Motion Artifact Reduction in Electrocardiogram Signals Through a Redundant Denoising Independent Component Analysis Method for Wearable Health Care Monitoring Systems: Algorithm Development and Validation

BACKGROUND: The quest for improved diagnosis and treatment in home health care models has led to the development of wearable medical devices for remote vital signs monitoring. An accurate signal and a high diagnostic yield are critical for the cost-effectiveness of wearable health care monitoring sy...

Descripción completa

Detalles Bibliográficos
Autores principales: Castaño Usuga, Fabian Andres, Gissel, Christian, Hernández, Alher Mauricio
Formato: Online Artículo Texto
Lenguaje:English
Publicado: JMIR Publications 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9736764/
https://www.ncbi.nlm.nih.gov/pubmed/36274196
http://dx.doi.org/10.2196/40826
_version_ 1784847115238768640
author Castaño Usuga, Fabian Andres
Gissel, Christian
Hernández, Alher Mauricio
author_facet Castaño Usuga, Fabian Andres
Gissel, Christian
Hernández, Alher Mauricio
author_sort Castaño Usuga, Fabian Andres
collection PubMed
description BACKGROUND: The quest for improved diagnosis and treatment in home health care models has led to the development of wearable medical devices for remote vital signs monitoring. An accurate signal and a high diagnostic yield are critical for the cost-effectiveness of wearable health care monitoring systems and their widespread application in resource-constrained environments. Despite technological advances, the information acquired by these devices can be contaminated by motion artifacts (MA) leading to misdiagnosis or repeated procedures with increases in associated costs. This makes it necessary to develop methods to improve the quality of the signal acquired by these devices. OBJECTIVE: We aimed to present a novel method for electrocardiogram (ECG) signal denoising to reduce MA. We aimed to analyze the method’s performance and to compare its performance to that of existing approaches. METHODS: We present the novel Redundant denoising Independent Component Analysis method for ECG signal denoising based on the redundant and simultaneous acquisition of ECG signals and movement information, multichannel processing, and performance assessment considering the information contained in the signal waveform. The method is based on data including ECG signals from the patient’s chest and back, the acquisition of triaxial movement signals from inertial measurement units, a reference signal synthesized from an autoregressive model, and the separation of interest and noise sources through multichannel independent component analysis. RESULTS: The proposed method significantly reduced MA, showing better performance and introducing a smaller distortion in the interest signal compared with other methods. Finally, the performance of the proposed method was compared to that of wavelet shrinkage and wavelet independent component analysis through the assessment of signal-to-noise ratio, dynamic time warping, and a proposed index based on the signal waveform evaluation with an ensemble average ECG. CONCLUSIONS: Our novel ECG denoising method is a contribution to converting wearable devices into medical monitoring tools that can be used to support the remote diagnosis and monitoring of cardiovascular diseases. A more accurate signal substantially improves the diagnostic yield of wearable devices. A better yield improves the devices’ cost-effectiveness and contributes to their widespread application.
format Online
Article
Text
id pubmed-9736764
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher JMIR Publications
record_format MEDLINE/PubMed
spelling pubmed-97367642022-12-11 Motion Artifact Reduction in Electrocardiogram Signals Through a Redundant Denoising Independent Component Analysis Method for Wearable Health Care Monitoring Systems: Algorithm Development and Validation Castaño Usuga, Fabian Andres Gissel, Christian Hernández, Alher Mauricio JMIR Med Inform Original Paper BACKGROUND: The quest for improved diagnosis and treatment in home health care models has led to the development of wearable medical devices for remote vital signs monitoring. An accurate signal and a high diagnostic yield are critical for the cost-effectiveness of wearable health care monitoring systems and their widespread application in resource-constrained environments. Despite technological advances, the information acquired by these devices can be contaminated by motion artifacts (MA) leading to misdiagnosis or repeated procedures with increases in associated costs. This makes it necessary to develop methods to improve the quality of the