Validation of Uright model TD 3127AT wireless-portable ambulatory blood pressure monitoring device with timer trigger modification, standard cuff size, in normotensive and mild hypertensive patients of Thailand registry (Thai valid ambulatory blood pressure monitoring)

Nowadays, automated blood pressure (BP) monitoring devices are commonly used by patients as a part of standard medical care for hypertension. The timer trigger was modified into a wireless automated home BP monitoring (HBPM) device to expand its potential use as ambulatory BP monitoring. However, th...

Descripción completa

Detalles Bibliográficos
Autores principales: Laosanguanek, Wanchat, Kaewkanha, Ponthakorn, Khowinthaseth, Sirhavich, Sirisuth, Sirawee, Phanthong, Tanattida, Chokesuwattanaskul, Ronpichai, Sangwatanaroj, Somkiat
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9632940/
https://www.ncbi.nlm.nih.gov/pubmed/36094361
http://dx.doi.org/10.1097/MBP.0000000000000620
_version_ 1784824150731259904
author Laosanguanek, Wanchat
Kaewkanha, Ponthakorn
Khowinthaseth, Sirhavich
Sirisuth, Sirawee
Phanthong, Tanattida
Chokesuwattanaskul, Ronpichai
Sangwatanaroj, Somkiat
author_facet Laosanguanek, Wanchat
Kaewkanha, Ponthakorn
Khowinthaseth, Sirhavich
Sirisuth, Sirawee
Phanthong, Tanattida
Chokesuwattanaskul, Ronpichai
Sangwatanaroj, Somkiat
author_sort Laosanguanek, Wanchat
collection PubMed
description Nowadays, automated blood pressure (BP) monitoring devices are commonly used by patients as a part of standard medical care for hypertension. The timer trigger was modified into a wireless automated home BP monitoring (HBPM) device to expand its potential use as ambulatory BP monitoring. However, the BP measurement accuracy in this modified device remains unknown. OBJECTIVE: We aimed to assess the accuracy of Uright model TD 3127AT, which is an automated HBPM device with a timer trigger modification, following an International Organization for Standardization (ISO) 81060-2:2018 guidelines in the Thai population. METHODS: This cross-sectional study included normotensive and hypertensive Thai participants following the ISO 81060-2:2018 guidelines from August 2021 to February 2022. This study aimed to compare the BP readings from an automated sphygmomanometer, Uright model TD 3127AT, TaiDoc Technology Corporation, with a timer trigger to a standard manual BP measurement. RESULT: BPs were measured in 85 participants with a mean age ± SD of 38.39 ± 13.91 years, and 69% were females. The mean SBP ± SD (range) was 117.46 ± 18.63 (84–176) mmHg and the mean DBP ± SD (range) was 74.84 ± 10.70 (42–108) mmHg. The mean BP difference between observers and devices was 0.66 ± 6.81 mmHg for SBP and −0.96 ± 6.33 mmHg for DBP. The SD of the averaged pair determination per individual was ±4.45 mmHg for SBP and ±3.46 mmHg for DBP. The accuracy of the timer-triggered device was found to be acceptable when evaluated according to the ISO 81060-2: 2018 guidelines. CONCLUSION: An automated sphygmomanometer, Uright model TD 3127AT, TaiDoc Technology Corporation, with timer trigger modification passed the ISO 81060-2:2018 guidelines.
format Online
Article
Text
id pubmed-9632940
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Lippincott Williams & Wilkins
record_format MEDLINE/PubMed
spelling pubmed-96329402022-11-04 Validation of Uright model TD 3127AT wireless-portable ambulatory blood pressure monitoring device with timer trigger modification, standard cuff size, in normotensive and mild hypertensive patients of Thailand registry (Thai valid ambulatory blood pressure monitoring) Laosanguanek, Wanchat Kaewkanha, Ponthakorn Khowinthaseth, Sirhavich Sirisuth, Sirawee Phanthong, Tanattida Chokesuwattanaskul, Ronpichai Sangwatanaroj, Somkiat Blood Press Monit Devices and Technology Nowadays, automated blood pressure (BP) monitoring devices are commonly used by patients as a part of standard medical care for hypertension. The timer trigger was modified into a wireless automated home BP monitoring (HBPM) device to expand its potential use as ambulatory BP monitoring. However, the BP measurement accuracy in this modified device remains unknown. OBJECTIVE: We aimed to assess the accuracy of Uright model TD 3127AT, which is an automated HBPM device with a timer trigger modification, following an International Organization for Standardization (ISO) 81060-2:2018 guidelines in the Thai population. METHODS: This cross-sectional study included normotensive and hypertensive Thai participants following the ISO 81060-2:2018 guidelines from August 2021 to February 2022. This study aimed to compare the BP readings from an automated sphygmomanometer, Uright model TD 3127AT, TaiDoc Technology Corporation, with a timer trigger to a standard manual BP measurement. RESULT: BPs were measured in 85 participants with a mean age ± SD of 38.39 ± 13.91 years, and 69% were females. The mean SBP ± SD (range) was 117.46 ± 18.63 (84–176) mmHg and the mean DBP ± SD (range) was 74.84 ± 10.70 (42–108) mmHg. The mean BP difference between observers and devices was 0.66 ± 6.81 mmHg for SBP and −0.96 ± 6.33 mmHg for DBP. The SD of the averaged pair determination per individual was ±4.45 mmHg for SBP and ±3.46 mmHg for DBP. The accuracy of the timer-triggered device was found to be acceptable when evaluated according to the ISO 81060-2: 2018 guidelines. CONCLUSION: An automated sphygmomanometer, Uright model TD 3127AT, TaiDoc Technology Corporation, with timer trigger modification passed the ISO 81060-2:2018 guidelines. Lippincott Williams & Wilkins 2022-09-09 2022-12 /pmc/articles/PMC9632940/ /pubmed/36094361 http://dx.doi.org/10.1097/MBP.0000000000000620 Text en Copyright © 2022 The Author(s). Published by Wolters Kluwer Health, Inc. