Cargando…

Intraperitoneal Loss of Pelvic Array Pin During Navigated Total Hip Arthroplasty

Pelvic array pin placement during navigated total hip arthroplasty has been known to cause complications; however, most of them are minor. We report a 78-year-old female who underwent a routine computer-navigated total hip replacement which was subsequently complicated by complete loss of the pelvic...

Descripción completa

Detalles Bibliográficos
Autores principales: Kan, Heng, Nusem, Iulian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9764167/
https://www.ncbi.nlm.nih.gov/pubmed/36561361
http://dx.doi.org/10.1016/j.artd.2022.11.003
_version_ 1784853218292924416
author Kan, Heng
Nusem, Iulian
author_facet Kan, Heng
Nusem, Iulian
author_sort Kan, Heng
collection PubMed
description Pelvic array pin placement during navigated total hip arthroplasty has been known to cause complications; however, most of them are minor. We report a 78-year-old female who underwent a routine computer-navigated total hip replacement which was subsequently complicated by complete loss of the pelvic array pin requiring retrieval via laparotomy. No structures were injured despite the surrounding urological, vascular, and visceral anatomy. The patient recovered and subsequently underwent a total hip arthroplasty without complication 6 weeks later. This case illustrates the dangers of inserting a pelvic array pin within patients with poor bone quality and highlights necessary considerations of pin design and surgical techniques to minimize this complication.
format Online
Article
Text
id pubmed-9764167
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-97641672022-12-21 Intraperitoneal Loss of Pelvic Array Pin During Navigated Total Hip Arthroplasty Kan, Heng Nusem, Iulian Arthroplast Today Case Report Pelvic array pin placement during navigated total hip arthroplasty has been known to cause complications; however, most of them are minor. We report a 78-year-old female who underwent a routine computer-navigated total hip replacement which was subsequently complicated by complete loss of the pelvic array pin requiring retrieval via laparotomy. No structures were injured despite the surrounding urological, vascular, and visceral anatomy. The patient recovered and subsequently underwent a total hip arthroplasty without complication 6 weeks later. This case illustrates the dangers of inserting a pelvic array pin within patients with poor bone quality and highlights necessary considerations of pin design and surgical techniques to minimize this complication. Elsevier 2022-12-12 /pmc/articles/PMC9764167/ /pubmed/36561361 http://dx.doi.org/10.1016/j.artd.2022.11.003 Text en © 2022 Published by Elsevier Inc. on behalf of The American Association of Hip and Knee Surgeons. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Case Report
Kan, Heng
Nusem, Iulian
Intraperitoneal Loss of Pelvic Array Pin During Navigated Total Hip Arthroplasty
title Intraperitoneal Loss of Pelvic Array Pin During Navigated Total Hip Arthroplasty
title_full Intraperitoneal Loss of Pelvic Array Pin During Navigated Total Hip Arthroplasty
title_fullStr Intraperitoneal Loss of Pelvic Array Pin During Navigated Total Hip Arthroplasty
title_full_unstemmed Intraperitoneal Loss of Pelvic Array Pin During Navigated Total Hip Arthroplasty
title_short Intraperitoneal Loss of Pelvic Array Pin During Navigated Total Hip Arthroplasty
title_sort intraperitoneal loss of pelvic array pin during navigated total hip arthroplasty
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9764167/
https://www.ncbi.nlm.nih.gov/pubmed/36561361
http://dx.doi.org/10.1016/j.artd.2022.11.003
work_keys_str_mv AT kanheng intraperitoneallossofpelvicarraypinduringnavigatedtotalhiparthroplasty
AT nusemiulian intraperitoneallossofpelvicarraypinduringnavigatedtotalhiparthroplasty