A novel blood pressure monitoring technique by smart HUAWEI WATCH: A validation study according to the ANSI/AAMI/ISO 81060-2:2018 guidelines

BACKGROUND: Given the rapid innovation of wearable technology, additional physical indicators can be detected, and blood pressure (BP) has become the focus of many emerging medical-device manufacturers. This study aimed to validate the accuracy of the newly developed HUAWEI WATCH in BP monitoring, a...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Lili, Xian, Hong, Guo, Jiajun, Li, Weihao, Wang, Jiaqi, Chen, Qing, Fu, Xiaoyu, Li, Hongbao, Chen, Qin, Zhang, Wei, Chen, Yucheng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9592900/
https://www.ncbi.nlm.nih.gov/pubmed/36304535
http://dx.doi.org/10.3389/fcvm.2022.923655
_version_ 1784815034645348352
author Wang, Lili
Xian, Hong
Guo, Jiajun
Li, Weihao
Wang, Jiaqi
Chen, Qing
Fu, Xiaoyu
Li, Hongbao
Chen, Qin
Zhang, Wei
Chen, Yucheng
author_facet Wang, Lili
Xian, Hong
Guo, Jiajun
Li, Weihao
Wang, Jiaqi
Chen, Qing
Fu, Xiaoyu
Li, Hongbao
Chen, Qin
Zhang, Wei
Chen, Yucheng
author_sort Wang, Lili
collection PubMed
description BACKGROUND: Given the rapid innovation of wearable technology, additional physical indicators can be detected, and blood pressure (BP) has become the focus of many emerging medical-device manufacturers. This study aimed to validate the accuracy of the newly developed HUAWEI WATCH in BP monitoring, according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO 81060-2:2018) guidelines. MATERIALS AND METHODS: The same arm sequential BP measurement was applied. One validation included four reference BP measurements taken simultaneously by two independent observers using a mercury sphygmomanometer, alternating with three test-watch measurements. Each test-watch measurement was compared against the average of the previous and subsequent reference BP readings. Two criteria were required for validation: (1) a mean BP difference of 5 mm Hg or less, with a standard deviation (SD) of 8 mm Hg or less for systolic blood pressure (SBP) and diastolic blood pressure (DBP) in the 255 pairs of measurements, and (2) an SD for the of 85 averaged BP differences within the threshold defined by the mean test-reference BP difference listed in the ANSI/AAMI/ISO 81060-2:2018 guidelines. RESULTS: The mean age of the 85 participants was 48 ± 18 years (range: 21–85), and 53 (62.4%) were male. The mean differences between the test and reference BPs were -0.25 ± 5.62 mm Hg and -1.33 ± 6.81 mm Hg for SBP and DBP, respectively (according to Criterion 1). The mean differences between the test BPs and reference BPs were -0.25 ± 5.00 mm Hg and -1.33 ± 6.31 mm Hg for SBP and DBP, respectively, according to Criterion 2. CONCLUSION: Blood pressure measurement using the HUAWEI WATCH showed excellent consistency with reference BPs, and fulfilled both validation criteria of the guidelines, show its promise as a wearable device for BP self-monitoring.
