Cargando…
Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement
Intraoperative defibrillation secondary to the usage of electrocautery in a patient with a cardiovascular implantable electronic device is a rare occurrence, and below-the-umbilicus electrocautery use causing inadvertent defibrillation is a near-zero risk. Defibrillation secondary to electrodispersi...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Cureus
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9585508/ https://www.ncbi.nlm.nih.gov/pubmed/36304382 http://dx.doi.org/10.7759/cureus.29391 |
_version_ | 1784813508708270080 |
---|---|
author | McGuire, Joseph A Hayanga, Jeremiah Barry, Charles Carpenter, Anna Frye, Benjamin Hughes, James Schwartzman, David Hayanga, Heather |
author_facet | McGuire, Joseph A Hayanga, Jeremiah Barry, Charles Carpenter, Anna Frye, Benjamin Hughes, James Schwartzman, David Hayanga, Heather |
author_sort | McGuire, Joseph A |
collection | PubMed |
description | Intraoperative defibrillation secondary to the usage of electrocautery in a patient with a cardiovascular implantable electronic device is a rare occurrence, and below-the-umbilicus electrocautery use causing inadvertent defibrillation is a near-zero risk. Defibrillation secondary to electrodispersive pad (EDP) radiofrequency dispersion has only ever been theorized. In this report, we describe the case of a 67-year-old male with an automatic implantable cardioverter defibrillator (AICD) undergoing a robotic-assisted left anterior total hip arthroplasty for left hip osteoarthritis who experienced inadvertent intraoperative defibrillation concurrent with electrocautery usage. The defibrillations ceased following contralateral EDP repositioning and application of a donut magnet overlying the patient’s AICD. |
format | Online Article Text |
id | pubmed-9585508 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Cureus |
record_format | MEDLINE/PubMed |
spelling | pubmed-95855082022-10-26 Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement McGuire, Joseph A Hayanga, Jeremiah Barry, Charles Carpenter, Anna Frye, Benjamin Hughes, James Schwartzman, David Hayanga, Heather Cureus Anesthesiology Intraoperative defibrillation secondary to the usage of electrocautery in a patient with a cardiovascular implantable electronic device is a rare occurrence, and below-the-umbilicus electrocautery use causing inadvertent defibrillation is a near-zero risk. Defibrillation secondary to electrodispersive pad (EDP) radiofrequency dispersion has only ever been theorized. In this report, we describe the case of a 67-year-old male with an automatic implantable cardioverter defibrillator (AICD) undergoing a robotic-assisted left anterior total hip arthroplasty for left hip osteoarthritis who experienced inadvertent intraoperative defibrillation concurrent with electrocautery usage. The defibrillations ceased following contralateral EDP repositioning and application of a donut magnet overlying the patient’s AICD. Cureus 2022-09-21 /pmc/articles/PMC9585508/ /pubmed/36304382 http://dx.doi.org/10.7759/cureus.29391 Text en Copyright © 2022, McGuire et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Anesthesiology McGuire, Joseph A Hayanga, Jeremiah Barry, Charles Carpenter, Anna Frye, Benjamin Hughes, James Schwartzman, David Hayanga, Heather Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement |
title | Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement |
title_full | Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement |
title_fullStr | Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement |
title_full_unstemmed | Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement |
title_short | Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement |
title_sort | inadvertent intraoperative defibrillation secondary to electrocautery grounding pad placement |
topic | Anesthesiology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9585508/ https://www.ncbi.nlm.nih.gov/pubmed/36304382 http://dx.doi.org/10.7759/cureus.29391 |
work_keys_str_mv | AT mcguirejosepha inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement AT hayangajeremiah inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement AT barrycharles inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement AT carpenteranna inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement AT fryebenjamin inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement AT hughesjames inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement AT schwartzmandavid inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement AT hayangaheather inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement |