Investigating the Equivalent Plastic Strain in a Variable Ring Length and Strut Width Thin-Strut Bioresorbable Scaffold

PURPOSE: The ArterioSorb[Formula: see text] bioresorbable scaffold (BRS) developed by Arterius Ltd is about to enter first in man clinical trials. Previous generations of BRS have been vulnerable to brittle fracture, when expanded via balloon inflation in-vivo, which can be extremely detrimental to...

Descripción completa

Detalles Bibliográficos
Autores principales: Hoddy, Ben, Ahmed, Naveed, Al-Lamee, Kadem, Bullett, Nial, Curzen, Nick, Bressloff, Neil W.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9750924/
https://www.ncbi.nlm.nih.gov/pubmed/35819580
http://dx.doi.org/10.1007/s13239-022-00625-3
_version_ 1784850364367896576
author Hoddy, Ben
Ahmed, Naveed
Al-Lamee, Kadem
Bullett, Nial
Curzen, Nick
Bressloff, Neil W.
author_facet Hoddy, Ben
Ahmed, Naveed
Al-Lamee, Kadem
Bullett, Nial
Curzen, Nick
Bressloff, Neil W.
author_sort Hoddy, Ben
collection PubMed
description PURPOSE: The ArterioSorb[Formula: see text] bioresorbable scaffold (BRS) developed by Arterius Ltd is about to enter first in man clinical trials. Previous generations of BRS have been vulnerable to brittle fracture, when expanded via balloon inflation in-vivo, which can be extremely detrimental to patient outcome. Therefore, this study explores the effect of variable ring length and strut width (as facilitated by the ArterioSorb[Formula: see text] design) on fracture resistance via analysis of the distribution of equivalent plastic strain in the scaffold struts post expansion. Scaffold performance is also assessed with respect to side branch access, radial strength, final deployed diameter and percentage recoil. METHODS: Finite element analysis was conducted of the crimping, expansion and radial crushing of five scaffold designs comprising different variations in ring length and strut width. The Abaqus/Explicit (DS SIMULIA) solution method was used for all simulations. Direct comparison between in-silico predictions and in-vitro measurements of the performance of the open cell variant of the ArterioSorb[Formula: see text] were made. Paths across the width of the crown apex and around the scaffold rings were defined along which the plastic strain distribution was analysed. RESULTS: The in-silico results demonstrated good predictions of final shape for the baseline scaffold design. Percentage recoil and radial strength were predicted to be, respectively, 2.8 and 1.7 times higher than the experimentally measured values, predominantly due to the limitations of the anisotropic elasto-plastic material property model used for the scaffold. Average maximum values of equivalent plastic strain were up to 2.4 times higher in the wide strut designs relative to the narrow strut scaffolds. As well as the concomitant risk of strut fracture, the wide strut designs also exhibited twisting and splaying behaviour at the crowns located on the scaffold end rings. Not only are these phenomena detrimental to the radial strength and risk of strut fracture but they also increase the likelihood of damage to the vessel wall. However, the baseline scaffold design was observed to tolerate significant over expansion without inducing excessive plastic strains, a result which is particularly encouraging, due to post-dilatation being commonplace in clinical practice. CONCLUSION: Therefore, the narrow strut designs investigated herein, are likely to offer optimal performance and potentially better patient outcomes. Further work should address the material modelling of next generation polymeric BRS to more accurately capture their mechanical behaviour. Observation of the in-vitro testing indicates that the ArterioSorb[Formula: see text] BRS can tolerate greater levels of over expansion than anticipated.
format Online
Article
Text
id pubmed-9750924
