A comparison of fluid status determination using bioelectric impedance and the isotope dilution method in hemodialysis and peritoneal dialysis patients

We aimed to compare fluid status as determined by multifrequency bioimpedance spectroscopy (MF-BIS, Xitron 4200, USA) with that determined by the isotope dilution method among a contemporary Chinese cohort. Healthy Chinese subjects (HS, n = 30) were recruited in Zhengzhou. Hemodialysis (HD, n = 49)...

Descripción completa

Detalles Bibliográficos
Autores principales: Dou, Yanna, Gong, Yanan, Li, Afang, Wang, Peipei, Song, Dongyan, Wang, Yuyang, Liu, Dong, Cheng, Genyang, Zhao, Zhanzheng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Taylor & Francis 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9586612/
https://www.ncbi.nlm.nih.gov/pubmed/36262107
http://dx.doi.org/10.1080/0886022X.2022.2132960
_version_ 1784813724533522432
author Dou, Yanna
Gong, Yanan
Li, Afang
Wang, Peipei
Song, Dongyan
Wang, Yuyang
Liu, Dong
Cheng, Genyang
Zhao, Zhanzheng
author_facet Dou, Yanna
Gong, Yanan
Li, Afang
Wang, Peipei
Song, Dongyan
Wang, Yuyang
Liu, Dong
Cheng, Genyang
Zhao, Zhanzheng
author_sort Dou, Yanna
collection PubMed
description We aimed to compare fluid status as determined by multifrequency bioimpedance spectroscopy (MF-BIS, Xitron 4200, USA) with that determined by the isotope dilution method among a contemporary Chinese cohort. Healthy Chinese subjects (HS, n = 30) were recruited in Zhengzhou. Hemodialysis (HD, n = 49) and peritoneal dialysis (PD, n = 48) patients were screened at the First Affiliated Hospital of Zhengzhou University. Total body water (TBW) and extracellular water (ECW) were measured by deuterium (TBW(D)) and bromide (ECW(Br)) dilution, respectively, and by MF-BIS using the Moissl equation (ME). The results of MF-BIS were compared to the reference method by Pearson analysis and Bland–Altman analysis in the three groups. The accuracy of overhydration as determined by MF-BIS was analyzed by receiver operating characteristic (ROC) curves. The TBW(D) and TBW(ME) values were 34.67 ± 7.31 and 35.41 ± 5.76 L, 37.30 ± 8.58 and 37.02 ± 8.10 L, and 38.61 ± 10.02 and 38.44 ± 7.59 L in the HS, HD and PD groups, respectively. The ECW(Br) and ECW(ME) values were 14.88 ± 3.33 and 15.53 ± 2.39 L, 16.24 ± 5.08 and 16.90 ± 3.93 L, and 19.08 ± 6.41 and 18.23 ± 3.61 L in the HS, HD and PD groups, respectively. The mean bias between TBW(D) and TBW(ME) was −0.74 L, 0.28 L, and 0.17 L in the HS, HD and PD groups, respectively. The mean bias between ECW(Br) and ECW(ME) was −0.65 L, −0.66 L, and 0.85 L in the HS, HD and PD groups, respectively. Compared to the ECW(Br)/TBW(D) ratio, the area under the ROC curve (AUC) of the ECW(ME)/TBW(ME) ratio for the diagnosis of overhydration was 0.76 and 0.68 in the HD and PD groups, respectively. In summary, MF-BIS with ME could be used in Chinese HD and PD patients.
format Online
Article
Text
id pubmed-9586612
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Taylor & Francis
record_format MEDLINE/PubMed
spelling pubmed-95866122022-10-22 A comparison of fluid status determination using bioelectric impedance and the isotope dilution method in hemodialysis and peritoneal dialysis patients Dou, Yanna Gong, Yanan Li, Afang Wang, Peipei Song, Dongyan Wang, Yuyang Liu, Dong Cheng, Genyang Zhao, Zhanzheng Ren Fail Clinical Study We aimed to compare fluid status as determined by multifrequency bioimpedance spectroscopy (MF-BIS, Xitron 4200, USA) with that determined by the isotope dilution method among a contemporary Chinese cohort. Healthy Chinese subjects (HS, n = 30) were recruited in Zhengzhou. Hemodialysis (HD, n = 49) and peritoneal dialysis (PD, n = 48) patients were screened at the First Affiliated Hospital of Zhengzhou University. Total body water (TBW) and extracellular water (ECW) were measured by deuterium (TBW(D)) and bromide (ECW(Br)) dilution, respectively, and by MF-BIS using the Moissl equation (ME). The results of MF-BIS were compared to the reference method by Pearson analysis and Bland–Altman analysis in the three groups. The accuracy of overhydration as determined by MF-BIS was