Flash Glucose Monitoring and Diabetes Mellitus Induced by Immune Checkpoint Inhibitors: An Approach to Clinical Practice
OBJECTIVES: The aim of this study is to investigate in depth diabetes mellitus associated with immune checkpoint inhibitors (DM-ICIs) by analysing a case series. We also evaluated the clinical impact of flash glucose monitoring (FGM) systems in the management of this entity. METHODS: We conducted an...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9652069/ https://www.ncbi.nlm.nih.gov/pubmed/36387939 http://dx.doi.org/10.1155/2022/4508633 |
_version_ | 1784828385276461056 |
---|---|
author | Rodríguez de Vera-Gómez, Pablo Piñar-Gutiérrez, Ana Guerrero-Vázquez, Raquel Bellido, Virginia Morales-Portillo, Cristóbal Sancho-Márquez, María Pilar Espejo-García, Pablo Gros-Herguido, Noelia López-Gallardo, Gema Martínez-Brocca, María Asunción Soto-Moreno, Alfonso |
author_facet | Rodríguez de Vera-Gómez, Pablo Piñar-Gutiérrez, Ana Guerrero-Vázquez, Raquel Bellido, Virginia Morales-Portillo, Cristóbal Sancho-Márquez, María Pilar Espejo-García, Pablo Gros-Herguido, Noelia López-Gallardo, Gema Martínez-Brocca, María Asunción Soto-Moreno, Alfonso |
author_sort | Rodríguez de Vera-Gómez, Pablo |
collection | PubMed |
description | OBJECTIVES: The aim of this study is to investigate in depth diabetes mellitus associated with immune checkpoint inhibitors (DM-ICIs) by analysing a case series. We also evaluated the clinical impact of flash glucose monitoring (FGM) systems in the management of this entity. METHODS: We conducted an observational cohort study of DM-ICIs diagnosed in two hospitals in Seville (Spain). Patients with a new diagnosis of diabetes mellitus (DM) or with sudden worsening of preexisting DM after starting treatment with ICIs, with a random 5 hour-postprandial C-peptide value of <0.6 nmol/L and without possibility of subsequent withdrawal of insulin treatment, were included. RESULTS: A total of 7 cases were identified, mostly males (n = 6; 85.7%), with a mean age of 64.9 years. The mean glycated hemoglobin (HbA1c) upon diagnosis was 8.1%, with diabetic ketoacidosis (DKA) observed in 6 cases (85.7%). Subcutaneous flash glucose monitoring (FGM) systems were used in six cases, with a mean follow-up period of 42.7 weeks. During the first 90 days of use, mean average glucose was 167.5 mg/dL, with a coefficient of variation (CV) of 34.6%. The mean time in the range 70-180 mg/dL (TIR) was 59.7%, with a mean time above range (TAR) 181-250 mg/dL of 27.8% and a mean TAR > 250 mg/dL of 10.2%. The mean time below range (TBR) 54-69 mg/dL was 2%, while the mean TBR < 54 mg/dL was 0.3%. The mean glucose management indicator (GMI) was 7.3%. No significant differences were observed in FGM values for the following 90 days of follow-up. A progressive improvement in all parameters of glycaemic control was observed between the first month of FGM use and the sixth month of FGM use. Of note, there was a decrease in mean CV (40.6% to 34.1%, p = 0.25), mean TAR 181-250 (30.3% to 26%, p = 0.49), mean TAR > 250 mg/dL (16.3% to 7.7%, p = 0.09), mean TBR 54-69 mg/dL (5.2% to 2%, p = 0.16), and mean TBR < 54 mg/dL (1.8% to 0.2%, p = 0.31), along with an increase in mean values of TIR 70-180 mg/dL (46.5% to 60.5%, p = 0.09). The lack of statistical significance in the differences observed in the mean FGM values over the follow-up period may be related to the small sample size. CONCLUSION: DM-ICI is recognised by a state of sudden-onset insulinopenia, often associated with DKA. The use of FGM systems may be a valid option for the effective management of DM-ICIs and for the prevention of severe hyperglycaemic and hypoglycaemic episodes in this condition. |
format | Online Article Text |
id | pubmed-9652069 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Hindawi |
record_format | MEDLINE/PubMed |
spelling | pubmed-96520692022-11-15 Flash Glucose Monitoring and Diabetes Mellitus Induced by Immune Checkpoint Inhibitors: An Approach to Clinical Practice Rodríguez de Vera-Gómez, Pablo Piñar-Gutiérrez, Ana Guerrero-Vázquez, Raquel Bellido, Virginia Morales-Portillo, Cristóbal Sancho-Márquez, María Pilar Espejo-García, Pablo Gros-Herguido, Noelia López-Gallardo, Gema Martínez-Brocca, María Asunción Soto-Moreno, Alfonso J Diabetes Res Research Article OBJECTIVES: The aim of this study is to investigate in depth diabetes mellitus associated with immune checkpoint inhibitors (DM-ICIs) by analysing a case series. We also evaluated the clinical impact of flash glucose monitoring (FGM) systems in the management of this entity. METHODS: We conducted an observational cohort study of DM-ICIs diagnosed in two hospitals in Seville (Spain). Patients with a new diagnosis of diabetes mellitus (DM) or with sudden worsening of preexisting DM after starting treatment with ICIs, with a random 5 hour-postprandial C-peptide value of <0.6 