Reliability of Marked Scales on Intravenous Fluid Plastic Bags
Precise fluid administration is important to prevent hypo- or hypervolemia. However, the accuracy of scales marked on intravenous (IV) fluid plastic bags had remained unknown. Ten 1 L sized IV crystalloids were prepared from each of three manufacturers (H, J, and D). At each scale, the actual volume...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Academy of Medical Sciences
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9763708/ https://www.ncbi.nlm.nih.gov/pubmed/36536546 http://dx.doi.org/10.3346/jkms.2022.37.e345 |
_version_ | 1784853118697078784 |
---|---|
author | Ko, Eunji Kim, Yun Hee Song, Yeon Jae Choe, Kwanyoung Heo, Junghyun Moon, Dae Eun Lim, Choon Hak |
author_facet | Ko, Eunji Kim, Yun Hee Song, Yeon Jae Choe, Kwanyoung Heo, Junghyun Moon, Dae Eun Lim, Choon Hak |
author_sort | Ko, Eunji |
collection | PubMed |
description | Precise fluid administration is important to prevent hypo- or hypervolemia. However, the accuracy of scales marked on intravenous (IV) fluid plastic bags had remained unknown. Ten 1 L sized IV crystalloids were prepared from each of three manufacturers (H, J, and D). At each scale, the actual volume of the IV fluid was measured. Differences with the measured volumes for each scale were investigated between the three manufacturers. All initial total volume was greater than 1 L. Except for the full-filled level, H overfilled, whereas J and D filled less. For J and D, the maximal differences between the scale and the measured volume were about 200 mL. Fluid volumes of each scale were significantly different among the three manufacturers (P < 0.001). It is inaccurate to measure the amount of fluid depending on the IV bag scales. Clinicians must use electronic infusion pumps for accurate fluid administration. |
format | Online Article Text |
id | pubmed-9763708 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | The Korean Academy of Medical Sciences |
record_format | MEDLINE/PubMed |
spelling | pubmed-97637082023-01-03 Reliability of Marked Scales on Intravenous Fluid Plastic Bags Ko, Eunji Kim, Yun Hee Song, Yeon Jae Choe, Kwanyoung Heo, Junghyun Moon, Dae Eun Lim, Choon Hak J Korean Med Sci Brief Communication Precise fluid administration is important to prevent hypo- or hypervolemia. However, the accuracy of scales marked on intravenous (IV) fluid plastic bags had remained unknown. Ten 1 L sized IV crystalloids were prepared from each of three manufacturers (H, J, and D). At each scale, the actual volume of the IV fluid was measured. Differences with the measured volumes for each scale were investigated between the three manufacturers. All initial total volume was greater than 1 L. Except for the full-filled level, H overfilled, whereas J and D filled less. For J and D, the maximal differences between the scale and the measured volume were about 200 mL. Fluid volumes of each scale were significantly different among the three manufacturers (P < 0.001). It is inaccurate to measure the amount of fluid depending on the IV bag scales. Clinicians must use electronic infusion pumps for accurate fluid administration. The Korean Academy of Medical Sciences 2022-11-24 /pmc/articles/PMC9763708/ /pubmed/36536546 http://dx.doi.org/10.3346/jkms.2022.37.e345 Text en © 2022 The Korean Academy of Medical Sciences. https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Brief Communication Ko, Eunji Kim, Yun Hee Song, Yeon Jae Choe, Kwanyoung Heo, Junghyun Moon, Dae Eun Lim, Choon Hak Reliability of Marked Scales on Intravenous Fluid Plastic Bags |
title | Reliability of Marked Scales on Intravenous Fluid Plastic Bags |
title_full | Reliability of Marked Scales on Intravenous Fluid Plastic Bags |
title_fullStr | Reliability of Marked Scales on Intravenous Fluid Plastic Bags |
title_full_unstemmed | Reliability of Marked Scales on Intravenous Fluid Plastic Bags |
title_short | Reliability of Marked Scales on Intravenous Fluid Plastic Bags |
title_sort | reliability of marked scales on intravenous fluid plastic bags |
topic | Brief Communication |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9763708/ https://www.ncbi.nlm.nih.gov/pubmed/36536546 http://dx.doi.org/10.3346/jkms.2022.37.e345 |
work_keys_str_mv | AT koeunji reliabilityofmarkedscalesonintravenousfluidplasticbags AT kimyunhee reliabilityofmarkedscalesonintravenousfluidplasticbags AT songyeonjae reliabilityofmarkedscalesonintravenousfluidplasticbags AT choekwanyoung reliabilityofmarkedscalesonintravenousfluidplasticbags AT heojunghyun reliabilityofmarkedscalesonintravenousfluidplasticbags AT moondaeeun reliabilityofmarkedscalesonintravenousfluidplasticbags AT limchoonhak reliabilityofmarkedscalesonintravenousfluidplasticbags |