Cargando…
First-in-Human Experience of Using a Universal Steerable Sheath in Implanting Left Atrial Appendage Closure Devices
Left atrial appendage (LAA) closure (LAAC) device implantation may be challenging in cases with difficult LAA anatomy. A deflectable sheath to approach multiple LAA positions may be helpful. We used a deflectable sheath to implant LAAC devices in 20 patients and included 60 cases using the standard...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9700041/ https://www.ncbi.nlm.nih.gov/pubmed/36444320 http://dx.doi.org/10.1016/j.jacasi.2022.08.003 |
Sumario: | Left atrial appendage (LAA) closure (LAAC) device implantation may be challenging in cases with difficult LAA anatomy. A deflectable sheath to approach multiple LAA positions may be helpful. We used a deflectable sheath to implant LAAC devices in 20 patients and included 60 cases using the standard sheaths for comparison. The procedures were successful in all patients without peri-procedural complications. After a median follow-up of 1.52 (IQR: 0.76-2.33) years, none of the patients in the deflectable sheath group had peri-device leak ≥3 mm and experienced thromboembolic stroke. In the standard sheath group, after follow-up of 2.03 (IQR: 0.87-3.19) years, 2 had peri-device leak ≥3 mm and 1 experienced thromboembolic stroke. We first proved the idea of using a universal steerable sheath for LAAC device implantation in difficult LAA anatomy, which also allows rapid switching of different LAAC devices. |
---|