RF35 | PSAT262 Comparative Study of Hertel Exophthalmometer and Luedde Exophthalmometer Versus Computed Tomography for Measurement of Proptosis in Normal Subjects

BACKGROUND: Clinical evaluation of proptosis is measured by an exophthalmometer. Hertel exophthalmometer has long been considered a standard instrument for measurement of eye protrusion. The reading may be subject to error while the examiners read through the integral reflection mirrors. Luedde exop...

Descripción completa

Detalles Bibliográficos
Autores principales: Traiwanatham, Sirinapa, Sirinvaravong, Sirinart, Plengvidhya, Nattachet, Wongsripuemtet, Jitsupa, Lertlam, Supassa
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9628883/
http://dx.doi.org/10.1210/jendso/bvac150.1779
_version_ 1784823284840267776
author Traiwanatham, Sirinapa
Sirinvaravong, Sirinart
Plengvidhya, Nattachet
Wongsripuemtet, Jitsupa
Lertlam, Supassa
author_facet Traiwanatham, Sirinapa
Sirinvaravong, Sirinart
Plengvidhya, Nattachet
Wongsripuemtet, Jitsupa
Lertlam, Supassa
author_sort Traiwanatham, Sirinapa
collection PubMed
description BACKGROUND: Clinical evaluation of proptosis is measured by an exophthalmometer. Hertel exophthalmometer has long been considered a standard instrument for measurement of eye protrusion. The reading may be subject to error while the examiners read through the integral reflection mirrors. Luedde exophthalmometer, which is a ruler correcting for parallax, is easier to use and less expensive than Hertel exophthalmometer. Since CT scan of orbits or brain provides the most objective assessment of proptosis measurements, we aim to study the correlation between Luedde and Hertel exophthalmometers in comparison with the CT scan for the evaluation of proptosis in normal subjects. METHODS: This prospective study was carried out between November 2020 and January 2021. Sixty-three patients were enrolled from a single tertiary hospital. All patients underwent CT scan of brain for medical reasons. On the same day, proptosis was measured with Hertel and Dual Luedde exophthalmometer by two independent examiners. Measurement of proptosis from CT scan was performed by one experienced neuroradiologist, using 2 reference measurement techniques. Correlation coefficients between proptosis values measured by each exophthalmometer and CT scan were compared. RESULTS: The mean age of the participants was 63 years old and 54% were female. For the right eye, mean proptosis measurement was 14.9+/-1.6 mm by Hertel exophthalmometer, 14.1+/-2.3mm by Luedde exophthalmometer, 16.6+/-2.3 mm by first CT technique, and 13.1 +/- 2.2 mm by second CT technique. For the left eye, mean proptosis measurement was 15.5 +/-1.5mm by Hertel exophthalmometer, 13.7+/-2.3 mm by Luedde exophthalmometer, 16.6+/-2.3 mm by first CT technique and 13.1 +/- 2.2 mm by second CT technique. The Intraclass correlation coefficient (ICC) of Hertel exophthalmometer for proptosis measurement compared to first CT technique and second CT technique were 0.63 and 0.60 for right eye; 0.61 and 0.55 for left eye, respectively. The ICC of Luedde exophthalmometer for proptosis measurement compared to first CT technique and second CT technique were 0.78 and 0.74 for right eye; 0.77 and 0.74 for left eye, respectively. CONCLUSION: There is a good correlation between Hertel and Luedde exophthalmometers in evaluation of proptosis. However, proptosis measured by Luedde exophthalmometer shows superior correlation with CT scan. The result suggests that Luedde exophthalmometer can be routinely used for measurement of proptosis in place of Hertel exophthalmometer, given the simplicity and the inexpensiveness of the instrument. Presentation: Saturday, June 11, 2022 1:00 p.m. - 3:00 p.m., Monday, June 13, 2022 12:30 p.m. - 12:35 p.m.
format Online
Article
Text
id pubmed-9628883
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-96288832022-11-04 RF35 | PSAT262 Comparative Study of Hertel Exophthalmometer and Luedde Exophthalmometer Versus Computed Tomography for Measurement of Proptosis in Normal Subjects Traiwanatham, Sirinapa Sirinvaravong, Sirinart Plengvidhya, Nattachet Wongsripuemtet, Jitsupa Lertlam, Supassa J Endocr Soc Thyroid BACKGROUND: Clinical evaluation of proptosis is measured by an exophthalmometer. Hertel exophthalmometer has long been considered a standard instrument for measurement of eye protrusion. The reading may be subject to error while the examiners read through the integral reflection mirrors. Luedde exophthalmometer, which is a ruler correcting for parallax, is easier to use and less expensive than Hertel exophthalmometer. Since CT scan of orbits or brain provides the most objective assessment of proptosis measurements, we aim to study the correlation between Luedde and Hertel exophthalmometers in comparison with the CT scan for the evaluation of proptosis in