PSAT096 Development and Validation of Mitotane Determination in Plasma Samples by LC-DAD
INTRODUCTION: Mitotane (o,p'–DDD) is the drug of choice for adrenocortical carcinomas (ACC) and its measurement in plasma is essential to control drug administration. OBJECTIVE: To develop and validate a simple, reliable and straightforward method for Mitotane determination in plasma samples. M...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9624653/ http://dx.doi.org/10.1210/jendso/bvac150.239 |
_version_ | 1784822285733986304 |
---|---|
author | Lima-Valassi, Helena P Marques, Anna Sylvia Ferrari Alves, Atecla Nunciata Lopes Mendonca, Berenice B |
author_facet | Lima-Valassi, Helena P Marques, Anna Sylvia Ferrari Alves, Atecla Nunciata Lopes Mendonca, Berenice B |
author_sort | Lima-Valassi, Helena P |
collection | PubMed |
description | INTRODUCTION: Mitotane (o,p'–DDD) is the drug of choice for adrenocortical carcinomas (ACC) and its measurement in plasma is essential to control drug administration. OBJECTIVE: To develop and validate a simple, reliable and straightforward method for Mitotane determination in plasma samples. METHOD: Drug free plasma samples were collected in potassium-ethylenediamine tetraacetate (K-EDTA) tubes and spiked with 1.0, 2.5, 10.0, 25.0 and 50.0 µg/mL of DDD. p,p´-DDD was used as internal standard and was added at 25.0 µg /mL to each sample, standards and controls. Samples were submitted to protein precipitation with acetonitrile and than centrifuged and 50 uL sobrenadant was injecting into a 1260 Infinity II liquid chromatography system (Agilent Technologies) and DDD and IS were detected at 230 nm by diodo array detector (DAD) in a 12 minutes isocratic mode with a solvent mixture of 60% acetonitrile and 40% formic acid in water 0.1% pump mixed, at 0.6 mL/min flow rate. Chromatography column Waters Acquity HSS T3 was kept at 28°C. We evaluated sensitivity, precision, presence of carry-over, recovery, linearity and the accuracy of this method. RESULTS: Our method resulted in a symmetrical peak shape and good baseline resolution for DDD (mitotane) and 4,4´-DDD (internal standard) with retention times of 6.0min, 6.4mim, respectively, with resolution higher than 1.5. Endogenous plasma compounds did not interfere with the evaluated peaks when blank plasma and spiked plasma with standards were compared. Linearity was assessed over the range of 1.0 to 50.0 µg/mL for DDD (R(2) > 0.9987 and a 97.8% - 105.5% extraction efficiency). Analytical sensitivity was 0.98 µg/mL, functional sensitivity was 1.00 µg/mL, intra-assay and inter-assay coefficient of variation were less than 9. 98%, carry-over was not detected, recovery ranged from 98% to 117% and high accuracy was observed (89.4% to 105.9%) with no matrix effects for mitotane measurements. Patient's samples results were compared with previous measurement by GC-MS method with a high correlation (r=0.88 and bias=-10.2%) showed that our method is usefull for routine mitotane measurement. CONCLUSION: DDD determination in plasma samples by our developed and validated method is simple, robust, efficient and sensitive for therapeutic drug monitoring and dose management in order to achieve therapeutic index of mitotane in patients with adrenocortical cancer. Presentation: Saturday, June 11, 2022 1:00 p.m. - 3:00 p.m. |
format | Online Article Text |
id | pubmed-9624653 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Oxford University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-96246532022-11-14 PSAT096 Development and Validation of Mitotane Determination in Plasma Samples by LC-DAD Lima-Valassi, Helena P Marques, Anna Sylvia Ferrari Alves, Atecla Nunciata Lopes Mendonca, Berenice B J Endocr Soc Adrenal INTRODUCTION: Mitotane (o,p'–DDD) is the drug of choice for adrenocortical carcinomas (ACC) and its measurement in plasma is essential to control drug administration. OBJECTIVE: To develop and validate a simple, reliable and straightforward method for Mitotane determination in plasma samples. METHOD: Drug free plasma samples were collected in potassium-ethylenediamine