signal acquired by these devices. OBJECTIVE: We aimed to present a novel method for electrocardiogram (ECG) signal denoising to reduce MA. We aimed to analyze the method’s performance and to compare its performance to that of existing approaches. METHODS: We present the novel Redundant denoising Independent Component Analysis method for ECG signal denoising based on the redundant and simultaneous acquisition of ECG signals and movement information, multichannel processing, and performance assessment considering the information contained in the signal waveform. The method is based on data including ECG signals from the patient’s chest and back, the acquisition of triaxial movement signals from inertial measurement units, a reference signal synthesized from an autoregressive model, and the separation of interest and noise sources through multichannel independent component analysis. RESULTS: The proposed method significantly reduced MA, showing better performance and introducing a smaller distortion in the interest signal compared with other methods. Finally, the performance of the proposed method was compared to that of wavelet shrinkage and wavelet independent component analysis through the assessment of signal-to-noise ratio, dynamic time warping, and a proposed index based on the signal waveform evaluation with an ensemble average ECG. CONCLUSIONS: Our novel ECG denoising method is a contribution to converting wearable devices into medical monitoring tools that can be used to support the remote diagnosis and monitoring of cardiovascular diseases. A more accurate signal substantially improves the diagnostic yield of wearable devices. A better yield improves the devices’ cost-effectiveness and contributes to their widespread application. JMIR Publications 2022-11-25 /pmc/articles/PMC9736764/ /pubmed/36274196 http://dx.doi.org/10.2196/40826 Text en ©Fabian Andres Castaño Usuga, Christian Gissel, Alher Mauricio Hernández. Originally published in JMIR Medical Informatics (https://medinform.jmir.org), 25.11.2022. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Medical Informatics, is properly cited. The complete bibliographic information, a link to the original publication on https://medinform.jmir.org/, as well as this copyright and license information must be included.
spellingShingle Original Paper
Castaño Usuga, Fabian Andres
Gissel, Christian
Hernández, Alher Mauricio
Motion Artifact Reduction in Electrocardiogram Signals Through a Redundant Denoising Independent Component Analysis Method for Wearable Health Care Monitoring Systems: Algorithm Development and Validation
title Motion Artifact Reduction in Electrocardiogram Signals Through a Redundant Denoising Independent Component Analysis Method for Wearable Health Care Monitoring Systems: Algorithm Development and Validation
title_full Motion Artifact Reduction in Electrocardiogram Signals Through a Redundant Denoising Independent Component Analysis Method for Wearable Health Care Monitoring Systems: Algorithm Development and Validation
title_fullStr Motion Artifact Reduction in Electrocardiogram Signals Through a Redundant Denoising Independent Component Analysis Method for Wearable Health Care Monitoring Systems: Algorithm Development and Validation
title_full_unstemmed Motion Artifact Reduction in Electrocardiogram Signals Through a Redundant Denoising Independent Component Analysis Method for Wearable Health Care Monitoring Systems: Algorithm Development and Validation
title_short Motion Artifact Reduction in Electrocardiogram Signals Through a Redundant Denoising Independent Component Analysis Method for Wearable Health Care Monitoring Systems: Algorithm Development and Validation
title_sort motion artifact reduction in electrocardiogram signals through a redundant denoising independent component analysis method for wearable health care monitoring systems: algorithm development and validation
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9736764/
https://www.ncbi.nlm.nih.gov/pubmed/36274196
http://dx.doi.org/10.2196/40826
work_keys_str_mv AT castanousugafabianandres motionartifactreductioninelectrocardiogramsignalsthrougharedundantdenoisingindependentcomponentanalysismethodforwearablehealthcaremonitoringsystemsalgorithmdevelopmentandvalidation
AT gisselchristian motionartifactreductioninelectrocardiogramsignalsthrougharedundantdenoisingindependentcomponentanalysismethodforwearablehealthcaremonitoringsystemsalgorithmdevelopmentandvalidation
AT hernandezalhermauricio motionartifactreductioninelectrocardiogramsignalsthrougharedundantdenoisingindependentcomponentanalysismethodforwearablehealthcaremonitoringsystemsalgorithmdevelopmentandvalidation