Cargando…

Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement

Intraoperative defibrillation secondary to the usage of electrocautery in a patient with a cardiovascular implantable electronic device is a rare occurrence, and below-the-umbilicus electrocautery use causing inadvertent defibrillation is a near-zero risk. Defibrillation secondary to electrodispersi...

Descripción completa

Detalles Bibliográficos
Autores principales: McGuire, Joseph A, Hayanga, Jeremiah, Barry, Charles, Carpenter, Anna, Frye, Benjamin, Hughes, James, Schwartzman, David, Hayanga, Heather
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9585508/
https://www.ncbi.nlm.nih.gov/pubmed/36304382
http://dx.doi.org/10.7759/cureus.29391
_version_ 1784813508708270080
author McGuire, Joseph A
Hayanga, Jeremiah
Barry, Charles
Carpenter, Anna
Frye, Benjamin
Hughes, James
Schwartzman, David
Hayanga, Heather
author_facet McGuire, Joseph A
Hayanga, Jeremiah
Barry, Charles
Carpenter, Anna
Frye, Benjamin
Hughes, James
Schwartzman, David
Hayanga, Heather
author_sort McGuire, Joseph A
collection PubMed
description Intraoperative defibrillation secondary to the usage of electrocautery in a patient with a cardiovascular implantable electronic device is a rare occurrence, and below-the-umbilicus electrocautery use causing inadvertent defibrillation is a near-zero risk. Defibrillation secondary to electrodispersive pad (EDP) radiofrequency dispersion has only ever been theorized. In this report, we describe the case of a 67-year-old male with an automatic implantable cardioverter defibrillator (AICD) undergoing a robotic-assisted left anterior total hip arthroplasty for left hip osteoarthritis who experienced inadvertent intraoperative defibrillation concurrent with electrocautery usage. The defibrillations ceased following contralateral EDP repositioning and application of a donut magnet overlying the patient’s AICD.
format Online
Article
Text
id pubmed-9585508
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-95855082022-10-26 Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement McGuire, Joseph A Hayanga, Jeremiah Barry, Charles Carpenter, Anna Frye, Benjamin Hughes, James Schwartzman, David Hayanga, Heather Cureus Anesthesiology Intraoperative defibrillation secondary to the usage of electrocautery in a patient with a cardiovascular implantable electronic device is a rare occurrence, and below-the-umbilicus electrocautery use causing inadvertent defibrillation is a near-zero risk. Defibrillation secondary to electrodispersive pad (EDP) radiofrequency dispersion has only ever been theorized. In this report, we describe the case of a 67-year-old male with an automatic implantable cardioverter defibrillator (AICD) undergoing a robotic-assisted left anterior total hip arthroplasty for left hip osteoarthritis who experienced inadvertent intraoperative defibrillation concurrent with electrocautery usage. The defibrillations ceased following contralateral EDP repositioning and application of a donut magnet overlying the patient’s AICD. Cureus 2022-09-21 /pmc/articles/PMC9585508/ /pubmed/36304382 http://dx.doi.org/10.7759/cureus.29391 Text en Copyright © 2022, McGuire et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Anesthesiology
McGuire, Joseph A
Hayanga, Jeremiah
Barry, Charles
Carpenter, Anna
Frye, Benjamin
Hughes, James
Schwartzman, David
Hayanga, Heather
Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement
title Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement
title_full Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement
title_fullStr Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement
title_full_unstemmed Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement
title_short Inadvertent Intraoperative Defibrillation Secondary to Electrocautery Grounding Pad Placement
title_sort inadvertent intraoperative defibrillation secondary to electrocautery grounding pad placement
topic Anesthesiology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9585508/
https://www.ncbi.nlm.nih.gov/pubmed/36304382
http://dx.doi.org/10.7759/cureus.29391
work_keys_str_mv AT mcguirejosepha inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement
AT hayangajeremiah inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement
AT barrycharles inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement
AT carpenteranna inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement
AT fryebenjamin inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement
AT hughesjames inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement
AT schwartzmandavid inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement
AT hayangaheather inadvertentintraoperativedefibrillationsecondarytoelectrocauterygroundingpadplacement