Cargando…

Reproducibility and repeatability of a semi-automated pipeline to quantify trapeziometacarpal joint angles using dynamic computed tomography

BACKGROUND: The trapeziometacarpal (TMC) joint is a mechanically complex joint and is commonly affected by musculoskeletal diseases such as osteoarthritis. Quantifying in vivo TMC joint biomechanics, such as joint angles, with traditional reflective marker-based methods can be difficult due to the j...

Descripción completa

Detalles Bibliográficos
Autores principales: Kuczynski, Michael T., Wang, Kendra, Tse, Justin J., Bugajski, Tomasz, Manske, Sarah L.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9644596/
https://www.ncbi.nlm.nih.gov/pubmed/36348287
http://dx.doi.org/10.1186/s12880-022-00922-2
_version_ 1784826776023728128
author Kuczynski, Michael T.
Wang, Kendra
Tse, Justin J.
Bugajski, Tomasz
Manske, Sarah L.
author_facet Kuczynski, Michael T.
Wang, Kendra
Tse, Justin J.
Bugajski, Tomasz
Manske, Sarah L.
author_sort Kuczynski, Michael T.
collection PubMed
description BACKGROUND: The trapeziometacarpal (TMC) joint is a mechanically complex joint and is commonly affected by musculoskeletal diseases such as osteoarthritis. Quantifying in vivo TMC joint biomechanics, such as joint angles, with traditional reflective marker-based methods can be difficult due to the joint’s location in the hand. Dynamic computed tomography (CT) can facilitate the quantification of TMC joint motion by continuously capturing three-dimensional volumes over time. However, post-processing of dynamic CT datasets can be time intensive and automated methods are needed to reduce processing times to allow for application to larger clinical studies. The purpose of this work is to introduce a fast, semi-automated pipeline to quantify joint angles from dynamic CT scans of the TMC joint and evaluate the associated error in joint angle and translation computation by means of a reproducibility and repeatability study. METHODS: Ten cadaveric hands were scanned with dynamic CT using a passive motion device to move thumbs in a radial abduction–adduction motion. Static CT scans and high-resolution peripheral quantitative CT scans were also acquired to generate high-resolution bone meshes. Abduction–adduction, flexion–extension, and axial rotation angles were computed using a joint coordinate system. Reproducibility and repeatability were assessed using intraclass correlation coefficients, Bland–Altman analysis, and root mean square errors. Target registration errors were computed to evaluate errors associated with image registration. RESULTS: We found good repeatability for flexion–extension, abduction–adduction, and axial rotation angles. Reproducibility was moderate for all three angles. Joint translations exhibited greater repeatability than reproducibility. Specimens with greater joint degeneration had lower repeatability and reproducibility. We found that the difference in resulting joint angles and translations were likely due to differences in segment coordinate system definition between multiple raters, rather than due to registration errors. CONCLUSIONS: The proposed semi-automatic processing pipeline was fast, repeatable, and moderately reproducible when quantifying TMC joint angles and translations. This work provides a range of errors for TMC joint angles from dynamic CT scans using manually selected anatomical landmarks.
format Online
Article
Text
id pubmed-9644596
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-96445962022-11-15 Reproducibility and repeatability of a semi-automated pipeline to quantify trapeziometacarpal joint angles using dynamic computed tomography Kuczynski, Michael T. Wang, Kendra Tse, Justin J. Bugajski, Tomasz Manske, Sarah L. BMC Med Imaging Research BACKGROUND: The trapeziometacarpal (TMC) joint is a mechanically complex joint and is commonly affected by musculoskeletal diseases such as osteoarthritis. Quantifying in vivo TMC joint biomechanics, such as joint angles, with traditional reflective marker-based methods can be difficult due to the joint’s location in the hand. Dynamic computed tomography (CT) can facilitate the quantification of TMC joint motion by continuously capturing three-dimensional volumes over time. However, post-processing of dynamic CT datasets can be time intensive and automated methods are needed to reduce processing times to allow for application to larger clinical studies. The purpose of this work is to introduce a fast, semi-automated pipeline to quantify joint angles from dynamic CT scans of the TMC joint and evaluate the associated error in joint angle and translation computation by means of a reproducibility and repeatability study. METHODS: Ten cadaveric hands were scanned with dynamic CT using a passive motion device to move thumbs in a radial abduction–adduction motion. Static CT scans and high-resolution peripheral quantitative CT scans were also acquired to generate high-resolution bone meshes. Abduction–adduction, flexion–extension, and axial rotation angles were computed using a joint coordinate system. Reproducibility and repeatability were assessed using intraclass correlation coefficients, Bland–Altman analysis, and root mean square errors. Target registration errors were computed to evaluate errors associated with image registration. RESULTS: We found good repeatability for flexion–extension, abduction–adduction, and axial rotation angles. Reproducibility was moderate for all three angles. Joint translations exhibited greater repeatability than reproducibility. Specimens with greater joint degeneration had lower repeatability and reproducibility. We found that the difference in resulting joint angles and translations were likely due to differences in segment coordinate system definition between multiple raters, rather than due to registration errors. CONCLUSIONS: The proposed semi-automatic processing pipeline was fast, repeatable, and moderately reproducible when quantifying TMC joint angles and translations. This work provides a range of errors for TMC joint angles from dynamic CT scans using manually selected anatomical landmarks. BioMed Central 2022-11-08 /pmc/articles/PMC9644596/ /pubmed/36348287 http://dx.doi.org/10.1186/s12880-022-00922-2 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Kuczynski, Michael T.
Wang, Kendra
Tse, Justin J.
Bugajski, Tomasz
Manske, Sarah L.
Reproducibility and repeatability of a semi-automated pipeline to quantify trapeziometacarpal joint angles using dynamic computed tomography
title Reproducibility and repeatability of a semi-automated pipeline to quantify trapeziometacarpal joint angles using dynamic computed tomography
title_full Reproducibility and repeatability of a semi-automated pipeline to quantify trapeziometacarpal joint angles using dynamic computed tomography
title_fullStr Reproducibility and repeatability of a semi-automated pipeline to quantify trapeziometacarpal joint angles using dynamic computed tomography
title_full_unstemmed Reproducibility and repeatability of a semi-automated pipeline to quantify trapeziometacarpal joint angles using dynamic computed tomography
title_short Reproducibility and repeatability of a semi-automated pipeline to quantify trapeziometacarpal joint angles using dynamic computed tomography
title_sort reproducibility and repeatability of a semi-automated pipeline to quantify trapeziometacarpal joint angles using dynamic computed tomography
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9644596/
https://www.ncbi.nlm.nih.gov/pubmed/36348287
http://dx.doi.org/10.1186/s12880-022-00922-2
work_keys_str_mv AT kuczynskimichaelt reproducibilityandrepeatabilityofasemiautomatedpipelinetoquantifytrapeziometacarpaljointanglesusingdynamiccomputedtomography
AT wangkendra reproducibilityandrepeatabilityofasemiautomatedpipelinetoquantifytrapeziometacarpaljointanglesusingdynamiccomputedtomography
AT tsejustinj reproducibilityandrepeatabilityofasemiautomatedpipelinetoquantifytrapeziometacarpaljointanglesusingdynamiccomputedtomography
AT bugajskitomasz reproducibilityandrepeatabilityofasemiautomatedpipelinetoquantifytrapeziometacarpaljointanglesusingdynamiccomputedtomography
AT manskesarahl reproducibilityandrepeatabilityofasemiautomatedpipelinetoquantifytrapeziometacarpaljointanglesusingdynamiccomputedtomography