High performance of the automated ADVIA Centaur Systems SARS-CoV-2 Antigen Assay in nasopharyngeal samples with high viral load
ADVIA Centaur SARS-CoV-2 Antigen (COV2Ag) Assay (Siemens Healthineers) was evaluated for SARS-CoV-2 detection. A total of 141 nasopharyngeal samples were analyzed by this technique and results were compared with those obtained by quantitative reverse-transcription polymerase chain reaction (RT-PCR)....
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9758452/ https://www.ncbi.nlm.nih.gov/pubmed/36527576 http://dx.doi.org/10.1007/s10123-022-00311-3 |
Sumario: | ADVIA Centaur SARS-CoV-2 Antigen (COV2Ag) Assay (Siemens Healthineers) was evaluated for SARS-CoV-2 detection. A total of 141 nasopharyngeal samples were analyzed by this technique and results were compared with those obtained by quantitative reverse-transcription polymerase chain reaction (RT-PCR). The overall sensitivity and specificity of the test were 68.70% and 70%, respectively. Regarding cycle threshold (Ct) values, the COV2Ag test showed a sensitivity of 93.75% and 100% for nasopharyngeal samples with Ct < 25 and < 20, respectively. ADVIA Centaur COV2Ag Assay is a useful, automated, and rapid technique for early SARS-CoV-2 diagnosis and isolation of the infected individuals, avoiding its transmission. |
---|