More Space, Less Noise—New-generation Low-Field Magnetic Resonance Imaging Systems Can Improve Patient Comfort: A Prospective 0.55T–1.5T-Scanner Comparison
Objectives: The objectives of this study were to assess patient comfort when imaged on a newly introduced 0.55T low-field magnetic resonance (MR) scanner system with a wider bore opening compared to a conventional 1.5T MR scanner system. Materials and Methods: In this prospective study, fifty patien...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9692643/ https://www.ncbi.nlm.nih.gov/pubmed/36431182 http://dx.doi.org/10.3390/jcm11226705 |
_version_ | 1784837318874497024 |
---|---|
author | Rusche, Thilo Vosshenrich, Jan Winkel, David J. Donners, Ricardo Segeroth, Martin Bach, Michael Merkle, Elmar M. Breit, Hanns-Christian |
author_facet | Rusche, Thilo Vosshenrich, Jan Winkel, David J. Donners, Ricardo Segeroth, Martin Bach, Michael Merkle, Elmar M. Breit, Hanns-Christian |
author_sort | Rusche, Thilo |
collection | PubMed |
description | Objectives: The objectives of this study were to assess patient comfort when imaged on a newly introduced 0.55T low-field magnetic resonance (MR) scanner system with a wider bore opening compared to a conventional 1.5T MR scanner system. Materials and Methods: In this prospective study, fifty patients (mean age: 66.2 ± 17.0 years, 22 females, 28 males) underwent subsequent magnetic resonance imaging (MRI) examinations with matched imaging protocols at 0.55T (MAGNETOM FreeMax, Siemens Healthineers; Erlangen, Germany) and 1.5T (MAGNETOM Avanto Fit, Siemens Healthineers; Erlangen, Germany) on the same day. MRI performed between 05/2021 and 07/2021 was included for analysis. The 0.55T MRI system had a bore opening of 80 cm, while the bore diameter of the 1.5T scanner system was 60 cm. Four patient groups were defined by imaged body regions: (1) cranial or cervical spine MRI using a head/neck coil (n = 27), (2) lumbar or thoracic spine MRI using only the in-table spine coils (n = 10), (3) hip MRI using a large flex coil (n = 8) and (4) upper- or lower-extremity MRI using small flex coils (n = 5). Following the MRI examinations, patients evaluated (1) sense of space, (2) noise level, (3) comfort, (4) coil comfort and (5) overall examination impression on a 5-point Likert-scale (range: 1= “much worse” to 5 = “much better”) using a questionnaire. Maximum noise levels of all performed imaging studies were measured in decibels (dB) by a sound level meter placed in the bore center. Results: Sense of space was perceived to be “better” or “much better” by 84% of patients for imaging examinations performed on the 0.55T MRI scanner system (mean score: 4.34 ± 0.75). Additionally, 84% of patients rated noise levels as “better” or “much better” when imaged on the low-field scanner system (mean score: 3.90 ± 0.61). Overall sensation during the imaging examination at 0.55T was rated as “better” or “much better” by 78% of patients (mean score: 3.96 ± 0.70). Quantitative assessment showed significantly reduced maximum noise levels for all 0.55T MRI studies, regardless of body region compared to 1.5T, i.e., brain MRI (83.8 ± 3.6 dB vs. 89.3 ± 5.4 dB; p = 0.04), spine MRI (83.7 ± 3.7 dB vs. 89.4 ± 2.6 dB; p = 0.004) and hip MRI (86.3 ± 5.0 dB vs. 89.1 ± 1.4 dB; p = 0.04). Conclusions: Patients perceived 0.55T new-generation low-field MRI to be more comfortable than conventional 1.5T MRI, given its larger bore opening and reduced noise levels during image acquisition. Therefore, new concepts regarding bore design and noise level reduction of MR scanner systems may help to reduce patient anxiety and improve well-being when undergoing MR imaging. |
format | Online Article Text |
id | pubmed-9692643 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-96926432022-11-26 More Space, Less Noise—New-generation Low-Field Magnetic Resonance Imaging Systems Can Improve Patient Comfort: A Prospective 0.55T–1.5T-Scanner Comparison Rusche, Thilo Vosshenrich, Jan Winkel, David J. Donners, Ricardo Segeroth, Martin Bach, Michael Merkle, Elmar M. Breit, Hanns-Christian J Clin Med Article Objectives: The objectives of this study were to assess patient comfort when imaged on a newly introduced 0.55T low-field magnetic resonance (MR) scanner system with a wider bore opening compared to a conventional 1.5T MR scanner system. Materials and Methods: In this prospective study, fifty patients (mean age: 66.2 ± 17.0 years, 22 females, 28 males) underwent subsequent magnetic resonance imaging (MRI) examinations with matched imaging protocols at 0.55T (MAGNETOM FreeMax, Siemens Healthineers; Erlangen, Germany) and 1.5T (MAGNETOM Avanto Fit, Siemens Healthineers; Erlangen, Germany) on the same day. MRI performed between 05/2021 and 07/2021 was included for analysis. The 0.55T