Cargando…

Comparison between Fourth-Generation FloTrac/Vigileo System and Continuous Thermodilution Technique for Cardiac Output Estimation after Time Adjustment during Off-Pump Coronary Artery Bypass Graft Surgery: A Retrospective Cohort Study

(1) Background: Previous studies reported limited performance of arterial pressure waveform-based cardiac output (CO) estimation (FloTrac/Vigileo system; CO-FloTrac) compared with the intermittent thermodilution technique (CO(int)). However, errors due to bolus maneuver and intermittent measurements...

Descripción completa

Detalles Bibliográficos
Autores principales: Oh, Chahyun, Lee, Soomin, Oh, Pyeonghwa, Chung, Woosuk, Ko, Youngkwon, Yoon, Seok-Hwa, Kim, Yoon-Hee, Ji, Sung-Mi, Hong, Boohwi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9605331/
https://www.ncbi.nlm.nih.gov/pubmed/36294414
http://dx.doi.org/10.3390/jcm11206093
Descripción
Sumario:(1) Background: Previous studies reported limited performance of arterial pressure waveform-based cardiac output (CO) estimation (FloTrac/Vigileo system; CO-FloTrac) compared with the intermittent thermodilution technique (CO(int)). However, errors due to bolus maneuver and intermittent measurements of CO(int) could limit its use as a reference. The continuous thermodilution technique (CO(cont)) may relieve such limitations. (2) Methods: The performance of CO-FloTrac was retrospectively assessed using continuous recordings of intraoperative physiological data acquired from patients who underwent off-pump coronary artery bypass graft (OPCAB) surgery with CO monitoring using both CO-FloTrac and CO(cont). Optimal time adjustments between the two measurements were determined based on R-squared values. (3) Results: A total of 134.2 h of data from 30 patients was included in the final analysis. The mean bias was –0.94 (95% CI, −1.35 to −0.52) L/min and the limits of agreements were −3.64 (95% CI, −4.44 to −3.08) L/min and 1.77 (95% CI, 1.21 to 2.57) L/min. The percentage error was 66.1% (95% CI, 52.4 to 85.8%). Depending on the time scale and the size of the exclusion zone, concordance rates ranged from 61.0% to 75.0%. (4) Conclusion: Despite the time adjustments, CO-FloTrac showed non-negligible overestimation, clinically unacceptable precision, and poor trending ability during OPCAB surgery.