Noninvasive Technique to Monitor the Pressure under a Cast: A Mobile Application-Friendly Bluetooth Pressure Sensor

AIM: The purpose of this study was to design a sensor that could monitor the skin-cast contact surface pressure (SCCSP) of a limb under a cast and inform the user via a mobile application when the pressure increases. METHODS: In this experimental study, an infant sphygmomanometer cuff was initially...

Descripción completa

Detalles Bibliográficos
Autores principales: Mahirogullari, Mahir, Surucu, Serkan, Cerci, Mehmet Halis, Aydin, Mahmud, Kayikli, Abdulkadir, Gunduz, Oguzhan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9640232/
https://www.ncbi.nlm.nih.gov/pubmed/36406479
http://dx.doi.org/10.1155/2022/9093612
_version_ 1784825805477511168
author Mahirogullari, Mahir
Surucu, Serkan
Cerci, Mehmet Halis
Aydin, Mahmud
Kayikli, Abdulkadir
Gunduz, Oguzhan
author_facet Mahirogullari, Mahir
Surucu, Serkan
Cerci, Mehmet Halis
Aydin, Mahmud
Kayikli, Abdulkadir
Gunduz, Oguzhan
author_sort Mahirogullari, Mahir
collection PubMed
description AIM: The purpose of this study was to design a sensor that could monitor the skin-cast contact surface pressure (SCCSP) of a limb under a cast and inform the user via a mobile application when the pressure increases. METHODS: In this experimental study, an infant sphygmomanometer cuff was initially placed on the forearm of 10 volunteers. A pressure sensor with a Bluetooth chip was then placed on the volar aspect of the forearm. Short arm plaster was applied with synthetic cast material. The SCCSP under the plaster was measured by the sensor and the measured values were transmitted to a mobile application via a Bluetooth chip. The mobile application processed the data from the chip and converted it to mmHg. RESULTS: Intracompartmental pressure (ICP) values were categorized as 0, 10, 20, 30, 40, 50, 60, and 75 mmHg. The highest SCCSP was 75 mmHg CP, while the lowest was 0 mmHg CP. The correlation coefficient of the mean pressure values was 0.993 (p ≤ 0.001) (SD 0.002, range 0.989–0.997), and there was a significant relationship between ICP and SCCSP values (p ≤ 0.05). CONCLUSION: We can monitor SCCSP, detect limb swelling, and notify the user via a mobile application by using Bluetooth pressure sensors.
format Online
Article
Text
id pubmed-9640232
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-96402322022-11-17 Noninvasive Technique to Monitor the Pressure under a Cast: A Mobile Application-Friendly Bluetooth Pressure Sensor Mahirogullari, Mahir Surucu, Serkan Cerci, Mehmet Halis Aydin, Mahmud Kayikli, Abdulkadir Gunduz, Oguzhan Int J Clin Pract Research Article AIM: The purpose of this study was to design a sensor that could monitor the skin-cast contact surface pressure (SCCSP) of a limb under a cast and inform the user via a mobile application when the pressure increases. METHODS: In this experimental study, an infant sphygmomanometer cuff was initially placed on the forearm of 10 volunteers. A pressure sensor with a Bluetooth chip was then placed on the volar aspect of the forearm. Short arm plaster was applied with synthetic cast material. The SCCSP under the plaster was measured by the sensor and the measured values were transmitted to a mobile application via a Bluetooth chip. The mobile application processed the data from the chip and converted it to mmHg. RESULTS: Intracompartmental pressure (ICP) values were categorized as 0, 10, 20, 30, 40, 50, 60, and 75 mmHg. The highest SCCSP was 75 mmHg CP, while the lowest was 0 mmHg CP. The correlation coefficient of the mean pressure values was 0.993 (p ≤ 0.001) (SD 0.002, range 0.989–0.997), and there was a significant relationship between ICP and SCCSP values (p ≤ 0.05). CONCLUSION: We can monitor SCCSP, detect limb swelling, and notify the user via a mobile application by using Bluetooth pressure sensors. Hindawi 2022-10-31 /pmc/articles/PMC9640232/ /pubmed/36406479 http://dx.doi.org/10.1155/2022/9093612 Text en Copyright © 2022 Mahir Mahirogullari et al. https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Mahirogullari, Mahir
Surucu, Serkan
Cerci, Mehmet Halis
Aydin, Mahmud
Kayikli, Abdulkadir
Gunduz, Oguzhan
Noninvasive Technique to Monitor the Pressure under a Cast: A Mobile Application-Friendly Bluetooth Pressure Sensor
title Noninvasive Technique to Monitor the Pressure under a Cast: A Mobile Application-Friendly Bluetooth Pressure Sensor
title_full Noninvasive Technique to Monitor the Pressure under a Cast: A Mobile Application-Friendly Bluetooth Pressure Sensor
title_fullStr Noninvasive Technique to Monitor the Pressure under a Cast: A Mobile Application-Friendly Bluetooth Pressure Sensor
title_full_unstemmed Noninvasive Technique to Monitor the Pressure under a Cast: A Mobile Application-Friendly Bluetooth Pressure Sensor
title_short Noninvasive Technique to Monitor the Pressure under a Cast: A Mobile Application-Friendly Bluetooth Pressure Sensor
title_sort noninvasive technique to monitor the pressure under a cast: a mobile application-friendly bluetooth pressure sensor
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9640232/
https://www.ncbi.nlm.nih.gov/pubmed/36406479
http://dx.doi.org/10.1155/2022/9093612
work_keys_str_mv AT mahirogullarimahir noninvasivetechniquetomonitorthepressureunderacastamobileapplicationfriendlybluetoothpressuresensor
AT surucuserkan noninvasivetechniquetomonitorthepressureunderacastamobileapplicationfriendlybluetoothpressuresensor
AT cercimehmethalis noninvasivetechniquetomonitorthepressureunderacastamobileapplicationfriendlybluetoothpressuresensor
AT aydinmahmud noninvasivetechniquetomonitorthepressureunderacastamobileapplicationfriendlybluetoothpressuresensor
AT kayikliabdulkadir noninvasivetechniquetomonitorthepressureunderacastamobileapplicationfriendlybluetoothpressuresensor
AT gunduzoguzhan noninvasivetechniquetomonitorthepressureunderacastamobileapplicationfriendlybluetoothpressuresensor