Cargando…
Application report of automatic unlocking baseplate in radiotherapy
PURPOSE: To reduce the potential risk during radiotherapy treatment of patients with head and neck tumors, we improved upon the design of an existing immobilization device by adding a feature to improve patient safety during emergency releases, and we verified its clinical application. METHOD: We de...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9588263/ https://www.ncbi.nlm.nih.gov/pubmed/36094026 http://dx.doi.org/10.1002/acm2.13778 |
_version_ | 1784814091373641728 |
---|---|
author | Li, Lintao Wang, Xianliang Xin, Xin Fan, Ming Lu, Shun Wang, Wei Yin, Gang |
author_facet | Li, Lintao Wang, Xianliang Xin, Xin Fan, Ming Lu, Shun Wang, Wei Yin, Gang |
author_sort | Li, Lintao |
collection | PubMed |
description | PURPOSE: To reduce the potential risk during radiotherapy treatment of patients with head and neck tumors, we improved upon the design of an existing immobilization device by adding a feature to improve patient safety during emergency releases, and we verified its clinical application. METHOD: We designed an improved automatic unlocking baseplate (AUB), and conducted a dosimetry comparison with Solo Align Full Body System (SAFBS, Klarity, China). The dosimetry comparison included dose‐attenuation measurements and results from human simulation. We selected four points for measurement to allow comparison between the SAFBS and our AUB. A simulated human body model was used for CT scanning, whereby the target area and structure and simulated radiotherapy plan were conducted according to the American Academy of Pain Medicine Task Group–119 report (TG‐119), whereby the dose differences were compared. The purpose of the clinical test was to verify the reliability of the AUB system in practical clinical applications. The application tests were conducted in CT simulation (CT‐sim) and treatment rooms. The test included assessments of the stability of the system and the reliability of our device. RESULTS: The dose‐attenuation measurements of the two baseplates were as follows: The transmission values with our unlocking system were 0.10% higher at the first point and 0.67% lower at the third. The same dose was obtained at points 2 and 4. In the simulation study, the PTV of the AUB was lower than that of the SAFBS, including 0.39% lower D (99) and 0.18% lower D (90). Among the organ‐at‐risk doses, the average dose of the AUB in the spinal cord was 0.6% higher than that of the SAFBS, and the average dose in the left and right parotid glands was more than 1.4% lower than that of SAFBS. The clinical test results were applied in treatment room and a CT‐sim room, which show a 100% success rate after being unlocked more than 5000 times. CONCLUSION: The AUB designed for head and neck patients had good functional versatility, the dose distribution met the requirements, and the automatic unlocking function was demonstrated to be stable and reliable. |
format | Online Article Text |
id | pubmed-9588263 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-95882632022-10-25 Application report of automatic unlocking baseplate in radiotherapy Li, Lintao Wang, Xianliang Xin, Xin Fan, Ming Lu, Shun Wang, Wei Yin, Gang J Appl Clin Med Phys Technical Notes PURPOSE: To reduce the potential risk during radiotherapy treatment of patients with head and neck tumors, we improved upon the design of an existing immobilization device by adding a feature to improve patient safety during emergency releases, and we verified its clinical application. METHOD: We designed an improved automatic unlocking baseplate (AUB), and conducted a dosimetry comparison with Solo Align Full Body System (SAFBS, Klarity, China). The dosimetry comparison included dose‐attenuation measurements and results from human simulation. We selected four points for measurement to allow comparison between the SAFBS and our AUB. A simulated human body model was used for CT scanning, whereby the target area and structure and simulated radiotherapy plan were conducted according to the American Academy of Pain Medicine Task Group–119 report (TG‐119), whereby the dose differences were compared. The purpose of the clinical test was to verify the reliability of the AUB system in practical clinical applications. The application tests were conducted in CT simulation (CT‐sim) and treatment rooms. The test included assessments of the stability of the system and the reliability of our device. RESULTS: The dose‐attenuation measurements of the two baseplates were as follows: The transmission values with our unlocking system were 0.10% higher at the first point and 0.67% lower at the third. The same dose was obtained at points 2 and 4. In the simulation study, the PTV of the AUB was lower than that of the SAFBS, including 0.39% lower D (99) and 0.18% lower D (90). Among the organ‐at‐risk doses, the average dose of the AUB in the spinal cord was 0.6% higher than that of the SAFBS, and the average dose in the left and right parotid glands was more than 1.4% lower than that of SAFBS. The clinical test results were applied in treatment room and a CT‐sim room, which show a 100% success rate after being unlocked more than 5000 times. CONCLUSION: The AUB designed for head and neck patients had good functional versatility, the dose distribution met the requirements, and the automatic unlocking function was demonstrated to be stable and reliable. John Wiley and Sons Inc. 2022-09-12 /pmc/articles/PMC9588263/ /pubmed/36094026 http://dx.doi.org/10.1002/acm2.13778 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Notes Li, Lintao Wang, Xianliang Xin, Xin Fan, Ming Lu, Shun Wang, Wei Yin, Gang Application report of automatic unlocking baseplate in radiotherapy |
title | Application report of automatic unlocking baseplate in radiotherapy |
title_full | Application report of automatic unlocking baseplate in radiotherapy |
title_fullStr | Application report of automatic unlocking baseplate in radiotherapy |
title_full_unstemmed | Application report of automatic unlocking baseplate in radiotherapy |
title_short | Application report of automatic unlocking baseplate in radiotherapy |
title_sort | application report of automatic unlocking baseplate in radiotherapy |
topic | Technical Notes |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9588263/ https://www.ncbi.nlm.nih.gov/pubmed/36094026 http://dx.doi.org/10.1002/acm2.13778 |
work_keys_str_mv | AT lilintao applicationreportofautomaticunlockingbaseplateinradiotherapy AT wangxianliang applicationreportofautomaticunlockingbaseplateinradiotherapy AT xinxin applicationreportofautomaticunlockingbaseplateinradiotherapy AT fanming applicationreportofautomaticunlockingbaseplateinradiotherapy AT lushun applicationreportofautomaticunlockingbaseplateinradiotherapy AT wangwei applicationreportofautomaticunlockingbaseplateinradiotherapy AT yingang applicationreportofautomaticunlockingbaseplateinradiotherapy |