Cargando…
Outcomes of novel 3D-printed fully porous titanium cup and a cemented highly cross-linked polyethylene liner in complex and revision total hip arthroplasty
INTRODUCTION: A novel fully porous acetabular titanium shell has been designed to reduce stiffness mismatch between bone and implant and promote osseointegration in complex (cTHA) and revision total hip arthroplasty (rTHA). A highly cross-linked polyethylene (XLPE) liner is cemented within the cup t...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9717502/ https://www.ncbi.nlm.nih.gov/pubmed/36457035 http://dx.doi.org/10.1186/s42836-022-00152-5 |
Sumario: | INTRODUCTION: A novel fully porous acetabular titanium shell has been designed to reduce stiffness mismatch between bone and implant and promote osseointegration in complex (cTHA) and revision total hip arthroplasty (rTHA). A highly cross-linked polyethylene (XLPE) liner is cemented within the cup to reduce wear rates and increase survivorship. This study reported the outcomes of an XLPE liner cemented into a novel 3D-printed fully porous cup in cTHA and rTHA. METHODS: Presented was a multicenter retrospective review of 40 patients (6 cTHA and 34 rTHA) who underwent THA with a fully porous titanium acetabular cup and cemented XLPE liner. Data were collected on demographics, surgical information, outcomes, including osseointegration and migration and implant survivorship. RESULTS: On average, patients were 71.42 ± 9.97 years old and obese (BMI: 30.36 ± 6.88 kg/m(2)) and were followed up for a mean time of 2.21 ± 0.77 years. Six patients underwent cTHA and 34 patients underwent rTHA. The mean hospital length of stay was 5.34 ± 3.34 days. Three (7.5%) 90-day readmissions were noted. Harris Hip Scores improved, on average, from 53.87 ± 12.58 preoperatively to 83.53 ± 12.15 postoperatively (P<0.001). One case of acetabular shell aspetic loosening with migration was noted. Thirty-nine of the 40 acetabular components were fully osseointegrated without migration. Two patients underwent re-revision surgery for PJI and one patient received acetabular shell+liner re-revision due to aseptic loosening. Kaplan-Meier analysis showed an all-cause revision-free survival rate of 95.0% at 6 months and 1 year, and 92.0% at 4-years. Aseptic acetabular cup, liner dislocation/loosening, and fracture-free survival was 100% at 6 months and 1-year, and 97.1% at 2 years. CONCLUSION: The combined use of a novel 3D-printed fully porous titanium acetabular shell and cemented XLPE acetabular liner yielded excellent rates of osseointegration, and all-cause and acetabular aseptic loosening survivorship at a minimum 1-year follow-up. Further long-term studies are needed to assess the longevity of this construct. |
---|