Catastrophic Trunnion Failure in an Anatomic Titanium Alloy Stem

INTRODUCTION: Certain titanium alloy stems have been shown to be susceptible to failure at the neck with catastrophic trunnion failure. Failure has been particularly noted in the single wedge Accolade 1 stem design. Other stems also used this alloy including the anatomic designed Citation stem. METH...

Descripción completa

Detalles Bibliográficos
Autores principales: Pruitt, Benjamin P., Mears, Simon C, Apple, Andrew E., Stambough, Jeffrey B., Barnes, C. Lowry, Stronach, Benjamin M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9709190/
https://www.ncbi.nlm.nih.gov/pubmed/36465993
http://dx.doi.org/10.1177/21514593221142726
_version_ 1784841093001510912
author Pruitt, Benjamin P.
Mears, Simon C
Apple, Andrew E.
Stambough, Jeffrey B.
Barnes, C. Lowry
Stronach, Benjamin M.
author_facet Pruitt, Benjamin P.
Mears, Simon C
Apple, Andrew E.
Stambough, Jeffrey B.
Barnes, C. Lowry
Stronach, Benjamin M.
author_sort Pruitt, Benjamin P.
collection PubMed
description INTRODUCTION: Certain titanium alloy stems have been shown to be susceptible to failure at the neck with catastrophic trunnion failure. Failure has been particularly noted in the single wedge Accolade 1 stem design. Other stems also used this alloy including the anatomic designed Citation stem. METHODS: This case series details 3 catastrophic failures of the TMZF version of the Citation femoral stem. RESULTS: Each of these failures appear to be attributed to cyclical wear of the TMZF trunnion against the cobalt chromium femoral head. Wear resulted in ultimate implant failure and significant metal debris in the joint capsule at the time of revision surgery. DISCUSSION: While surgeons are aware of the risk of catastrophic failure for the Accolade 1 stem, failure may similarly happen in the TMZF Citation stem. Surgeons should monitor these implants with care and discuss the potential for trunnion failure with their patients.
format Online
Article
Text
id pubmed-9709190
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-97091902022-12-01 Catastrophic Trunnion Failure in an Anatomic Titanium Alloy Stem Pruitt, Benjamin P. Mears, Simon C Apple, Andrew E. Stambough, Jeffrey B. Barnes, C. Lowry Stronach, Benjamin M. Geriatr Orthop Surg Rehabil Medical Student Corner INTRODUCTION: Certain titanium alloy stems have been shown to be susceptible to failure at the neck with catastrophic trunnion failure. Failure has been particularly noted in the single wedge Accolade 1 stem design. Other stems also used this alloy including the anatomic designed Citation stem. METHODS: This case series details 3 catastrophic failures of the TMZF version of the Citation femoral stem. RESULTS: Each of these failures appear to be attributed to cyclical wear of the TMZF trunnion against the cobalt chromium femoral head. Wear resulted in ultimate implant failure and significant metal debris in the joint capsule at the time of revision surgery. DISCUSSION: While surgeons are aware of the risk of catastrophic failure for the Accolade 1 stem, failure may similarly happen in the TMZF Citation stem. Surgeons should monitor these implants with care and discuss the potential for trunnion failure with their patients. SAGE Publications 2022-11-28 /pmc/articles/PMC9709190/ /pubmed/36465993 http://dx.doi.org/10.1177/21514593221142726 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by-nc/4.0/This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Medical Student Corner
Pruitt, Benjamin P.
Mears, Simon C
Apple, Andrew E.
Stambough, Jeffrey B.
Barnes, C. Lowry
Stronach, Benjamin M.
Catastrophic Trunnion Failure in an Anatomic Titanium Alloy Stem
title Catastrophic Trunnion Failure in an Anatomic Titanium Alloy Stem
title_full Catastrophic Trunnion Failure in an Anatomic Titanium Alloy Stem
title_fullStr Catastrophic Trunnion Failure in an Anatomic Titanium Alloy Stem
title_full_unstemmed Catastrophic Trunnion Failure in an Anatomic Titanium Alloy Stem
title_short Catastrophic Trunnion Failure in an Anatomic Titanium Alloy Stem
title_sort catastrophic trunnion failure in an anatomic titanium alloy stem
topic Medical Student Corner
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9709190/
https://www.ncbi.nlm.nih.gov/pubmed/36465993
http://dx.doi.org/10.1177/21514593221142726
work_keys_str_mv AT pruittbenjaminp catastrophictrunnionfailureinananatomictitaniumalloystem
AT mearssimonc catastrophictrunnionfailureinananatomictitaniumalloystem
AT appleandrewe catastrophictrunnionfailureinananatomictitaniumalloystem
AT stamboughjeffreyb catastrophictrunnionfailureinananatomictitaniumalloystem
AT barnesclowry catastrophictrunnionfailureinananatomictitaniumalloystem
AT stronachbenjaminm catastrophictrunnionfailureinananatomictitaniumalloystem