normal subjects. METHODS: This prospective study was carried out between November 2020 and January 2021. Sixty-three patients were enrolled from a single tertiary hospital. All patients underwent CT scan of brain for medical reasons. On the same day, proptosis was measured with Hertel and Dual Luedde exophthalmometer by two independent examiners. Measurement of proptosis from CT scan was performed by one experienced neuroradiologist, using 2 reference measurement techniques. Correlation coefficients between proptosis values measured by each exophthalmometer and CT scan were compared. RESULTS: The mean age of the participants was 63 years old and 54% were female. For the right eye, mean proptosis measurement was 14.9+/-1.6 mm by Hertel exophthalmometer, 14.1+/-2.3mm by Luedde exophthalmometer, 16.6+/-2.3 mm by first CT technique, and 13.1 +/- 2.2 mm by second CT technique. For the left eye, mean proptosis measurement was 15.5 +/-1.5mm by Hertel exophthalmometer, 13.7+/-2.3 mm by Luedde exophthalmometer, 16.6+/-2.3 mm by first CT technique and 13.1 +/- 2.2 mm by second CT technique. The Intraclass correlation coefficient (ICC) of Hertel exophthalmometer for proptosis measurement compared to first CT technique and second CT technique were 0.63 and 0.60 for right eye; 0.61 and 0.55 for left eye, respectively. The ICC of Luedde exophthalmometer for proptosis measurement compared to first CT technique and second CT technique were 0.78 and 0.74 for right eye; 0.77 and 0.74 for left eye, respectively. CONCLUSION: There is a good correlation between Hertel and Luedde exophthalmometers in evaluation of proptosis. However, proptosis measured by Luedde exophthalmometer shows superior correlation with CT scan. The result suggests that Luedde exophthalmometer can be routinely used for measurement of proptosis in place of Hertel exophthalmometer, given the simplicity and the inexpensiveness of the instrument. Presentation: Saturday, June 11, 2022 1:00 p.m. - 3:00 p.m., Monday, June 13, 2022 12:30 p.m. - 12:35 p.m. Oxford University Press 2022-11-01 /pmc/articles/PMC9628883/ http://dx.doi.org/10.1210/jendso/bvac150.1779 Text en © The Author(s) 2022. Published by Oxford University Press on behalf of the Endocrine Society. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs licence (https://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial reproduction and distribution of the work, in any medium, provided the original work is not altered or transformed in any way, and that the work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle Thyroid
Traiwanatham, Sirinapa
Sirinvaravong, Sirinart
Plengvidhya, Nattachet
Wongsripuemtet, Jitsupa
Lertlam, Supassa
RF35 | PSAT262 Comparative Study of Hertel Exophthalmometer and Luedde Exophthalmometer Versus Computed Tomography for Measurement of Proptosis in Normal Subjects
title RF35 | PSAT262 Comparative Study of Hertel Exophthalmometer and Luedde Exophthalmometer Versus Computed Tomography for Measurement of Proptosis in Normal Subjects
title_full RF35 | PSAT262 Comparative Study of Hertel Exophthalmometer and Luedde Exophthalmometer Versus Computed Tomography for Measurement of Proptosis in Normal Subjects
title_fullStr RF35 | PSAT262 Comparative Study of Hertel Exophthalmometer and Luedde Exophthalmometer Versus Computed Tomography for Measurement of Proptosis in Normal Subjects
title_full_unstemmed RF35 | PSAT262 Comparative Study of Hertel Exophthalmometer and Luedde Exophthalmometer Versus Computed Tomography for Measurement of Proptosis in Normal Subjects
title_short RF35 | PSAT262 Comparative Study of Hertel Exophthalmometer and Luedde Exophthalmometer Versus Computed Tomography for Measurement of Proptosis in Normal Subjects
title_sort rf35 | psat262 comparative study of hertel exophthalmometer and luedde exophthalmometer versus computed tomography for measurement of proptosis in normal subjects
topic Thyroid
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9628883/
http://dx.doi.org/10.1210/jendso/bvac150.1779
work_keys_str_mv AT traiwanathamsirinapa rf35psat262comparativestudyofhertelexophthalmometerandlueddeexophthalmometerversuscomputedtomographyformeasurementofproptosisinnormalsubjects
AT sirinvaravongsirinart rf35psat262comparativestudyofhertelexophthalmometerandlueddeexophthalmometerversuscomputedtomographyformeasurementofproptosisinnormalsubjects
AT plengvidhyanattachet rf35psat262comparativestudyofhertelexophthalmometerandlueddeexophthalmometerversuscomputedtomographyformeasurementofproptosisinnormalsubjects
AT wongsripuemtetjitsupa rf35psat262comparativestudyofhertelexophthalmometerandlueddeexophthalmometerversuscomputedtomographyformeasurementofproptosisinnormalsubjects
AT lertlamsupassa rf35psat262comparativestudyofhertelexophthalmometerandlueddeexophthalmometerversuscomputedtomographyformeasurementofproptosisinnormalsubjects