Decreased Sensitivity of Rapid Antigen Test Is Associated with a Lower Viral Load of Omicron than Delta SARS-CoV-2 Variant

Large-scale screening for SARS-CoV-2 infection is an important tool for epidemic prevention and control. The appearance of new variants associated with specific mutations can call into question the effectiveness of rapid diagnostic tests (RDTs) deployed massively at national and international levels...

Descripción completa

Detalles Bibliográficos
Autores principales: Cocherie, Théophile, Bastide, Mathilda, Sakhi, Soraya, Zafilaza, Karen, Flandre, Philippe, Leducq, Valentin, Jary, Aude, Burrel, Sonia, Louet, Martine, Calvez, Vincent, Marcelin, Anne-Geneviève, Marot, Stéphane
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Society for Microbiology 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9603576/
https://www.ncbi.nlm.nih.gov/pubmed/36125269
http://dx.doi.org/10.1128/spectrum.01922-22
_version_ 1784817588781449216
author Cocherie, Théophile
Bastide, Mathilda
Sakhi, Soraya
Zafilaza, Karen
Flandre, Philippe
Leducq, Valentin
Jary, Aude
Burrel, Sonia
Louet, Martine
Calvez, Vincent
Marcelin, Anne-Geneviève
Marot, Stéphane
author_facet Cocherie, Théophile
Bastide, Mathilda
Sakhi, Soraya
Zafilaza, Karen
Flandre, Philippe
Leducq, Valentin
Jary, Aude
Burrel, Sonia
Louet, Martine
Calvez, Vincent
Marcelin, Anne-Geneviève
Marot, Stéphane
author_sort Cocherie, Théophile
collection PubMed
description Large-scale screening for SARS-CoV-2 infection is an important tool for epidemic prevention and control. The appearance of new variants associated with specific mutations can call into question the effectiveness of rapid diagnostic tests (RDTs) deployed massively at national and international levels. We compared the clinical and virological characteristics of individuals infected by Delta or Omicron variants to assess which factors were associated with a reduced performance of RDT. A commercially available RDT as well as the evaluation of the viral load (VL) and the detection of replicate intermediates (RIs) were carried out retrospectively on positive SARS-CoV-2 nasopharyngeal specimens from health care workers of the Pitié-Salpêtrière Hospital infected by the Delta or Omicron variant between July 2021 and January 2022. Of the 205 samples analyzed (104 from individuals infected with Delta and 101 with Omicron), 176 were analyzed by RDT and 200 by RT-PCR for VL and RIs. The sensitivity of the TDR for Omicron was significantly lower than that observed for Delta (53.8% versus 74.7%, respectively, P < 0.01). Moreover, the Delta VL was significantly higher than that measured for Omicron (median Ct 21.2 versus 24.1, respectively, P < 0.01) and associated with the positivity of the RDT in multivariate analysis. We demonstrate a lower RDT sensitivity associated with a lower VL at the time of diagnosis on Omicron-infected individuals in comparison to those infected with the Delta variant. This RDT lower sensitivity should be taken into account in the large-scale screening strategy and in particular in case of strong suspicion of infection where testing should be repeated. IMPORTANCE Previous reports have shown a variability in the diagnostic performance of RDTs. In the era of SARS-CoV-2 variants and the use of RDT, mutation associated with these variants could affect the test performance. We evaluate the sensitivity of the RDT Panbio COVID-19 Ag (Abbott) with two variants of concern (VOC), the Delta and Omicron variants. In order to investigate whether clinical characteristics or virological characteristics can affect this sensitivity, we collected clinical information and performed a specific RT-PCR that detected the RIs as a marker of the viral replication and viral cycle stage. Our results showed that Omicron was less detected than the Delta variant. A lower viral load of Omicron variant in comparison to Delta variant explained this decreased sensitivity, even if they are at the same stage of the disease and the viral cycle and should be taken into account with the use of RDT as diagnostic tool.
format Online
Article
Text
id pubmed-9603576
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher American Society for Microbiology
record_format MEDLINE/PubMed