format Online
Article
Text
id pubmed-9592900
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-95929002022-10-26 A novel blood pressure monitoring technique by smart HUAWEI WATCH: A validation study according to the ANSI/AAMI/ISO 81060-2:2018 guidelines Wang, Lili Xian, Hong Guo, Jiajun Li, Weihao Wang, Jiaqi Chen, Qing Fu, Xiaoyu Li, Hongbao Chen, Qin Zhang, Wei Chen, Yucheng Front Cardiovasc Med Cardiovascular Medicine BACKGROUND: Given the rapid innovation of wearable technology, additional physical indicators can be detected, and blood pressure (BP) has become the focus of many emerging medical-device manufacturers. This study aimed to validate the accuracy of the newly developed HUAWEI WATCH in BP monitoring, according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO 81060-2:2018) guidelines. MATERIALS AND METHODS: The same arm sequential BP measurement was applied. One validation included four reference BP measurements taken simultaneously by two independent observers using a mercury sphygmomanometer, alternating with three test-watch measurements. Each test-watch measurement was compared against the average of the previous and subsequent reference BP readings. Two criteria were required for validation: (1) a mean BP difference of 5 mm Hg or less, with a standard deviation (SD) of 8 mm Hg or less for systolic blood pressure (SBP) and diastolic blood pressure (DBP) in the 255 pairs of measurements, and (2) an SD for the of 85 averaged BP differences within the threshold defined by the mean test-reference BP difference listed in the ANSI/AAMI/ISO 81060-2:2018 guidelines. RESULTS: The mean age of the 85 participants was 48 ± 18 years (range: 21–85), and 53 (62.4%) were male. The mean differences between the test and reference BPs were -0.25 ± 5.62 mm Hg and -1.33 ± 6.81 mm Hg for SBP and DBP, respectively (according to Criterion 1). The mean differences between the test BPs and reference BPs were -0.25 ± 5.00 mm Hg and -1.33 ± 6.31 mm Hg for SBP and DBP, respectively, according to Criterion 2. CONCLUSION: Blood pressure measurement using the HUAWEI WATCH showed excellent consistency with reference BPs, and fulfilled both validation criteria of the guidelines, show its promise as a wearable device for BP self-monitoring. Frontiers Media S.A. 2022-10-11 /pmc/articles/PMC9592900/ /pubmed/36304535 http://dx.doi.org/10.3389/fcvm.2022.923655 Text en Copyright © 2022 Wang, Xian, Guo, Li, Wang, Chen, Fu, Li, Chen, Zhang and Chen. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Cardiovascular Medicine
Wang, Lili
Xian, Hong
Guo, Jiajun
Li, Weihao
Wang, Jiaqi
Chen, Qing
Fu, Xiaoyu
Li, Hongbao
Chen, Qin
Zhang, Wei
Chen, Yucheng
A novel blood pressure monitoring technique by smart HUAWEI WATCH: A validation study according to the ANSI/AAMI/ISO 81060-2:2018 guidelines
title A novel blood pressure monitoring technique by smart HUAWEI WATCH: A validation study according to the ANSI/AAMI/ISO 81060-2:2018 guidelines
title_full A novel blood pressure monitoring technique by smart HUAWEI WATCH: A validation study according to the ANSI/AAMI/ISO 81060-2:2018 guidelines
title_fullStr A novel blood pressure monitoring technique by smart HUAWEI WATCH: A validation study according to the ANSI/AAMI/ISO 81060-2:2018 guidelines
title_full_unstemmed A novel blood pressure monitoring technique by smart HUAWEI WATCH: A validation study according to the ANSI/AAMI/ISO 81060-2:2018 guidelines
title_short A novel blood pressure monitoring technique by smart HUAWEI WATCH: A validation study according to the ANSI/AAMI/ISO 81060-2:2018 guidelines
title_sort novel blood pressure monitoring technique by smart huawei watch: a validation study according to the ansi/aami/iso 81060-2:2018 guidelines
topic Cardiovascular Medicine
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9592900/
https://www.ncbi.nlm.nih.gov/pubmed/36304535
http://dx.doi.org/10.3389/fcvm.2022.923655
work_keys_str_mv AT wanglili anovelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT xianhong anovelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT guojiajun anovelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT liweihao anovelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT wangjiaqi anovelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT chenqing anovelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT fuxiaoyu anovelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT lihongbao anovelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT chenqin anovelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT zhangwei anovelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT chenyucheng anovelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT wanglili novelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT xianhong novelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT guojiajun novelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT liweihao novelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT wangjiaqi novelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT chenqing novelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT fuxiaoyu novelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT lihongbao novelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT chenqin novelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT zhangwei novelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines
AT chenyucheng novelbloodpressuremonitoringtechniquebysmarthuaweiwatchavalidationstudyaccordingtotheansiaamiiso8106022018guidelines