Thyroglobulin Assay Interferences: Clinical Usefulness of Mass-Spectrometry Methods
CONTEXT: Thyroglobulin autoantibodies (TgAbs) affect thyroglobulin immunometric assays (TgIMAs), causing falsely low results. Conversely, heterophilic antibodies (HAs) may cause falsely elevated results. Thyroglobulin (Tg) measurements by mass spectrometry (MS) resist antibody interference. The most...
Autores principales: | Barbesino, Giuseppe, Algeciras-Schimnich, Alicia, Bornhorst, Joshua |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9683495/ https://www.ncbi.nlm.nih.gov/pubmed/36438548 http://dx.doi.org/10.1210/jendso/bvac169 |
Ejemplares similares
-
Development of a PTHrP Chemiluminescent Immunoassay to Assess Humoral Hypercalcemia of Malignancy
por: Ashrafzadeh-Kian, Susan Louise, et al.
Publicado: (2021) -
Comparison of intact protein and digested peptide techniques for high throughput proteotyping of ApoE
por: Maus, Anthony, et al.
Publicado: (2022) -
New Cutoffs for the Biochemical Diagnosis of Adrenal Insufficiency after ACTH Stimulation using Specific Cortisol Assays
por: Javorsky, Bradley R, et al.
Publicado: (2021) -
Impact of Thyroglobulin and Thyroglobulin Antibody Assay Performance on the Differential Classification of DTC Patients
por: Schoonen, Lise, et al.
Publicado: (2021) -
SAT-390 New Cortisol Threshold for Diagnosis of Adrenal Insufficiency After Cosyntropin Stimulation Testing Using the Elecsys Cortisol II, Access Cortisol, and LC-MS/MS Assays
por: Javorsky, Bradley, et al.