Diferencia de presión arterial entre brazos: concordancia entre 2 métodos automáticos de medición simultánea y reproducibilidad entre visitas

OBJECTIVE: An increased interarm blood pressure difference (IAD) (≥10 mmHg) has been associated with increased cardiovascular morbidity and mortality. There are few studies determining how IAD has to be measured and its reliability between visits. The objectives of our study were twofold. First, to...

Descripción completa

Detalles Bibliográficos
Autores principales: Martínez-Sánchez, Nuria, Palasí, Antonio, Pera, Guillem, Martínez, Luis Miguel, Albaladejo, Raquel, Torán, Pere
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9681639/
https://www.ncbi.nlm.nih.gov/pubmed/36423568
http://dx.doi.org/10.1016/j.aprim.2022.102514
_version_ 1784834659910156288
author Martínez-Sánchez, Nuria
Palasí, Antonio
Pera, Guillem
Martínez, Luis Miguel
Albaladejo, Raquel
Torán, Pere
author_facet Martínez-Sánchez, Nuria
Palasí, Antonio
Pera, Guillem
Martínez, Luis Miguel
Albaladejo, Raquel
Torán, Pere
author_sort Martínez-Sánchez, Nuria
collection PubMed
description OBJECTIVE: An increased interarm blood pressure difference (IAD) (≥10 mmHg) has been associated with increased cardiovascular morbidity and mortality. There are few studies determining how IAD has to be measured and its reliability between visits. The objectives of our study were twofold. First, to evaluate the concordance between two automatic oscillometric devices for IAD measurement (two OMRON devices and one Microlife WatchBP™ device (WBP™)) and to analyse the reproducibility of IAD between visits in the general population attending a primary care centre. DESIGN: Descriptive cross-sectional study of concordance between the two methods and reproducibility of IAD between two visits separated by one week. SITE: Parets del Vallès primary care centre (Barcelona). PARTICIPANTS: General population aged 35-74 years. INTERVENTIONS AND MAIN MEASUREMENTS: One hundred and forty-nine patients completed the two visits. At each visit, IAD was measured three times with both methods and the mean of the three determinations was considered. Other variables such as sociodemographic and anthropometric variables, pathological antecedents and pharmacological treatment were collected through a review of the medical history and an interview with the patient. Concordance between the two devices and between visits reproducibility were calculated using the Lin concordance coefficient (CCL) for IAD expressed continuously and kappa(k) indices for IAD categorised as normal or pathological. RESULTS: Concordance for IAD expressed continuously was low: CCL = 0.13 (0.02-0.24). Concordance was also low for IAD categorised as normal or pathological (k = −0.03 (−0.05-0.00)). Reproducibility between visits was low for both methods and for both continuous and categorised IAD: with OMRON™ CCL = 0.19 (0.03-0.34) and k = −0.02 (−0.16-0.12) and for WBP™ CCL = 0.14 (−0.01-0.29) and k = 0.49 (0.33-0.64). CONCLUSIONS: Concordance between two automatic oscillometers in the simultaneous IAD measurement was low. Reproducibility between visits was also low for both methods.
format Online
Article
Text
id pubmed-9681639
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-96816392022-11-24 Diferencia de presión arterial entre brazos: concordancia entre 2 métodos automáticos de medición simultánea y reproducibilidad entre visitas Martínez-Sánchez, Nuria Palasí, Antonio Pera, Guillem Martínez, Luis Miguel Albaladejo, Raquel Torán, Pere Aten Primaria Original OBJECTIVE: An increased interarm blood pressure difference (IAD) (≥10 mmHg) has been associated with increased cardiovascular morbidity and mortality. There are few studies determining how IAD has to be measured and its reliability between visits. The objectives of our study were twofold. First, to evaluate the concordance between two automatic oscillometric devices for IAD measurement (two OMRON devices and one Microlife WatchBP™ device (WBP™)) and