Artificial Intelligence Detection of Diabetic Retinopathy: Subgroup Comparison of the EyeArt System with Ophthalmologists’ Dilated Examinations

OBJECTIVE: To compare general ophthalmologists, retina specialists, and the EyeArt Artificial Intelligence (AI) system to the clinical reference standard for detecting more than mild diabetic retinopathy (mtmDR). DESIGN: Prospective, pivotal, multicenter trial conducted from April 2017 to May 2018....

Descripción completa

Detalles Bibliográficos
Autores principales: Lim, Jennifer Irene, Regillo, Carl D., Sadda, SriniVas R., Ipp, Eli, Bhaskaranand, Malavika, Ramachandra, Chaithanya, Solanki, Kaushal
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9636573/
https://www.ncbi.nlm.nih.gov/pubmed/36345378
http://dx.doi.org/10.1016/j.xops.2022.100228
_version_ 1784824974582743040
author Lim, Jennifer Irene
Regillo, Carl D.
Sadda, SriniVas R.
Ipp, Eli
Bhaskaranand, Malavika
Ramachandra, Chaithanya
Solanki, Kaushal
author_facet Lim, Jennifer Irene
Regillo, Carl D.
Sadda, SriniVas R.
Ipp, Eli
Bhaskaranand, Malavika
Ramachandra, Chaithanya
Solanki, Kaushal
author_sort Lim, Jennifer Irene
collection PubMed
description OBJECTIVE: To compare general ophthalmologists, retina specialists, and the EyeArt Artificial Intelligence (AI) system to the clinical reference standard for detecting more than mild diabetic retinopathy (mtmDR). DESIGN: Prospective, pivotal, multicenter trial conducted from April 2017 to May 2018. PARTICIPANTS: Participants were aged ≥ 18 years who had diabetes mellitus and underwent dilated ophthalmoscopy. A total of 521 of 893 participants met these criteria and completed the study protocol. TESTING: Participants underwent 2-field fundus photography (macula centered, disc centered) for the EyeArt system, dilated ophthalmoscopy, and 4-widefield stereoscopic dilated fundus photography for reference standard grading. MAIN OUTCOME MEASURES: For mtmDR detection, sensitivity and specificity of EyeArt gradings of 2-field, fundus photographs and ophthalmoscopy grading versus a rigorous clinical reference standard comprising Reading Center grading of 4-widefield stereoscopic dilated fundus photographs using the ETDRS severity scale. The AI system provided automatic eye-level results regarding mtmDR. RESULTS: Overall, 521 participants (999 eyes) at 10 centers underwent dilated ophthalmoscopy: 406 by nonretina and 115 by retina specialists. Reading Center graded 207 positive and 792 eyes negative for mtmDR. Of these 999 eyes, 26 eyes were ungradable by the EyeArt system, leaving 973 eyes with both EyeArt and Reading Center gradings. Retina specialists correctly identified 22 of 37 eyes as positive (sensitivity 59.5%) and 182 of 184 eyes as negative (specificity 98.9%) for mtmDR versus the EyeArt AI system that identified 36 of 37 as positive (sensitivity 97%) and 162 of 184 eyes as negative (specificity of 88%) for mtmDR. General ophthalmologists correctly identified 35 of 170 eyes as positive (sensitivity 20.6%) and 607 of 608 eyes as negative (specificity 99.8%) for mtmDR compared with the EyeArt AI system that identified 164 of 170 as positive (sensitivity 96.5%) and 525 of 608 eyes as negative (specificity 86%) for mtmDR. CONCLUSIONS: The AI system had a higher sensitivity for detecting mtmDR than either general ophthalmologists or retina specialists compared with the clinical reference standard. It can potentially serve as a low-cost point-of-care diabetic retinopathy detection tool and help address the diabetic eye screening burden.
format Online
Article
Text
id pubmed-9636573
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-96365732022-11-06 Artificial Intelligence Detection of Diabetic Retinopathy: Subgroup Comparison of the EyeArt System with Ophthalmologists’ Dilated Examinations Lim, Jennifer Irene Regillo, Carl D. Sadda, SriniVas R. Ipp, Eli Bhaskaranand, Malavika Ramachandra, Chaithanya Solanki, Kaushal Ophthalmol Sci Artificial Intelligence and Big Data OBJECTIVE: To compare general ophthalmologists, retina specialists, and the EyeArt Artificial Intelligence (AI) system to the clinical reference standard for detecting more than mild diabetic retinopathy (mtmDR). DESIGN: Prospective, pivotal, multicenter trial conducted from April 2017 to May 2018. PARTICIPANTS: Participants were aged ≥ 18 years who had diabetes mellitus and underwent dilated ophthalmoscopy. A total of 521 of 893 participants met these criteria and completed the study protocol. TESTING: Participants underwent 2-field fundus