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-97509242022-12-16 Investigating the Equivalent Plastic Strain in a Variable Ring Length and Strut Width Thin-Strut Bioresorbable Scaffold Hoddy, Ben Ahmed, Naveed Al-Lamee, Kadem Bullett, Nial Curzen, Nick Bressloff, Neil W. Cardiovasc Eng Technol Original Article PURPOSE: The ArterioSorb[Formula: see text] bioresorbable scaffold (BRS) developed by Arterius Ltd is about to enter first in man clinical trials. Previous generations of BRS have been vulnerable to brittle fracture, when expanded via balloon inflation in-vivo, which can be extremely detrimental to patient outcome. Therefore, this study explores the effect of variable ring length and strut width (as facilitated by the ArterioSorb[Formula: see text] design) on fracture resistance via analysis of the distribution of equivalent plastic strain in the scaffold struts post expansion. Scaffold performance is also assessed with respect to side branch access, radial strength, final deployed diameter and percentage recoil. METHODS: Finite element analysis was conducted of the crimping, expansion and radial crushing of five scaffold designs comprising different variations in ring length and strut width. The Abaqus/Explicit (DS SIMULIA) solution method was used for all simulations. Direct comparison between in-silico predictions and in-vitro measurements of the performance of the open cell variant of the ArterioSorb[Formula: see text] were made. Paths across the width of the crown apex and around the scaffold rings were defined along which the plastic strain distribution was analysed. RESULTS: The in-silico results demonstrated good predictions of final shape for the baseline scaffold design. Percentage recoil and radial strength were predicted to be, respectively, 2.8 and 1.7 times higher than the experimentally measured values, predominantly due to the limitations of the anisotropic elasto-plastic material property model used for the scaffold. Average maximum values of equivalent plastic strain were up to 2.4 times higher in the wide strut designs relative to the narrow strut scaffolds. As well as the concomitant risk of strut fracture, the wide strut designs also exhibited twisting and splaying behaviour at the crowns located on the scaffold end rings. Not only are these phenomena detrimental to the radial strength and risk of strut fracture but they also increase the likelihood of damage to the vessel wall. However, the baseline scaffold design was observed to tolerate significant over expansion without inducing excessive plastic strains, a result which is particularly encouraging, due to post-dilatation being commonplace in clinical practice. CONCLUSION: Therefore, the narrow strut designs investigated herein, are likely to offer optimal performance and potentially better patient outcomes. Further work should address the material modelling of next generation polymeric BRS to more accurately capture their mechanical behaviour. Observation of the in-vitro testing indicates that the ArterioSorb[Formula: see text] BRS can tolerate greater levels of over expansion than anticipated. Springer International Publishing 2022-07-11 2022 /pmc/articles/PMC9750924/ /pubmed/35819580 http://dx.doi.org/10.1007/s13239-022-00625-3 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Article
Hoddy, Ben
Ahmed, Naveed
Al-Lamee, Kadem
Bullett, Nial
Curzen, Nick
Bressloff, Neil W.
Investigating the Equivalent Plastic Strain in a Variable Ring Length and Strut Width Thin-Strut Bioresorbable Scaffold
title Investigating the Equivalent Plastic Strain in a Variable Ring Length and Strut Width Thin-Strut Bioresorbable Scaffold
title_full Investigating the Equivalent Plastic Strain in a Variable Ring Length and Strut Width Thin-Strut Bioresorbable Scaffold
title_fullStr Investigating the Equivalent Plastic Strain in a Variable Ring Length and Strut Width Thin-Strut Bioresorbable Scaffold
title_full_unstemmed Investigating the Equivalent Plastic Strain in a Variable Ring Length and Strut Width Thin-Strut Bioresorbable Scaffold
title_short Investigating the Equivalent Plastic Strain in a Variable Ring Length and Strut Width Thin-Strut Bioresorbable Scaffold
title_sort investigating the equivalent plastic strain in a variable ring length and strut width thin-strut bioresorbable scaffold
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9750924/
https://www.ncbi.nlm.nih.gov/pubmed/35819580
http://dx.doi.org/10.1007/s13239-022-00625-3
work_keys_str_mv AT hoddyben investigatingtheequivalentplasticstraininavariableringlengthandstrutwidththinstrutbioresorbablescaffold
AT ahmednaveed investigatingtheequivalentplasticstraininavariableringlengthandstrutwidththinstrutbioresorbablescaffold
AT allameekadem investigatingtheequivalentplasticstraininavariableringlengthandstrutwidththinstrutbioresorbablescaffold
AT bullettnial investigatingtheequivalentplasticstraininavariableringlengthandstrutwidththinstrutbioresorbablescaffold
AT curzennick investigatingtheequivalentplasticstraininavariableringlengthandstrutwidththinstrutbioresorbablescaffold
AT bressloffneilw investigatingtheequivalentplasticstraininavariableringlengthandstrutwidththinstrutbioresorbablescaffold