Cargando…
Cause analysis of the liner dissociation of a customized reverse shoulder prosthesis based on finite element analysis
Background: Dissociation of the polyethylene liner after reverse shoulder arthroplasty could cause shoulder dislocation that could not achieve closed reduction. The cause of liner dissociation is currently unclear. Method: Non-homogeneous model of the bone was constructed and dynamic finite element...
Autores principales: | Wan, Qian, Zhang, Aobo, Bai, Haotian, Liu, Yang, Chen, Hao, Zhao, Xin, Wang, Xiaonan, Han, Qing, Wang, Jincheng |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Frontiers Media S.A.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9755191/ https://www.ncbi.nlm.nih.gov/pubmed/36532579 http://dx.doi.org/10.3389/fbioe.2022.1038116 |
Ejemplares similares
-
Medial-lateral translational malalignment of the prosthesis on tibial stress distribution in total knee arthroplasty: A finite element analysis
por: Zheng, Zhiqian, et al.
Publicado: (2023) -
Improving the Stability of a Hemipelvic Prosthesis Based on Bone Mineral Density Screw Channel and Prosthesis Optimization Design
por: Zhou, Rongqi, et al.
Publicado: (2022) -
Biomechanical comparison between unilateral and bilateral percutaneous vertebroplasty for osteoporotic vertebral compression fractures: A finite element analysis
por: Dai, Haowen, et al.
Publicado: (2022) -
Biomechanical effects of individualized artificial titanium alloy lamina implantation after laminectomy: A finite element analysis
por: Gong, Xuqiang, et al.
Publicado: (2022) -
Total Talar Replacement With Custom-Made Vitallium Prosthesis for Talar Avascular Necrosis
por: Luo, Wenbin, et al.
Publicado: (2022)