The delivered dose assessment in pancreas SBRT with the target position determined using an in-house position monitoring system

PURPOSE: This study assessed the delivered dose accuracy in pancreas SBRT by incorporating the real-time target position determined using an in-house position monitoring system. METHODS AND MATERIALS: An online image-based position monitoring system, SeedTracker, was developed to monitor radiopaque...

Descripción completa

Detalles Bibliográficos
Autores principales: Arumugam, Sankar, Young, Tony, Johnston, Meredith, Pavey, Darren, Lee, Mark
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9743991/
https://www.ncbi.nlm.nih.gov/pubmed/36518308
http://dx.doi.org/10.3389/fonc.2022.1009916
_version_ 1784848822220881920
author Arumugam, Sankar
Young, Tony
Johnston, Meredith
Pavey, Darren
Lee, Mark
author_facet Arumugam, Sankar
Young, Tony
Johnston, Meredith
Pavey, Darren
Lee, Mark
author_sort Arumugam, Sankar
collection PubMed
description PURPOSE: This study assessed the delivered dose accuracy in pancreas SBRT by incorporating the real-time target position determined using an in-house position monitoring system. METHODS AND MATERIALS: An online image-based position monitoring system, SeedTracker, was developed to monitor radiopaque marker positions using monoscopic x-ray images, available from the Elekta XVI imaging system. This system was applied to patients receiving SBRT for pancreatic cancer on the MASTERPLAN Pilot trial (ACTRN 12617001642370). All patients were implanted pre-treatment with at least three peri-tumoral radiopaque markers for target localisation. During treatment delivery, marker positions were compared to expected positions delineated from the planning CT. The position tolerance of ±3mm from the expected position of the markers was set to trigger a gating event (GE) during treatment. The dosimetric impact of position deviations and actual dose delivered with position corrections was assessed by convolving the plan control point dose matrices with temporal target positions determined during treatment. RESULTS: Eight patients were treated within this study. At least one GE was observed in 38% of the treatment fractions and more than one GE was observed in 10% of the fractions. The position deviations resulted in the mean(range) difference of -0.1(-1.1 - 0.4)Gy in minimum dose to tumour and 1.9(-0.1- 4.6)Gy increase to Dmax to duodenum compared to planned dose. In actual treatment delivery with the patient realignment, the mean difference of tumour min dose and duodenal Dmax was reduced to 0.1(-1.0 – 1.1)Gy and 1.1 (-0.7 - 3.3)Gy respectively compared to the planned dose. CONCLUSIONS: The in-house real-time position monitoring system improved the treatment accuracy of pancreatic SBRT in a general-purpose linac and enabled assessment of delivered dose by incorporating the temporal target position during delivery. The intrafraction motion impacts the dose to tumour even if target position is maintained within a 3mm position tolerance.
format Online
Article
Text
id pubmed-9743991
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-97439912022-12-13 The delivered dose assessment in pancreas SBRT with the target position determined using an in-house position monitoring system Arumugam, Sankar Young, Tony Johnston, Meredith Pavey, Darren Lee, Mark Front Oncol Oncology PURPOSE: This study assessed the delivered dose accuracy in pancreas SBRT by incorporating the real-time target position determined using an in-house position monitoring system. METHODS AND MATERIALS: An online image-based position monitoring system, SeedTracker, was developed to monitor radiopaque marker positions using monoscopic x-ray images, available from the Elekta XVI imaging system. This system was applied to patients receiving SBRT for pancreatic cancer on the MASTERPLAN Pilot trial (ACTRN 12617001642370). All patients were implanted pre-treatment with at least three peri-tumoral radiopaque markers for target localisation. During treatment delivery, marker positions were compared to expected positions delineated from the planning CT. The position tolerance of ±3mm from the expected position of the markers was set to trigger a gating event (GE) during treatment. The dosimetric impact of position deviations and actual dose delivered with position corrections was assessed by convolving the plan control point dose matrices with temporal target positions determined during treatment. RESULTS: Eight patients were treated within this study. At least one GE was observed in 38% of the treatment fractions and more than one GE was observed in 10% of the fractions. The position deviations resulted in the mean(range) difference of -0.1(-1.1 - 0.4)Gy in minimum dose to tumour and 1.9(-0.1- 4.6)Gy increase to Dmax to duodenum compared to planned dose. In actual treatment delivery with the patient realignment, the mean difference of tumour min dose and duodenal Dmax was reduced to 0.1(-1.0 – 1.1)Gy and 1.1 (-0.7 - 3.3)Gy respectively compared to the planned dose. CONCLUSIONS: The in-house real-time position monitoring system improved the treatment accuracy of pancreatic SBRT in a general-purpose linac and enabled assessment of delivered dose by incorporating the temporal target position during delivery. The intrafraction motion impacts the dose to tumour even if target position is maintained within a 3mm position tolerance. Frontiers Media S.A. 2022-11-28 /pmc/articles/PMC9743991/ /pubmed/36518308 http://dx.doi.org/10.3389/fonc.2022.1009916 Text en Copyright © 2022 Arumugam, Young, Johnston, Pavey and Lee https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Oncology
Arumugam, Sankar
Young, Tony
Johnston, Meredith
Pavey, Darren
Lee, Mark
The delivered dose assessment in pancreas SBRT with the target position determined using an in-house position monitoring system
title The delivered dose assessment in pancreas SBRT with the target position determined using an in-house position monitoring system
title_full The delivered dose assessment in pancreas SBRT with the target position determined using an in-house position monitoring system
title_fullStr The delivered dose assessment in pancreas SBRT with the target position determined using an in-house position monitoring system
title_full_unstemmed The delivered dose assessment in pancreas SBRT with the target position determined using an in-house position monitoring system
title_short The delivered dose assessment in pancreas SBRT with the target position determined using an in-house position monitoring system
title_sort delivered dose assessment in pancreas sbrt with the target position determined using an in-house position monitoring system
topic Oncology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9743991/
https://www.ncbi.nlm.nih.gov/pubmed/36518308
http://dx.doi.org/10.3389/fonc.2022.1009916
work_keys_str_mv AT arumugamsankar thedelivereddoseassessmentinpancreassbrtwiththetargetpositiondeterminedusinganinhousepositionmonitoringsystem
AT youngtony thedelivereddoseassessmentinpancreassbrtwiththetargetpositiondeterminedusinganinhousepositionmonitoringsystem
AT johnstonmeredith thedelivereddoseassessmentinpancreassbrtwiththetargetpositiondeterminedusinganinhousepositionmonitoringsystem
AT paveydarren thedelivereddoseassessmentinpancreassbrtwiththetargetpositiondeterminedusinganinhousepositionmonitoringsystem
AT leemark thedelivereddoseassessmentinpancreassbrtwiththetargetpositiondeterminedusinganinhousepositionmonitoringsystem
AT arumugamsankar delivereddoseassessmentinpancreassbrtwiththetargetpositiondeterminedusinganinhousepositionmonitoringsystem
AT youngtony delivereddoseassessmentinpancreassbrtwiththetargetpositiondeterminedusinganinhousepositionmonitoringsystem
AT johnstonmeredith delivereddoseassessmentinpancreassbrtwiththetargetpositiondeterminedusinganinhousepositionmonitoringsystem
AT paveydarren delivereddoseassessmentinpancreassbrtwiththetargetpositiondeterminedusinganinhousepositionmonitoringsystem
AT leemark delivereddoseassessmentinpancreassbrtwiththetargetpositiondeterminedusinganinhousepositionmonitoringsystem