MRI system had a bore opening of 80 cm, while the bore diameter of the 1.5T scanner system was 60 cm. Four patient groups were defined by imaged body regions: (1) cranial or cervical spine MRI using a head/neck coil (n = 27), (2) lumbar or thoracic spine MRI using only the in-table spine coils (n = 10), (3) hip MRI using a large flex coil (n = 8) and (4) upper- or lower-extremity MRI using small flex coils (n = 5). Following the MRI examinations, patients evaluated (1) sense of space, (2) noise level, (3) comfort, (4) coil comfort and (5) overall examination impression on a 5-point Likert-scale (range: 1= “much worse” to 5 = “much better”) using a questionnaire. Maximum noise levels of all performed imaging studies were measured in decibels (dB) by a sound level meter placed in the bore center. Results: Sense of space was perceived to be “better” or “much better” by 84% of patients for imaging examinations performed on the 0.55T MRI scanner system (mean score: 4.34 ± 0.75). Additionally, 84% of patients rated noise levels as “better” or “much better” when imaged on the low-field scanner system (mean score: 3.90 ± 0.61). Overall sensation during the imaging examination at 0.55T was rated as “better” or “much better” by 78% of patients (mean score: 3.96 ± 0.70). Quantitative assessment showed significantly reduced maximum noise levels for all 0.55T MRI studies, regardless of body region compared to 1.5T, i.e., brain MRI (83.8 ± 3.6 dB vs. 89.3 ± 5.4 dB; p = 0.04), spine MRI (83.7 ± 3.7 dB vs. 89.4 ± 2.6 dB; p = 0.004) and hip MRI (86.3 ± 5.0 dB vs. 89.1 ± 1.4 dB; p = 0.04). Conclusions: Patients perceived 0.55T new-generation low-field MRI to be more comfortable than conventional 1.5T MRI, given its larger bore opening and reduced noise levels during image acquisition. Therefore, new concepts regarding bore design and noise level reduction of MR scanner systems may help to reduce patient anxiety and improve well-being when undergoing MR imaging. MDPI 2022-11-12 /pmc/articles/PMC9692643/ /pubmed/36431182 http://dx.doi.org/10.3390/jcm11226705 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Rusche, Thilo Vosshenrich, Jan Winkel, David J. Donners, Ricardo Segeroth, Martin Bach, Michael Merkle, Elmar M. Breit, Hanns-Christian More Space, Less Noise—New-generation Low-Field Magnetic Resonance Imaging Systems Can Improve Patient Comfort: A Prospective 0.55T–1.5T-Scanner Comparison |
title | More Space, Less Noise—New-generation Low-Field Magnetic Resonance Imaging Systems Can Improve Patient Comfort: A Prospective 0.55T–1.5T-Scanner Comparison |
title_full | More Space, Less Noise—New-generation Low-Field Magnetic Resonance Imaging Systems Can Improve Patient Comfort: A Prospective 0.55T–1.5T-Scanner Comparison |
title_fullStr | More Space, Less Noise—New-generation Low-Field Magnetic Resonance Imaging Systems Can Improve Patient Comfort: A Prospective 0.55T–1.5T-Scanner Comparison |
title_full_unstemmed | More Space, Less Noise—New-generation Low-Field Magnetic Resonance Imaging Systems Can Improve Patient Comfort: A Prospective 0.55T–1.5T-Scanner Comparison |
title_short | More Space, Less Noise—New-generation Low-Field Magnetic Resonance Imaging Systems Can Improve Patient Comfort: A Prospective 0.55T–1.5T-Scanner Comparison |
title_sort | more space, less noise—new-generation low-field magnetic resonance imaging systems can improve patient comfort: a prospective 0.55t–1.5t-scanner comparison |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9692643/ https://www.ncbi.nlm.nih.gov/pubmed/36431182 http://dx.doi.org/10.3390/jcm11226705 |
work_keys_str_mv | AT ruschethilo morespacelessnoisenewgenerationlowfieldmagneticresonanceimagingsystemscanimprovepatientcomfortaprospective055t15tscannercomparison AT vosshenrichjan morespacelessnoisenewgenerationlowfieldmagneticresonanceimagingsystemscanimprovepatientcomfortaprospective055t15tscannercomparison AT winkeldavidj morespacelessnoisenewgenerationlowfieldmagneticresonanceimagingsystemscanimprovepatientcomfortaprospective055t15tscannercomparison AT donnersricardo morespacelessnoisenewgenerationlowfieldmagneticresonanceimagingsystemscanimprovepatientcomfortaprospective055t15tscannercomparison AT segerothmartin morespacelessnoisenewgenerationlowfieldmagneticresonanceimagingsystemscanimprovepatientcomfortaprospective055t15tscannercomparison AT bachmichael morespacelessnoisenewgenerationlowfieldmagneticresonanceimagingsystemscanimprovepatientcomfortaprospective055t15tscannercomparison AT merkleelmarm morespacelessnoisenewgenerationlowfieldmagneticresonanceimagingsystemscanimprovepatientcomfortaprospective055t15tscannercomparison AT breithannschristian morespacelessnoisenewgenerationlowfieldmagneticresonanceimagingsystemscanimprovepatientcomfortaprospective055t15tscannercomparison |