to analyse the reproducibility of IAD between visits in the general population attending a primary care centre. DESIGN: Descriptive cross-sectional study of concordance between the two methods and reproducibility of IAD between two visits separated by one week. SITE: Parets del Vallès primary care centre (Barcelona). PARTICIPANTS: General population aged 35-74 years. INTERVENTIONS AND MAIN MEASUREMENTS: One hundred and forty-nine patients completed the two visits. At each visit, IAD was measured three times with both methods and the mean of the three determinations was considered. Other variables such as sociodemographic and anthropometric variables, pathological antecedents and pharmacological treatment were collected through a review of the medical history and an interview with the patient. Concordance between the two devices and between visits reproducibility were calculated using the Lin concordance coefficient (CCL) for IAD expressed continuously and kappa(k) indices for IAD categorised as normal or pathological. RESULTS: Concordance for IAD expressed continuously was low: CCL = 0.13 (0.02-0.24). Concordance was also low for IAD categorised as normal or pathological (k = −0.03 (−0.05-0.00)). Reproducibility between visits was low for both methods and for both continuous and categorised IAD: with OMRON™ CCL = 0.19 (0.03-0.34) and k = −0.02 (−0.16-0.12) and for WBP™ CCL = 0.14 (−0.01-0.29) and k = 0.49 (0.33-0.64). CONCLUSIONS: Concordance between two automatic oscillometers in the simultaneous IAD measurement was low. Reproducibility between visits was also low for both methods. Elsevier 2022-12 2022-11-21 /pmc/articles/PMC9681639/ /pubmed/36423568 http://dx.doi.org/10.1016/j.aprim.2022.102514 Text en © 2022 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original
Martínez-Sánchez, Nuria
Palasí, Antonio
Pera, Guillem
Martínez, Luis Miguel
Albaladejo, Raquel
Torán, Pere
Diferencia de presión arterial entre brazos: concordancia entre 2 métodos automáticos de medición simultánea y reproducibilidad entre visitas
title Diferencia de presión arterial entre brazos: concordancia entre 2 métodos automáticos de medición simultánea y reproducibilidad entre visitas
title_full Diferencia de presión arterial entre brazos: concordancia entre 2 métodos automáticos de medición simultánea y reproducibilidad entre visitas
title_fullStr Diferencia de presión arterial entre brazos: concordancia entre 2 métodos automáticos de medición simultánea y reproducibilidad entre visitas
title_full_unstemmed Diferencia de presión arterial entre brazos: concordancia entre 2 métodos automáticos de medición simultánea y reproducibilidad entre visitas
title_short Diferencia de presión arterial entre brazos: concordancia entre 2 métodos automáticos de medición simultánea y reproducibilidad entre visitas
title_sort diferencia de presión arterial entre brazos: concordancia entre 2 métodos automáticos de medición simultánea y reproducibilidad entre visitas
topic Original
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9681639/
https://www.ncbi.nlm.nih.gov/pubmed/36423568
http://dx.doi.org/10.1016/j.aprim.2022.102514
work_keys_str_mv AT martinezsancheznuria diferenciadepresionarterialentrebrazosconcordanciaentre2metodosautomaticosdemedicionsimultaneayreproducibilidadentrevisitas
AT palasiantonio diferenciadepresionarterialentrebrazosconcordanciaentre2metodosautomaticosdemedicionsimultaneayreproducibilidadentrevisitas
AT peraguillem diferenciadepresionarterialentrebrazosconcordanciaentre2metodosautomaticosdemedicionsimultaneayreproducibilidadentrevisitas
AT martinezluismiguel diferenciadepresionarterialentrebrazosconcordanciaentre2metodosautomaticosdemedicionsimultaneayreproducibilidadentrevisitas
AT albaladejoraquel diferenciadepresionarterialentrebrazosconcordanciaentre2metodosautomaticosdemedicionsimultaneayreproducibilidadentrevisitas
AT toranpere diferenciadepresionarterialentrebrazosconcordanciaentre2metodosautomaticosdemedicionsimultaneayreproducibilidadentrevisitas