Integration of 3D nuclear imaging in 3D mapping system for ventricular tachycardia ablation in patients with implanted devices: Perfusion/voltage retrospective assessment of scar location

BACKGROUND: The identification of low-voltage proarrhythmic areas for catheter ablation of scar-mediated ventricular tachycardia (VT) remains challenging. Integration of myocardial perfusion imaging (single-photon emission computed tomography/computed tomography; SPECT/CT) and electroanatomical mapp...

Descripción completa

Detalles Bibliográficos
Autores principales: Thibault, Bernard, Richer, Louis-Philippe, McSpadden, Luke C., Ryu, Kyungmoo, Aguilar, Martin, Cadrin-Tourigny, Julia, Tadros, Rafik, Mondésert, Blandine, Rivard, Léna, Dyrda, Katia, Dubuc, Marc, Macle, Laurent, Talajic, Mario, Khairy, Paul, Guerra, Peter G., Roy, Denis, Grégoire, Jean, Harel, François
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9626736/
https://www.ncbi.nlm.nih.gov/pubmed/36340481
http://dx.doi.org/10.1016/j.hroo.2022.06.008
Descripción
Sumario:BACKGROUND: The identification of low-voltage proarrhythmic areas for catheter ablation of scar-mediated ventricular tachycardia (VT) remains challenging. Integration of myocardial perfusion imaging (single-photon emission computed tomography/computed tomography; SPECT/CT) and electroanatomical mapping (EAM) may improve delineation of the arrhythmogenic substrate. OBJECTIVE: To assess the feasibility of SPECT/CT image integration with voltage maps using the EnSite Precision system (Abbott) in patients undergoing scar-mediated VT ablation. METHODS: Patients underwent SPECT/CT imaging prior to left ventricular (LV) EAM with the EnSite Precision mapping system. The SPECT/CT, EAM data, and ablation lesions were retrospectively co-registered in the EnSite Precision system and exported for analysis. Segmental tissue viability scores were calculated based on SPECT/CT perfusion and electrogram bipolar voltage amplitude. Concordance, specificity, and sensitivity between the 2 modalities as well as the impact of SPECT/CT spatial resolution were evaluated. RESULTS: Twenty subjects (95% male, 67 ± 7 years old, left ventricular ejection fraction 36% ± 11%) underwent EAM and SPECT/CT integration. A concordance of 70% was found between EAM and SPECT/CT for identification of cardiac segments as scar vs viable, with EAM showing a 68.5% sensitivity and 76.4% specificity when using SPECT/CT as a gold standard. Projection on low-resolution 3D geometries led to an average decrease of 38% ± 22% of the voltage points used. CONCLUSION: The study demonstrated the feasibility of integrating SPECT/CT with EAM performed retrospectively for characterization of anatomical substrates during VT ablation procedures.