Determination of Glyphosate in White and Brown Rice with HPLC-ICP-MS/MS

Background: In 2017, the European Commission renewed the approval of glyphosate (GLY) but only for five years. GLY remains one of the most controversial and studied molecules. Method: A simplified method was tested for the determination of GLY in white rice (WR) and brown rice (BR), after extraction...

Descripción completa

Detalles Bibliográficos
Autores principales: Fontanella, Maria Chiara, Lamastra, Lucrezia, Beone, Gian Maria
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9696991/
https://www.ncbi.nlm.nih.gov/pubmed/36432148
http://dx.doi.org/10.3390/molecules27228049
Descripción
Sumario:Background: In 2017, the European Commission renewed the approval of glyphosate (GLY) but only for five years. GLY remains one of the most controversial and studied molecules. Method: A simplified method was tested for the determination of GLY in white rice (WR) and brown rice (BR), after extraction only with a methanol solution, by liquid chromatography coupled with inductively coupled mass triple quadrupole (HPLC-ICP-MS/MS) with a PRP-X100 anionic column. After performing a test on groundwater, the quantification of GLY in WR and BR was validated in terms of the LOD, LOQ, accuracy, precision, linearity, and the matrix effect. Results: The LOD was 0.0027 mg kg(−1) for WR and 0.0136 mg kg(−1) for BR. The LOQ was 0.0092 mg kg(−1) for WR and 0.0456 mg kg(−1) for BR. The mean recoveries were within 76–105% at three fortification levels. The relative standard deviation for the analysis (five replicates for three spike levels) was < 11% for both matrices. A linear response was confirmed in all cases in the entire concentration range (R(2)(WR) = 1.000 and R(2)(BR) = 0.9818). Conclusion: The proposed method could be considered useful for the determination of GLY in different types of rice and designed and adapted for other cereals. The matrix effect, quantified in BR matrix extraction, could be avoided by using a matrix-matched calibration line.