tetraacetate (K-EDTA) tubes and spiked with 1.0, 2.5, 10.0, 25.0 and 50.0 µg/mL of DDD. p,p´-DDD was used as internal standard and was added at 25.0 µg /mL to each sample, standards and controls. Samples were submitted to protein precipitation with acetonitrile and than centrifuged and 50 uL sobrenadant was injecting into a 1260 Infinity II liquid chromatography system (Agilent Technologies) and DDD and IS were detected at 230 nm by diodo array detector (DAD) in a 12 minutes isocratic mode with a solvent mixture of 60% acetonitrile and 40% formic acid in water 0.1% pump mixed, at 0.6 mL/min flow rate. Chromatography column Waters Acquity HSS T3 was kept at 28°C. We evaluated sensitivity, precision, presence of carry-over, recovery, linearity and the accuracy of this method. RESULTS: Our method resulted in a symmetrical peak shape and good baseline resolution for DDD (mitotane) and 4,4´-DDD (internal standard) with retention times of 6.0min, 6.4mim, respectively, with resolution higher than 1.5. Endogenous plasma compounds did not interfere with the evaluated peaks when blank plasma and spiked plasma with standards were compared. Linearity was assessed over the range of 1.0 to 50.0 µg/mL for DDD (R(2) > 0.9987 and a 97.8% - 105.5% extraction efficiency). Analytical sensitivity was 0.98 µg/mL, functional sensitivity was 1.00 µg/mL, intra-assay and inter-assay coefficient of variation were less than 9. 98%, carry-over was not detected, recovery ranged from 98% to 117% and high accuracy was observed (89.4% to 105.9%) with no matrix effects for mitotane measurements. Patient's samples results were compared with previous measurement by GC-MS method with a high correlation (r=0.88 and bias=-10.2%) showed that our method is usefull for routine mitotane measurement. CONCLUSION: DDD determination in plasma samples by our developed and validated method is simple, robust, efficient and sensitive for therapeutic drug monitoring and dose management in order to achieve therapeutic index of mitotane in patients with adrenocortical cancer. Presentation: Saturday, June 11, 2022 1:00 p.m. - 3:00 p.m. Oxford University Press 2022-11-01 /pmc/articles/PMC9624653/ http://dx.doi.org/10.1210/jendso/bvac150.239 Text en © The Author(s) 2022. Published by Oxford University Press on behalf of the Endocrine Society. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs licence (https://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial reproduction and distribution of the work, in any medium, provided the original work is not altered or transformed in any way, and that the work is properly cited. For commercial re-use, please contact journals.permissions@oup.com |
spellingShingle | Adrenal Lima-Valassi, Helena P Marques, Anna Sylvia Ferrari Alves, Atecla Nunciata Lopes Mendonca, Berenice B PSAT096 Development and Validation of Mitotane Determination in Plasma Samples by LC-DAD |
title | PSAT096 Development and Validation of Mitotane Determination in Plasma Samples by LC-DAD |
title_full | PSAT096 Development and Validation of Mitotane Determination in Plasma Samples by LC-DAD |
title_fullStr | PSAT096 Development and Validation of Mitotane Determination in Plasma Samples by LC-DAD |
title_full_unstemmed | PSAT096 Development and Validation of Mitotane Determination in Plasma Samples by LC-DAD |
title_short | PSAT096 Development and Validation of Mitotane Determination in Plasma Samples by LC-DAD |
title_sort | psat096 development and validation of mitotane determination in plasma samples by lc-dad |
topic | Adrenal |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9624653/ http://dx.doi.org/10.1210/jendso/bvac150.239 |
work_keys_str_mv | AT limavalassihelenap psat096developmentandvalidationofmitotanedeterminationinplasmasamplesbylcdad AT marquesannasylviaferrari psat096developmentandvalidationofmitotanedeterminationinplasmasamplesbylcdad AT alvesateclanunciatalopes psat096developmentandvalidationofmitotanedeterminationinplasmasamplesbylcdad AT mendoncabereniceb psat096developmentandvalidationofmitotanedeterminationinplasmasamplesbylcdad |