Cargando…
Εffect of cleansers on the composition and mechanical properties of orthodontic aligners in vitro
BACKGROUND: The aim of the study was to investigate the effect of three aligner cleaners on the composition and mechanical properties of two types of orthodontic aligners. MATERIALS AND METHODS: The cleaners tested were two alkaline peroxide solutions (Retainer Brite—RB; Retainer Cleaner—RC) and one...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9751251/ https://www.ncbi.nlm.nih.gov/pubmed/36517652 http://dx.doi.org/10.1186/s40510-022-00449-w |
Sumario: | BACKGROUND: The aim of the study was to investigate the effect of three aligner cleaners on the composition and mechanical properties of two types of orthodontic aligners. MATERIALS AND METHODS: The cleaners tested were two alkaline peroxide solutions (Retainer Brite—RB; Retainer Cleaner—RC) and one peroxide-free (Steraligner—ST) and the aligners Clear Aligner (C, polyester) and Invisalign (I, polyester–urethane). The aligners were immersed in the cleaner solutions as instructed every day (15 min for RB, RC; 5 min for ST) for a two-week period. The acidity of the solutions was tested with a pH meter. The changes in the chemical composition of the aligners were studied by attenuated total-reflection Fourier transform infrared spectrometry (ATR-FTIR), while Instrumented Indentation Testing (IIT) was used for assessment of changes in Martens Hardness (HM), modulus (E(IT)), elastic index (n(IT)) and relaxation (R(IT)). RESULTS: RB and RC were weakly acidic (pH = 6.3), whereas ST was mildly acidic (pH = 4.8). The ATR-FTIR analysis demonstrated evidence of acidic hydrolysis of C in ST and I in RB. The IIT-derived properties of I were not affected by the cleaners. However, for C a significant change was found in HM (all cleaners), n(IT) (all cleaners) and R(IT) (RB, ST). Although the chemical changes support a hydrolytic material deterioration, the results of mechanical properties may interfere with the material residual stresses during fabrication. CONCLUSIONS: Caution should be exerted in the selection of aligner cleaners. The mild acidic cleanser was more aggressive to the polyester, whereas an alkaline peroxide to the polyester–urethane aligner. |
---|