Colon mucosal injury caused by water jet malfunction during a screening colonoscopy: A case report

BACKGROUND: Screening colonoscopies are routinely performed and have low occurrences of adverse events such as perforation, bleeding, infection, and post-polypectomy syndrome. True device related adverse events are rarely reported in the literature. CASE SUMMARY: We report a case of a 51-year-old pa...

Descripción completa

Detalles Bibliográficos
Autores principales: Patel, Parth, Chen, Chien-Huan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Baishideng Publishing Group Inc 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9724532/
https://www.ncbi.nlm.nih.gov/pubmed/36483814
http://dx.doi.org/10.12998/wjcc.v10.i33.12375
_version_ 1784844438616408064
author Patel, Parth
Chen, Chien-Huan
author_facet Patel, Parth
Chen, Chien-Huan
author_sort Patel, Parth
collection PubMed
description BACKGROUND: Screening colonoscopies are routinely performed and have low occurrences of adverse events such as perforation, bleeding, infection, and post-polypectomy syndrome. True device related adverse events are rarely reported in the literature. CASE SUMMARY: We report a case of a 51-year-old patient without past medical history who presented for her first screening colonoscopy. The patient was thought to have friable mucosa in the cecum and oozed upon water irrigation during screening colonoscopy. It was later identified that the colonoscope used during the index procedure had malfunctioned and produced a pin-point water jet which damaged the colon mucosa of cecum. The maintenance service identified a piece of rubber fragment lodged in the instrument component at the tip of the scope, resulting in high pressure water jet. Repeat colonoscopy with a functioning colonoscope confirmed normal colon mucosa without friability. CONCLUSION: This is the first report of mucosa injury from a colonoscope water jet malfunction. Endoscopists should recognize the potential for endoscopic malfunction.
format Online
Article
Text
id pubmed-9724532
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Baishideng Publishing Group Inc
record_format MEDLINE/PubMed
spelling pubmed-97245322022-12-07 Colon mucosal injury caused by water jet malfunction during a screening colonoscopy: A case report Patel, Parth Chen, Chien-Huan World J Clin Cases Case Report BACKGROUND: Screening colonoscopies are routinely performed and have low occurrences of adverse events such as perforation, bleeding, infection, and post-polypectomy syndrome. True device related adverse events are rarely reported in the literature. CASE SUMMARY: We report a case of a 51-year-old patient without past medical history who presented for her first screening colonoscopy. The patient was thought to have friable mucosa in the cecum and oozed upon water irrigation during screening colonoscopy. It was later identified that the colonoscope used during the index procedure had malfunctioned and produced a pin-point water jet which damaged the colon mucosa of cecum. The maintenance service identified a piece of rubber fragment lodged in the instrument component at the tip of the scope, resulting in high pressure water jet. Repeat colonoscopy with a functioning colonoscope confirmed normal colon mucosa without friability. CONCLUSION: This is the first report of mucosa injury from a colonoscope water jet malfunction. Endoscopists should recognize the potential for endoscopic malfunction. Baishideng Publishing Group Inc 2022-11-26 2022-11-26 /pmc/articles/PMC9724532/ /pubmed/36483814 http://dx.doi.org/10.12998/wjcc.v10.i33.12375 Text en ©The Author(s) 2022. Published by Baishideng Publishing Group Inc. All rights reserved. https://creativecommons.org/licenses/by-nc/4.0/This article is an open-access article that was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution NonCommercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: https://creativecommons.org/Licenses/by-nc/4.0/
spellingShingle Case Report
Patel, Parth
Chen, Chien-Huan
Colon mucosal injury caused by water jet malfunction during a screening colonoscopy: A case report
title Colon mucosal injury caused by water jet malfunction during a screening colonoscopy: A case report
title_full Colon mucosal injury caused by water jet malfunction during a screening colonoscopy: A case report
title_fullStr Colon mucosal injury caused by water jet malfunction during a screening colonoscopy: A case report
title_full_unstemmed Colon mucosal injury caused by water jet malfunction during a screening colonoscopy: A case report
title_short Colon mucosal injury caused by water jet malfunction during a screening colonoscopy: A case report
title_sort colon mucosal injury caused by water jet malfunction during a screening colonoscopy: a case report
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9724532/
https://www.ncbi.nlm.nih.gov/pubmed/36483814
http://dx.doi.org/10.12998/wjcc.v10.i33.12375
work_keys_str_mv AT patelparth colonmucosalinjurycausedbywaterjetmalfunctionduringascreeningcolonoscopyacasereport
AT chenchienhuan colonmucosalinjurycausedbywaterjetmalfunctionduringascreeningcolonoscopyacasereport