Cargando…
3D planning and patient specific instrumentation for intraarticular corrective osteotomy of trapeziometacarpal-, metacarpal and finger joints
BACKGROUND: Intra-articular malunions of the finger can lead to deformity and loss of function and can be treated with intra-articular corrective osteotomies. The aim of this study was to evaluate radiographic joint congruency, feasibility and functional outcome of three-dimensional (3D) printed pat...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9641882/ https://www.ncbi.nlm.nih.gov/pubmed/36348352 http://dx.doi.org/10.1186/s12891-022-05946-x |
Sumario: | BACKGROUND: Intra-articular malunions of the finger can lead to deformity and loss of function and can be treated with intra-articular corrective osteotomies. The aim of this study was to evaluate radiographic joint congruency, feasibility and functional outcome of three-dimensional (3D) printed patient-specific instrumentation (PSI) for corrective osteotomies at the trapeziometacarpal and finger joints. METHODS: Computer-tomography (CT) scans were acquired preoperatively for standard 3D planning, which was followed by calculation of cutting planes and the design of individualized bone surface contact drilling, sawing and reposition guides. Follow-up CT scans and clinical examinations (range of motion, grip strength) were performed. Postoperative complications were documented and patient-reported outcome measurements were assessed (Single Assessment Numeric Evaluation (SANE) score, brief Michigan Hand Questionnaire (MHQ)). RESULTS: Ten patients (mean age 28.4 ± 12.8,range 13.8–51.3) years) were included with a mean follow-up of 21 ± 18 (3–59) months including seven osteotomies at the trapeziometacarpal or metacarpophalangeal joints and three at the proximal interphalangeal joint (PIP). All radiographic follow-up examinations showed the planned correction with good joint congruency and regular osseous consolidation. At the latest follow-up, the range of motion (ROM) increased and the average grip strength recovered to the level of the contralateral side. No postoperative complication was detected. The mean SANE score improved from 44 ± 23 (0–70) to 82 ± 12 (60–90) after a mean of 72 ± 20 (44–114) months. The mean postoperative brief MHQ was 92 ± 8 (71–98). CONCLUSION: The use of 3D PSI in treating intra-articular malunions at the trapeziometacarpal and finger joints restored articular congruency accurately. ROM and grip strength improved postoperatively comparable to the healthy contralateral side and patient-reported outcome measures improved after medium-term follow-up. |
---|