Arrhythmia Following Pulse Generator Change: What Is the Mechanism?
We present an unusual case of a “reverse” pacemaker-mediated endless loop arrhythmia with native atrioventricular conduction serving as the anterograde limb of the tachycardia circuit and the atrial depolarization stimulated by the pacemaker in response to the sensed ventricular QRS. The electrocard...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MediaSphere Medical
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9721300/ https://www.ncbi.nlm.nih.gov/pubmed/36570479 http://dx.doi.org/10.19102/icrm.2022.13115 |
Sumario: | We present an unusual case of a “reverse” pacemaker-mediated endless loop arrhythmia with native atrioventricular conduction serving as the anterograde limb of the tachycardia circuit and the atrial depolarization stimulated by the pacemaker in response to the sensed ventricular QRS. The electrocardiogram findings can be explained by lead reversal in the header during pulse generator change, with the ventricular lead connected to the atrial port and the atrial lead connected to the ventricular port. Careful examination of the electrocardiogram with attention to the 2 closely spaced pacing stimuli separated by the paced atrioventricular delay provided an important clue for diagnosis this case. |
---|