A ruptured balloon shaft during an angioplasty
We report a case of balloon shaft rupture during percutaneous coronary intervention. Although the entrapped balloon was not yet deflated when the complication occurred, we successfully retrieved it percutaneously using a trapping technique. This case described a cheap and straightforward technique o...