analyzed by receiver operating characteristic (ROC) curves. The TBW(D) and TBW(ME) values were 34.67 ± 7.31 and 35.41 ± 5.76 L, 37.30 ± 8.58 and 37.02 ± 8.10 L, and 38.61 ± 10.02 and 38.44 ± 7.59 L in the HS, HD and PD groups, respectively. The ECW(Br) and ECW(ME) values were 14.88 ± 3.33 and 15.53 ± 2.39 L, 16.24 ± 5.08 and 16.90 ± 3.93 L, and 19.08 ± 6.41 and 18.23 ± 3.61 L in the HS, HD and PD groups, respectively. The mean bias between TBW(D) and TBW(ME) was −0.74 L, 0.28 L, and 0.17 L in the HS, HD and PD groups, respectively. The mean bias between ECW(Br) and ECW(ME) was −0.65 L, −0.66 L, and 0.85 L in the HS, HD and PD groups, respectively. Compared to the ECW(Br)/TBW(D) ratio, the area under the ROC curve (AUC) of the ECW(ME)/TBW(ME) ratio for the diagnosis of overhydration was 0.76 and 0.68 in the HD and PD groups, respectively. In summary, MF-BIS with ME could be used in Chinese HD and PD patients. Taylor & Francis 2022-10-19 /pmc/articles/PMC9586612/ /pubmed/36262107 http://dx.doi.org/10.1080/0886022X.2022.2132960 Text en © 2022 The Author(s). Published by Informa UK Limited, trading as Taylor & Francis Group. https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial License (http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ), which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Clinical Study
Dou, Yanna
Gong, Yanan
Li, Afang
Wang, Peipei
Song, Dongyan
Wang, Yuyang
Liu, Dong
Cheng, Genyang
Zhao, Zhanzheng
A comparison of fluid status determination using bioelectric impedance and the isotope dilution method in hemodialysis and peritoneal dialysis patients
title A comparison of fluid status determination using bioelectric impedance and the isotope dilution method in hemodialysis and peritoneal dialysis patients
title_full A comparison of fluid status determination using bioelectric impedance and the isotope dilution method in hemodialysis and peritoneal dialysis patients
title_fullStr A comparison of fluid status determination using bioelectric impedance and the isotope dilution method in hemodialysis and peritoneal dialysis patients
title_full_unstemmed A comparison of fluid status determination using bioelectric impedance and the isotope dilution method in hemodialysis and peritoneal dialysis patients
title_short A comparison of fluid status determination using bioelectric impedance and the isotope dilution method in hemodialysis and peritoneal dialysis patients
title_sort comparison of fluid status determination using bioelectric impedance and the isotope dilution method in hemodialysis and peritoneal dialysis patients
topic Clinical Study
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9586612/
https://www.ncbi.nlm.nih.gov/pubmed/36262107
http://dx.doi.org/10.1080/0886022X.2022.2132960
work_keys_str_mv AT douyanna acomparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT gongyanan acomparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT liafang acomparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT wangpeipei acomparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT songdongyan acomparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT wangyuyang acomparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT liudong acomparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT chenggenyang acomparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT zhaozhanzheng acomparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT douyanna comparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT gongyanan comparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT liafang comparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT wangpeipei comparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT songdongyan comparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT wangyuyang comparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT liudong comparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT chenggenyang comparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients
AT zhaozhanzheng comparisonoffluidstatusdeterminationusingbioelectricimpedanceandtheisotopedilutionmethodinhemodialysisandperitonealdialysispatients