Evaluation of a new cemented highly cross-linked all-polyethylene cup: a prospective and randomised study assessing wear and fixation characteristics using radiostereometric analysis
BACKGROUND AND PURPOSE: The aim of this prospective, randomised and controlled study was to evaluate the wear and fixation properties of a new cemented highly cross-linked all-polyethylene (HXLPE) cup in comparison with a conventional cemented ultra-high molecular weight polyethylene (ConvPE) cup us...
Autores principales: | Otten, Volker, Wästerlund, Daniel, Lindbjörn, John, Mertens, Carl, Mukka, Sebastian, Crnalic, Sead, Nilsson, Kjell G |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9726741/ https://www.ncbi.nlm.nih.gov/pubmed/33566703 http://dx.doi.org/10.1177/1120700021989991 |
Ejemplares similares
-
Uncemented cups with and without screw holes in primary THA: a Swedish Hip Arthroplasty Register study with 22,725 hips
por: Otten, Volker, et al.
Publicado: (2019) -
Precision of radiostereometric analysis (RSA) of acetabular cup stability and polyethylene wear improved by adding tantalum beads to the liner
por: Nebergall, Audrey K, et al.
Publicado: (2015) -
Direct comparison of polyethylene wear in cemented and uncemented acetabular cups
por: Bjerkholt, Håvard, et al.
Publicado: (2010) -
Uncemented monoblock trabecular metal posterior stabilized high-flex total knee arthroplasty: similar pattern of migration to the cruciate-retaining design — a prospective radiostereometric analysis (RSA) and clinical evaluation of 40 patients (49 knees) 60 years or younger with 9 years’ follow-up
por: Wojtowicz, Radoslaw, et al.
Publicado: (2019) -
Superior accuracy of model-based radiostereometric
analysis for measurement of polyethylene wear: A phantom study
por: Stilling, M., et al.
Publicado: (2012)