Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy

BACKGROUND: We evaluated the setup error and dose absorption of an immobilization system with a shell and wooden baseplate (SW) for lung stereotactic body radiotherapy (SBRT). MATERIALS AND METHODS: Setup errors in 109 patients immobilized with an SW or BodyFix system (BF) were compared. Dose attenu...

Descripción completa

Detalles Bibliográficos
Autores principales: Nitta, Yuya, Ueda, Yoshihiro, Murata, Seiya, Isono, Masaru, Ohira, Shingo, Masaoka, Akira, Inui, Shoki, Washio, Hayate, Yoshinaka, Reimi, Sagawa, Tomohiro, Miyazaki, Masayoshi, Teshima, Teruki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Via Medica 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9746647/
https://www.ncbi.nlm.nih.gov/pubmed/36523803
http://dx.doi.org/10.5603/RPOR.a2022.0089
_version_ 1784849411410493440
author Nitta, Yuya
Ueda, Yoshihiro
Murata, Seiya
Isono, Masaru
Ohira, Shingo
Masaoka, Akira
Inui, Shoki
Washio, Hayate
Yoshinaka, Reimi
Sagawa, Tomohiro
Miyazaki, Masayoshi
Teshima, Teruki
author_facet Nitta, Yuya
Ueda, Yoshihiro
Murata, Seiya
Isono, Masaru
Ohira, Shingo
Masaoka, Akira
Inui, Shoki
Washio, Hayate
Yoshinaka, Reimi
Sagawa, Tomohiro
Miyazaki, Masayoshi
Teshima, Teruki
author_sort Nitta, Yuya
collection PubMed
description BACKGROUND: We evaluated the setup error and dose absorption of an immobilization system with a shell and wooden baseplate (SW) for lung stereotactic body radiotherapy (SBRT). MATERIALS AND METHODS: Setup errors in 109 patients immobilized with an SW or BodyFix system (BF) were compared. Dose attenuation rates of materials for baseplates were measured with an ion-chamber. Ionization measurements were performed from 90° to 180° gantry angle in 10° increments, with the ball water equivalent phantom placed at the center of the wood and carbon baseplates whose effects on dose distribution were compared using an electron portal imaging device. RESULTS: The ratio for the anterior-posterior, cranial-caudal, and right-left of the cases within 3-mm registered shifts in interfractional setup error were 90.9%, 89.2%, and 97.4% for the SW, and 93.2%, 91.6%, and 98.0% for the BF, respectively. For intrafractional setup error, 98.3%, 97.4%, and 99.1% for the SW and 96.6%, 95.8%, and 98.7% for the BF were within 3-mm registered shifts, respectively. In the center position, the average (minimum/maximum) dose attenuation rates from 90° to 180° for the wooden and carbon baseplates were 0.5 (0.1/2.8)% and 1.0 (–0.1/10.1)% with 6 MV, respectively. The gamma passing rates of 2%/2 mm for the wooden and carbon baseplates were 99.7% and 98.3% (p < 0.01). CONCLUSIONS: The immobilization system with an SW is effective for lung SBRT since it is comparable to the BF in setup accuracy. Moreover, the wooden baseplate had lower radiation attenuation rates and affected the dose distribution less than the carbon baseplate.
format Online
Article
Text
id pubmed-9746647
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Via Medica
record_format MEDLINE/PubMed
spelling pubmed-97466472022-12-14 Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy Nitta, Yuya Ueda, Yoshihiro Murata, Seiya Isono, Masaru Ohira, Shingo Masaoka, Akira Inui, Shoki Washio, Hayate Yoshinaka, Reimi Sagawa, Tomohiro Miyazaki, Masayoshi Teshima, Teruki Rep Pract Oncol Radiother Research Paper BACKGROUND: We evaluated the setup error and dose absorption of an immobilization system with a shell and wooden baseplate (SW) for lung stereotactic body radiotherapy (SBRT). MATERIALS AND METHODS: Setup errors in 109 patients immobilized with an SW or BodyFix system (BF) were compared. Dose attenuation rates of materials for baseplates were measured with an ion-chamber. Ionization measurements were performed from 90° to 180° gantry angle in 10° increments, with the ball water equivalent phantom placed at the center of the wood and carbon baseplates whose effects on dose distribution were compared using an electron portal imaging device. RESULTS: The ratio for the anterior-posterior, cranial-caudal, and right-left of the cases within 3-mm registered shifts in interfractional setup error were 90.9%, 89.2%, and 97.4% for the SW, and 93.2%, 91.6%, and 98.0% for the BF, respectively. For intrafractional setup error, 98.3%, 97.4%, and 99.1% for the SW and 96.6%, 95.8%, and 98.7% for the BF were within 3-mm registered shifts, respectively. In the center position, the average (minimum/maximum) dose attenuation rates from 90° to 180° for the wooden and carbon baseplates were 0.5 (0.1/2.8)% and 1.0 (–0.1/10.1)% with 6 MV, respectively. The gamma passing rates of 2%/2 mm for the wooden and carbon baseplates were 99.7% and 98.3% (p < 0.01). CONCLUSIONS: The immobilization system with an SW is effective for lung SBRT since it is comparable to the BF in setup accuracy. Moreover, the wooden baseplate had lower radiation attenuation rates and affected the dose distribution less than the carbon baseplate. Via Medica 2022-10-31 /pmc/articles/PMC9746647/ /pubmed/36523803 http://dx.doi.org/10.5603/RPOR.a2022.0089 Text en © 2022 Greater Poland Cancer Centre https://creativecommons.org/licenses/by-nc-nd/4.0/This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially
spellingShingle Research Paper
Nitta, Yuya
Ueda, Yoshihiro
Murata, Seiya
Isono, Masaru
Ohira, Shingo
Masaoka, Akira
Inui, Shoki
Washio, Hayate
Yoshinaka, Reimi
Sagawa, Tomohiro
Miyazaki, Masayoshi
Teshima, Teruki
Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy
title Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy
title_full Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy
title_fullStr Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy
title_full_unstemmed Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy
title_short Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy
title_sort setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy
topic Research Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9746647/
https://www.ncbi.nlm.nih.gov/pubmed/36523803
http://dx.doi.org/10.5603/RPOR.a2022.0089
work_keys_str_mv AT nittayuya setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy
AT uedayoshihiro setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy
AT murataseiya setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy
AT isonomasaru setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy
AT ohirashingo setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy
AT masaokaakira setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy
AT inuishoki setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy
AT washiohayate setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy
AT yoshinakareimi setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy
AT sagawatomohiro setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy
AT miyazakimasayoshi setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy
AT teshimateruki setupaccuracyanddoseattenuationofawoodenimmobilizationsystemforlungstereotacticbodyradiotherapy