Cargando…

Catastrophic failure of a titanium locking plate in a proximal humeral fracture: case report and literature review

BACKGROUND: Angular stable locking plates have shown good clinical results in treating proximal humeral fractures, but complications are not uncommon. This study reported a rare case of catastrophic failure of a titanium locking plate. A retrieval analysis of the implants was performed using an opti...

Descripción completa

Detalles Bibliográficos
Autores principales: Lian, Yan-Shiang, Huang, Chang-Hung, Chuang, Min-Yao
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9636662/
https://www.ncbi.nlm.nih.gov/pubmed/36333815
http://dx.doi.org/10.1186/s12891-022-05931-4
_version_ 1784824997936627712
author Lian, Yan-Shiang
Huang, Chang-Hung
Chuang, Min-Yao
author_facet Lian, Yan-Shiang
Huang, Chang-Hung
Chuang, Min-Yao
author_sort Lian, Yan-Shiang
collection PubMed
description BACKGROUND: Angular stable locking plates have shown good clinical results in treating proximal humeral fractures, but complications are not uncommon. This study reported a rare case of catastrophic failure of a titanium locking plate. A retrieval analysis of the implants was performed using an optic microscope and a scanning electron microscope. CASE PRESENTATION: A 69-year-old male reported a right proximal humeral fracture at the surgical neck and was treated by open reduction and internal fixation with a locking plate system. Ninety-six days after surgery, the patient came to clinic for acute local pain over the shoulder without any trauma. The radiographs showed a complete breakage of the implant accompanying displaced fracture. Revision surgery was performed to restabilize the fracture with a longer locking plate. The follow-up radiographs at 9 months showed complete union of the bone fracture. CONCLUSIONS: From the retrieval analysis, repetitive torsion loads on the vulnerable area of the implant are assumed to cause this catastrophic event. It is recommended that adequate activity restriction, such as reaching, be undertaken to avoid this rare complication. Current study also provides contributive information for the modification of plate design and pre-operative planning for device configuration to improve the success rate of locking plate fixation.
format Online
Article
Text
id pubmed-9636662
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-96366622022-11-06 Catastrophic failure of a titanium locking plate in a proximal humeral fracture: case report and literature review Lian, Yan-Shiang Huang, Chang-Hung Chuang, Min-Yao BMC Musculoskelet Disord Case Report BACKGROUND: Angular stable locking plates have shown good clinical results in treating proximal humeral fractures, but complications are not uncommon. This study reported a rare case of catastrophic failure of a titanium locking plate. A retrieval analysis of the implants was performed using an optic microscope and a scanning electron microscope. CASE PRESENTATION: A 69-year-old male reported a right proximal humeral fracture at the surgical neck and was treated by open reduction and internal fixation with a locking plate system. Ninety-six days after surgery, the patient came to clinic for acute local pain over the shoulder without any trauma. The radiographs showed a complete breakage of the implant accompanying displaced fracture. Revision surgery was performed to restabilize the fracture with a longer locking plate. The follow-up radiographs at 9 months showed complete union of the bone fracture. CONCLUSIONS: From the retrieval analysis, repetitive torsion loads on the vulnerable area of the implant are assumed to cause this catastrophic event. It is recommended that adequate activity restriction, such as reaching, be undertaken to avoid this rare complication. Current study also provides contributive information for the modification of plate design and pre-operative planning for device configuration to improve the success rate of locking plate fixation. BioMed Central 2022-11-05 /pmc/articles/PMC9636662/ /pubmed/36333815 http://dx.doi.org/10.1186/s12891-022-05931-4 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Case Report
Lian, Yan-Shiang
Huang, Chang-Hung
Chuang, Min-Yao
Catastrophic failure of a titanium locking plate in a proximal humeral fracture: case report and literature review
title Catastrophic failure of a titanium locking plate in a proximal humeral fracture: case report and literature review
title_full Catastrophic failure of a titanium locking plate in a proximal humeral fracture: case report and literature review
title_fullStr Catastrophic failure of a titanium locking plate in a proximal humeral fracture: case report and literature review
title_full_unstemmed Catastrophic failure of a titanium locking plate in a proximal humeral fracture: case report and literature review
title_short Catastrophic failure of a titanium locking plate in a proximal humeral fracture: case report and literature review
title_sort catastrophic failure of a titanium locking plate in a proximal humeral fracture: case report and literature review
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9636662/
https://www.ncbi.nlm.nih.gov/pubmed/36333815
http://dx.doi.org/10.1186/s12891-022-05931-4
work_keys_str_mv AT lianyanshiang catastrophicfailureofatitaniumlockingplateinaproximalhumeralfracturecasereportandliteraturereview
AT huangchanghung catastrophicfailureofatitaniumlockingplateinaproximalhumeralfracturecasereportandliteraturereview
AT chuangminyao catastrophicfailureofatitaniumlockingplateinaproximalhumeralfracturecasereportandliteraturereview