Validación preclínica de un respirador de turbina para la ventilación invasiva: el respirador ACUTE-19

BACKGROUND: The severe acute respiratory syndrome-coronavirus 2 pandemic pressure on healthcare systems can exhaust ventilator resources, especially where resources are restricted. Our objective was a rapid preclinical evaluation of a newly developed turbine-based ventilator, named the ACUTE-19, for...

Descripción completa

Detalles Bibliográficos
Autores principales: Alonso-Iñigo, J.M., Mazzinari, G., Casañ-Pallardó, M., Redondo-García, J.I., Viscasillas-Monteagudo, J., Gutierrez-Bautista, A., Ramirez-Faz, J., Alonso-Pérez, P., Díaz-Lobato, S., Neto, A.S., Diaz-Cambronero, O., Argente-Navarro, P., Gama de Abreu, M., Pelosi, P., Schultz, M.J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Sociedad Española de Anestesiología, Reanimación y Terapéutica del Dolor. Published by Elsevier España, S.L.U. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9617684/
https://www.ncbi.nlm.nih.gov/pubmed/36337377
http://dx.doi.org/10.1016/j.redar.2021.09.008
Descripción
Sumario:BACKGROUND: The severe acute respiratory syndrome-coronavirus 2 pandemic pressure on healthcare systems can exhaust ventilator resources, especially where resources are restricted. Our objective was a rapid preclinical evaluation of a newly developed turbine-based ventilator, named the ACUTE-19, for invasive ventilation. METHODS: Validation consisted of (a) testing tidal volume delivery in 11 simulated models, with various resistances and compliances; (b) comparison with a commercial ventilator (VIVO-50) adapting the United Kingdom Medicines and Healthcare products Regulatory Agency-recommendations for rapidly manufactured ventilators; and (c) in vivo testing in a sheep before and after inducing acute respiratory distress syndrome by saline lavage. RESULTS: Differences in tidal volume in the simulated models were marginally different (largest difference 33 ml [95% CI 31 to 36]; P < .001). Plateau pressure was not different (−0.3 cmH(2)O [95% CI −0.9 to 0.3]; P = .409), and positive end-expiratory pressure was marginally different (0.3 cmH(2)O [95% CI 0.2 to 0.3]; P < .001) between the ACUTE-19 and the commercial ventilator. Bland-Altman analyses showed good agreement (mean bias −0.29 [limits of agreement 0.82 to −1.42], and mean bias 0.56 [limits of agreement 1.94 to −0.81], at a plateau pressure of 15 and 30 cmH(2)O, respectively). The ACUTE-19 achieved optimal oxygenation and ventilation before and after acute respiratory distress syndrome induction. CONCLUSIONS: The ACUTE-19 performed accurately in simulated and animal models yielding a comparable performance with a VIVO-50 commercial device. The ACUTE-19 can provide the basis for the development of a future affordable commercial ventilator.