fit, qualitative fit, seal check, and filtration rate. Methods: We conducted a prospective cohort study from June 15 to August 31, 2020. In total, 7 participants were fitted to a 3M 1860 small or regular N95 respirator based on qualitative and quantitative fit testing. Respirators underwent 25 disinfection cycles with the Bioquell BQ-50 VHP generator. After each cycle, participants donned and doffed respirators and performed a seal check. Participants were given 2 attempts to pass their seal check. Every 10 cycles, qualitative fit testing was done using an aerosolized Bitrex solution. Quantitative fit testing was conducted using a PortaCount Pro 8038 Fit Tester to generate a fit factor score. Appropriate fit is defined as a fit factor score of 100 or greater. Quantitative testing was done at cycles 1, 3, 5, 7, 10, 15, 20, and 25. Filtration efficiencies of particles ≥0.3 µm in diameter were measured using the TSI Optical Particle Sizer 3330 at cycles 1, 5, 10, 15, 20, and 25. The Fisher exact test was used to assess qualitative fit and seal check. The Kruskal-Wallis test was used to analyze quantitative fit and filtration rate. Results: We observed no seal-check or quantitative-fit test failures during the study window. All participants passed qualitative fit testing. Although there was a significant degree of variability in fit factor scores across disinfection cycles (mean score 163.5, p <0.05), there was no significant difference between participants (p = 0.6) (Figure 1). There was no statistically significant change in mean filtration rate from cycle 10 to 25 (P = .05), and the filtration rate remained >95% by cycle 25 (Figure 2). Conclusions: VHP reprocessing did not diminish the efficacy of N95 respirators based on the 4 metrics we assessed: filtration rate, seal check, qualitative fit, and quantitative fit. Of significance, the filtration rate remained well above the 95% standard filtration for N95 respirators—even through 25 cycles of reprocessing. VHP reprocessing is a safe, viable strategy to disinfect N95 respirators and extend their use, particularly during supply shortages. Funding: No Disclosures: None Cambridge University Press 2021-07-29 /pmc/articles/PMC9551554/ http://dx.doi.org/10.1017/ash.2021.38 Text en © The Society for Healthcare Epidemiology of America 2021 https://creativecommons.org/licenses/by/4.0/This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Disinfection/Sterilization
Yen, Christina
Mehrotra, Preeti
Pepe, Dana
Wright, Sharon
Gordon, Patrick
Parameswaran, Lalitha
Evaluating N95 Respirator Filtration, Seal, Qualitative and Quantitative Fit with Vaporous Hydrogen Peroxide Reprocessing
title Evaluating N95 Respirator Filtration, Seal, Qualitative and Quantitative Fit with Vaporous Hydrogen Peroxide Reprocessing
title_full Evaluating N95 Respirator Filtration, Seal, Qualitative and Quantitative Fit with Vaporous Hydrogen Peroxide Reprocessing
title_fullStr Evaluating N95 Respirator Filtration, Seal, Qualitative and Quantitative Fit with Vaporous Hydrogen Peroxide Reprocessing
title_full_unstemmed Evaluating N95 Respirator Filtration, Seal, Qualitative and Quantitative Fit with Vaporous Hydrogen Peroxide Reprocessing
title_short Evaluating N95 Respirator Filtration, Seal, Qualitative and Quantitative Fit with Vaporous Hydrogen Peroxide Reprocessing
title_sort evaluating n95 respirator filtration, seal, qualitative and quantitative fit with vaporous hydrogen peroxide reprocessing
topic Disinfection/Sterilization
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9551554/
http://dx.doi.org/10.1017/ash.2021.38
work_keys_str_mv AT yenchristina evaluatingn95respiratorfiltrationsealqualitativeandquantitativefitwithvaporoushydrogenperoxidereprocessing
AT mehrotrapreeti evaluatingn95respiratorfiltrationsealqualitativeandquantitativefitwithvaporoushydrogenperoxidereprocessing
AT pepedana evaluatingn95respiratorfiltrationsealqualitativeandquantitativefitwithvaporoushydrogenperoxidereprocessing
AT wrightsharon evaluatingn95respiratorfiltrationsealqualitativeandquantitativefitwithvaporoushydrogenperoxidereprocessing
AT gordonpatrick evaluatingn95respiratorfiltrationsealqualitativeandquantitativefitwithvaporoushydrogenperoxidereprocessing
AT parameswaranlalitha evaluatingn95respiratorfiltrationsealqualitativeandquantitativefitwithvaporoushydrogenperoxidereprocessing