nmol/L and without possibility of subsequent withdrawal of insulin treatment, were included. RESULTS: A total of 7 cases were identified, mostly males (n = 6; 85.7%), with a mean age of 64.9 years. The mean glycated hemoglobin (HbA1c) upon diagnosis was 8.1%, with diabetic ketoacidosis (DKA) observed in 6 cases (85.7%). Subcutaneous flash glucose monitoring (FGM) systems were used in six cases, with a mean follow-up period of 42.7 weeks. During the first 90 days of use, mean average glucose was 167.5 mg/dL, with a coefficient of variation (CV) of 34.6%. The mean time in the range 70-180 mg/dL (TIR) was 59.7%, with a mean time above range (TAR) 181-250 mg/dL of 27.8% and a mean TAR > 250 mg/dL of 10.2%. The mean time below range (TBR) 54-69 mg/dL was 2%, while the mean TBR < 54 mg/dL was 0.3%. The mean glucose management indicator (GMI) was 7.3%. No significant differences were observed in FGM values for the following 90 days of follow-up. A progressive improvement in all parameters of glycaemic control was observed between the first month of FGM use and the sixth month of FGM use. Of note, there was a decrease in mean CV (40.6% to 34.1%, p = 0.25), mean TAR 181-250 (30.3% to 26%, p = 0.49), mean TAR > 250 mg/dL (16.3% to 7.7%, p = 0.09), mean TBR 54-69 mg/dL (5.2% to 2%, p = 0.16), and mean TBR < 54 mg/dL (1.8% to 0.2%, p = 0.31), along with an increase in mean values of TIR 70-180 mg/dL (46.5% to 60.5%, p = 0.09). The lack of statistical significance in the differences observed in the mean FGM values over the follow-up period may be related to the small sample size. CONCLUSION: DM-ICI is recognised by a state of sudden-onset insulinopenia, often associated with DKA. The use of FGM systems may be a valid option for the effective management of DM-ICIs and for the prevention of severe hyperglycaemic and hypoglycaemic episodes in this condition. Hindawi 2022-11-04 /pmc/articles/PMC9652069/ /pubmed/36387939 http://dx.doi.org/10.1155/2022/4508633 Text en Copyright © 2022 Pablo Rodríguez de Vera-Gómez et al. https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Rodríguez de Vera-Gómez, Pablo Piñar-Gutiérrez, Ana Guerrero-Vázquez, Raquel Bellido, Virginia Morales-Portillo, Cristóbal Sancho-Márquez, María Pilar Espejo-García, Pablo Gros-Herguido, Noelia López-Gallardo, Gema Martínez-Brocca, María Asunción Soto-Moreno, Alfonso Flash Glucose Monitoring and Diabetes Mellitus Induced by Immune Checkpoint Inhibitors: An Approach to Clinical Practice |
title | Flash Glucose Monitoring and Diabetes Mellitus Induced by Immune Checkpoint Inhibitors: An Approach to Clinical Practice |
title_full | Flash Glucose Monitoring and Diabetes Mellitus Induced by Immune Checkpoint Inhibitors: An Approach to Clinical Practice |
title_fullStr | Flash Glucose Monitoring and Diabetes Mellitus Induced by Immune Checkpoint Inhibitors: An Approach to Clinical Practice |
title_full_unstemmed | Flash Glucose Monitoring and Diabetes Mellitus Induced by Immune Checkpoint Inhibitors: An Approach to Clinical Practice |
title_short | Flash Glucose Monitoring and Diabetes Mellitus Induced by Immune Checkpoint Inhibitors: An Approach to Clinical Practice |
title_sort | flash glucose monitoring and diabetes mellitus induced by immune checkpoint inhibitors: an approach to clinical practice |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9652069/ https://www.ncbi.nlm.nih.gov/pubmed/36387939 http://dx.doi.org/10.1155/2022/4508633 |
work_keys_str_mv | AT rodriguezdeveragomezpablo flashglucosemonitoringanddiabetesmellitusinducedbyimmunecheckpointinhibitorsanapproachtoclinicalpractice AT pinargutierrezana flashglucosemonitoringanddiabetesmellitusinducedbyimmunecheckpointinhibitorsanapproachtoclinicalpractice AT guerrerovazquezraquel flashglucosemonitoringanddiabetesmellitusinducedbyimmunecheckpointinhibitorsanapproachtoclinicalpractice AT bellidovirginia flashglucosemonitoringanddiabetesmellitusinducedbyimmunecheckpointinhibitorsanapproachtoclinicalpractice AT moralesportillocristobal flashglucosemonitoringanddiabetesmellitusinducedbyimmunecheckpointinhibitorsanapproachtoclinicalpractice AT sanchomarquezmariapilar flashglucosemonitoringanddiabetesmellitusinducedbyimmunecheckpointinhibitorsanapproachtoclinicalpractice AT espejogarciapablo flashglucosemonitoringanddiabetesmellitusinducedbyimmunecheckpointinhibitorsanapproachtoclinicalpractice AT grosherguidonoelia flashglucosemonitoringanddiabetesmellitusinducedbyimmunecheckpointinhibitorsanapproachtoclinicalpractice AT lopezgallardogema flashglucosemonitoringanddiabetesmellitusinducedbyimmunecheckpointinhibitorsanapproachtoclinicalpractice AT martinezbroccamariaasuncion flashglucosemonitoringanddiabetesmellitusinducedbyimmunecheckpointinhibitorsanapproachtoclinicalpractice AT sotomorenoalfonso flashglucosemonitoringanddiabetesmellitusinducedbyimmunecheckpointinhibitorsanapproachtoclinicalpractice |