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND) (https://creativecommons.org/licenses/by-nc-nd/4.0/) , where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal.
spellingShingle Devices and Technology
Laosanguanek, Wanchat
Kaewkanha, Ponthakorn
Khowinthaseth, Sirhavich
Sirisuth, Sirawee
Phanthong, Tanattida
Chokesuwattanaskul, Ronpichai
Sangwatanaroj, Somkiat
Validation of Uright model TD 3127AT wireless-portable ambulatory blood pressure monitoring device with timer trigger modification, standard cuff size, in normotensive and mild hypertensive patients of Thailand registry (Thai valid ambulatory blood pressure monitoring)
title Validation of Uright model TD 3127AT wireless-portable ambulatory blood pressure monitoring device with timer trigger modification, standard cuff size, in normotensive and mild hypertensive patients of Thailand registry (Thai valid ambulatory blood pressure monitoring)
title_full Validation of Uright model TD 3127AT wireless-portable ambulatory blood pressure monitoring device with timer trigger modification, standard cuff size, in normotensive and mild hypertensive patients of Thailand registry (Thai valid ambulatory blood pressure monitoring)
title_fullStr Validation of Uright model TD 3127AT wireless-portable ambulatory blood pressure monitoring device with timer trigger modification, standard cuff size, in normotensive and mild hypertensive patients of Thailand registry (Thai valid ambulatory blood pressure monitoring)
title_full_unstemmed Validation of Uright model TD 3127AT wireless-portable ambulatory blood pressure monitoring device with timer trigger modification, standard cuff size, in normotensive and mild hypertensive patients of Thailand registry (Thai valid ambulatory blood pressure monitoring)
title_short Validation of Uright model TD 3127AT wireless-portable ambulatory blood pressure monitoring device with timer trigger modification, standard cuff size, in normotensive and mild hypertensive patients of Thailand registry (Thai valid ambulatory blood pressure monitoring)
title_sort validation of uright model td 3127at wireless-portable ambulatory blood pressure monitoring device with timer trigger modification, standard cuff size, in normotensive and mild hypertensive patients of thailand registry (thai valid ambulatory blood pressure monitoring)
topic Devices and Technology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9632940/
https://www.ncbi.nlm.nih.gov/pubmed/36094361
http://dx.doi.org/10.1097/MBP.0000000000000620
work_keys_str_mv AT laosanguanekwanchat validationofurightmodeltd3127atwirelessportableambulatorybloodpressuremonitoringdevicewithtimertriggermodificationstandardcuffsizeinnormotensiveandmildhypertensivepatientsofthailandregistrythaivalidambulatorybloodpressuremonitoring
AT kaewkanhaponthakorn validationofurightmodeltd3127atwirelessportableambulatorybloodpressuremonitoringdevicewithtimertriggermodificationstandardcuffsizeinnormotensiveandmildhypertensivepatientsofthailandregistrythaivalidambulatorybloodpressuremonitoring
AT khowinthasethsirhavich validationofurightmodeltd3127atwirelessportableambulatorybloodpressuremonitoringdevicewithtimertriggermodificationstandardcuffsizeinnormotensiveandmildhypertensivepatientsofthailandregistrythaivalidambulatorybloodpressuremonitoring
AT sirisuthsirawee validationofurightmodeltd3127atwirelessportableambulatorybloodpressuremonitoringdevicewithtimertriggermodificationstandardcuffsizeinnormotensiveandmildhypertensivepatientsofthailandregistrythaivalidambulatorybloodpressuremonitoring
AT phanthongtanattida validationofurightmodeltd3127atwirelessportableambulatorybloodpressuremonitoringdevicewithtimertriggermodificationstandardcuffsizeinnormotensiveandmildhypertensivepatientsofthailandregistrythaivalidambulatorybloodpressuremonitoring
AT chokesuwattanaskulronpichai validationofurightmodeltd3127atwirelessportableambulatorybloodpressuremonitoringdevicewithtimertriggermodificationstandardcuffsizeinnormotensiveandmildhypertensivepatientsofthailandregistrythaivalidambulatorybloodpressuremonitoring
AT sangwatanarojsomkiat validationofurightmodeltd3127atwirelessportableambulatorybloodpressuremonitoringdevicewithtimertriggermodificationstandardcuffsizeinnormotensiveandmildhypertensivepatientsofthailandregistrythaivalidambulatorybloodpressuremonitoring