spelling pubmed-96035762022-10-27 Decreased Sensitivity of Rapid Antigen Test Is Associated with a Lower Viral Load of Omicron than Delta SARS-CoV-2 Variant Cocherie, Théophile Bastide, Mathilda Sakhi, Soraya Zafilaza, Karen Flandre, Philippe Leducq, Valentin Jary, Aude Burrel, Sonia Louet, Martine Calvez, Vincent Marcelin, Anne-Geneviève Marot, Stéphane Microbiol Spectr Observation Large-scale screening for SARS-CoV-2 infection is an important tool for epidemic prevention and control. The appearance of new variants associated with specific mutations can call into question the effectiveness of rapid diagnostic tests (RDTs) deployed massively at national and international levels. We compared the clinical and virological characteristics of individuals infected by Delta or Omicron variants to assess which factors were associated with a reduced performance of RDT. A commercially available RDT as well as the evaluation of the viral load (VL) and the detection of replicate intermediates (RIs) were carried out retrospectively on positive SARS-CoV-2 nasopharyngeal specimens from health care workers of the Pitié-Salpêtrière Hospital infected by the Delta or Omicron variant between July 2021 and January 2022. Of the 205 samples analyzed (104 from individuals infected with Delta and 101 with Omicron), 176 were analyzed by RDT and 200 by RT-PCR for VL and RIs. The sensitivity of the TDR for Omicron was significantly lower than that observed for Delta (53.8% versus 74.7%, respectively, P < 0.01). Moreover, the Delta VL was significantly higher than that measured for Omicron (median Ct 21.2 versus 24.1, respectively, P < 0.01) and associated with the positivity of the RDT in multivariate analysis. We demonstrate a lower RDT sensitivity associated with a lower VL at the time of diagnosis on Omicron-infected individuals in comparison to those infected with the Delta variant. This RDT lower sensitivity should be taken into account in the large-scale screening strategy and in particular in case of strong suspicion of infection where testing should be repeated. IMPORTANCE Previous reports have shown a variability in the diagnostic performance of RDTs. In the era of SARS-CoV-2 variants and the use of RDT, mutation associated with these variants could affect the test performance. We evaluate the sensitivity of the RDT Panbio COVID-19 Ag (Abbott) with two variants of concern (VOC), the Delta and Omicron variants. In order to investigate whether clinical characteristics or virological characteristics can affect this sensitivity, we collected clinical information and performed a specific RT-PCR that detected the RIs as a marker of the viral replication and viral cycle stage. Our results showed that Omicron was less detected than the Delta variant. A lower viral load of Omicron variant in comparison to Delta variant explained this decreased sensitivity, even if they are at the same stage of the disease and the viral cycle and should be taken into account with the use of RDT as diagnostic tool. American Society for Microbiology 2022-09-20 /pmc/articles/PMC9603576/ /pubmed/36125269 http://dx.doi.org/10.1128/spectrum.01922-22 Text en Copyright © 2022 Cocherie et al. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 International license (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Observation
Cocherie, Théophile
Bastide, Mathilda
Sakhi, Soraya
Zafilaza, Karen
Flandre, Philippe
Leducq, Valentin
Jary, Aude
Burrel, Sonia
Louet, Martine
Calvez, Vincent
Marcelin, Anne-Geneviève
Marot, Stéphane
Decreased Sensitivity of Rapid Antigen Test Is Associated with a Lower Viral Load of Omicron than Delta SARS-CoV-2 Variant
title Decreased Sensitivity of Rapid Antigen Test Is Associated with a Lower Viral Load of Omicron than Delta SARS-CoV-2 Variant
title_full Decreased Sensitivity of Rapid Antigen Test Is Associated with a Lower Viral Load of Omicron than Delta SARS-CoV-2 Variant
title_fullStr Decreased Sensitivity of Rapid Antigen Test Is Associated with a Lower Viral Load of Omicron than Delta SARS-CoV-2 Variant
title_full_unstemmed Decreased Sensitivity of Rapid Antigen Test Is Associated with a Lower Viral Load of Omicron than Delta SARS-CoV-2 Variant
title_short Decreased Sensitivity of Rapid Antigen Test Is Associated with a Lower Viral Load of Omicron than Delta SARS-CoV-2 Variant
title_sort decreased sensitivity of rapid antigen test is associated with a lower viral load of omicron than delta sars-cov-2 variant
topic Observation
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9603576/
https://www.ncbi.nlm.nih.gov/pubmed/36125269
http://dx.doi.org/10.1128/spectrum.01922-22
work_keys_str_mv AT cocherietheophile decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant
AT bastidemathilda decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant
AT sakhisoraya decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant
AT zafilazakaren decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant
AT flandrephilippe decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant
AT leducqvalentin decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant
AT jaryaude decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant
AT burrelsonia decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant
AT louetmartine decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant
AT calvezvincent decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant
AT marcelinannegenevieve decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant
AT marotstephane decreasedsensitivityofrapidantigentestisassociatedwithalowerviralloadofomicronthandeltasarscov2variant