photography (macula centered, disc centered) for the EyeArt system, dilated ophthalmoscopy, and 4-widefield stereoscopic dilated fundus photography for reference standard grading. MAIN OUTCOME MEASURES: For mtmDR detection, sensitivity and specificity of EyeArt gradings of 2-field, fundus photographs and ophthalmoscopy grading versus a rigorous clinical reference standard comprising Reading Center grading of 4-widefield stereoscopic dilated fundus photographs using the ETDRS severity scale. The AI system provided automatic eye-level results regarding mtmDR. RESULTS: Overall, 521 participants (999 eyes) at 10 centers underwent dilated ophthalmoscopy: 406 by nonretina and 115 by retina specialists. Reading Center graded 207 positive and 792 eyes negative for mtmDR. Of these 999 eyes, 26 eyes were ungradable by the EyeArt system, leaving 973 eyes with both EyeArt and Reading Center gradings. Retina specialists correctly identified 22 of 37 eyes as positive (sensitivity 59.5%) and 182 of 184 eyes as negative (specificity 98.9%) for mtmDR versus the EyeArt AI system that identified 36 of 37 as positive (sensitivity 97%) and 162 of 184 eyes as negative (specificity of 88%) for mtmDR. General ophthalmologists correctly identified 35 of 170 eyes as positive (sensitivity 20.6%) and 607 of 608 eyes as negative (specificity 99.8%) for mtmDR compared with the EyeArt AI system that identified 164 of 170 as positive (sensitivity 96.5%) and 525 of 608 eyes as negative (specificity 86%) for mtmDR. CONCLUSIONS: The AI system had a higher sensitivity for detecting mtmDR than either general ophthalmologists or retina specialists compared with the clinical reference standard. It can potentially serve as a low-cost point-of-care diabetic retinopathy detection tool and help address the diabetic eye screening burden. Elsevier 2022-09-30 /pmc/articles/PMC9636573/ /pubmed/36345378 http://dx.doi.org/10.1016/j.xops.2022.100228 Text en © 2022 by the American Academy of Ophthalmology. https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Artificial Intelligence and Big Data
Lim, Jennifer Irene
Regillo, Carl D.
Sadda, SriniVas R.
Ipp, Eli
Bhaskaranand, Malavika
Ramachandra, Chaithanya
Solanki, Kaushal
Artificial Intelligence Detection of Diabetic Retinopathy: Subgroup Comparison of the EyeArt System with Ophthalmologists’ Dilated Examinations
title Artificial Intelligence Detection of Diabetic Retinopathy: Subgroup Comparison of the EyeArt System with Ophthalmologists’ Dilated Examinations
title_full Artificial Intelligence Detection of Diabetic Retinopathy: Subgroup Comparison of the EyeArt System with Ophthalmologists’ Dilated Examinations
title_fullStr Artificial Intelligence Detection of Diabetic Retinopathy: Subgroup Comparison of the EyeArt System with Ophthalmologists’ Dilated Examinations
title_full_unstemmed Artificial Intelligence Detection of Diabetic Retinopathy: Subgroup Comparison of the EyeArt System with Ophthalmologists’ Dilated Examinations
title_short Artificial Intelligence Detection of Diabetic Retinopathy: Subgroup Comparison of the EyeArt System with Ophthalmologists’ Dilated Examinations
title_sort artificial intelligence detection of diabetic retinopathy: subgroup comparison of the eyeart system with ophthalmologists’ dilated examinations
topic Artificial Intelligence and Big Data
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9636573/
https://www.ncbi.nlm.nih.gov/pubmed/36345378
http://dx.doi.org/10.1016/j.xops.2022.100228
work_keys_str_mv AT limjenniferirene artificialintelligencedetectionofdiabeticretinopathysubgroupcomparisonoftheeyeartsystemwithophthalmologistsdilatedexaminations
AT regillocarld artificialintelligencedetectionofdiabeticretinopathysubgroupcomparisonoftheeyeartsystemwithophthalmologistsdilatedexaminations
AT saddasrinivasr artificialintelligencedetectionofdiabeticretinopathysubgroupcomparisonoftheeyeartsystemwithophthalmologistsdilatedexaminations
AT ippeli artificialintelligencedetectionofdiabeticretinopathysubgroupcomparisonoftheeyeartsystemwithophthalmologistsdilatedexaminations
AT bhaskaranandmalavika artificialintelligencedetectionofdiabeticretinopathysubgroupcomparisonoftheeyeartsystemwithophthalmologistsdilatedexaminations
AT ramachandrachaithanya artificialintelligencedetectionofdiabeticretinopathysubgroupcomparisonoftheeyeartsystemwithophthalmologistsdilatedexaminations
AT solankikaushal artificialintelligencedetectionofdiabeticretinopathysubgroupcomparisonoftheeyeartsystemwithophthalmologistsdilatedexaminations