Cargando…

Portable Colorimetric Hydrogel Test Kits and On-Mobile Digital Image Colorimetry for On-Site Determination of Nutrients in Water

Portable colorimetric hydrogel test kits are newly developed for the on-site detection of nitrite, nitrate, and phosphate in water. Griess-doped hydrogel was prepared at the bottom of a 1.5 mL plastic tube for nitrite detection, a nitrate reduction film based on zinc powder was placed on the inner l...

Descripción completa

Detalles Bibliográficos
Autores principales: Wongniramaikul, Worawit, Kleangklao, Bussakorn, Boonkanon, Chanita, Taweekarn, Tarawee, Phatthanawiwat, Kharittha, Sriprom, Wilasinee, Limsakul, Wadcharawadee, Towanlong, Wanchitra, Tipmanee, Danai, Choodum, Aree
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9655143/
https://www.ncbi.nlm.nih.gov/pubmed/36364112
http://dx.doi.org/10.3390/molecules27217287
Descripción
Sumario:Portable colorimetric hydrogel test kits are newly developed for the on-site detection of nitrite, nitrate, and phosphate in water. Griess-doped hydrogel was prepared at the bottom of a 1.5 mL plastic tube for nitrite detection, a nitrate reduction film based on zinc powder was placed on the inner lid of a second 1.5 mL plastic tube for use in conjunction with the Griess-doped hydrogel for nitrate detection, and a molybdenum blue-based reagent was entrapped within a poly(vinyl alcohol) hydrogel matrix placed at the bottom of a third 1.5 mL plastic tube to detect phosphate. These test kits are usable with on-mobile digital image colorimetry (DIC) for the on-site determination of nutrients with good analytical performance. The detection limits were 0.02, 0.04, and 0.14 mg L(−1) for nitrite, nitrate, and phosphate, respectively, with good accuracy (<4.8% relative error) and precision (<1.85% relative standard deviation). These test kits and on-mobile DIC were used for the on-site determination of nutrients in the Pak Bang and Bang Yai canals, the main canals in Phuket, Thailand. The concentrations of nitrite, nitrate, and phosphate were undetectable to 0.60 mg L(−1), undetectable to 2.98 mg L(−1), and undetectable to 0.52 mg L(−1), respectively.