Investigating the Origin of Mycobacterium chimaera Contamination in Heater-Cooler Units: Integrated Analysis with Fourier Transform Infrared Spectroscopy and Whole-Genome Sequencing

Mycobacterium chimaera is ubiquitously spread in the environment, including factory and hospital water systems. Invasive cases of M. chimaera infection have been associated with aerosols produced by the use of heater-cooler units (HCU) during cardiac surgery. The aim of this study was to evaluate fo...

Descripción completa

Detalles Bibliográficos
Autores principales: Bisognin, F., Messina, F., Butera, O., Nisii, C., Mazzarelli, A., Cristino, S., Pascale, M. R., Lombardi, G., Cannas, A., Dal Monte, P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Society for Microbiology 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9769643/
https://www.ncbi.nlm.nih.gov/pubmed/36222693
http://dx.doi.org/10.1128/spectrum.02893-22
_version_ 1784854416190341120
author Bisognin, F.
Messina, F.
Butera, O.
Nisii, C.
Mazzarelli, A.
Cristino, S.
Pascale, M. R.
Lombardi, G.
Cannas, A.
Dal Monte, P.
author_facet Bisognin, F.
Messina, F.
Butera, O.
Nisii, C.
Mazzarelli, A.
Cristino, S.
Pascale, M. R.
Lombardi, G.
Cannas, A.
Dal Monte, P.
author_sort Bisognin, F.
collection PubMed
description Mycobacterium chimaera is ubiquitously spread in the environment, including factory and hospital water systems. Invasive cases of M. chimaera infection have been associated with aerosols produced by the use of heater-cooler units (HCU) during cardiac surgery. The aim of this study was to evaluate for the first time the performance of IR-Biotyper system on a large number of M. chimaera isolates collected from longitudinal environmental HCUs samples and water sources from hospitals located in three Italian provinces. In addition, IR-Biotyper results were compared with whole-genome sequencing (WGS) analysis, the reference method for molecular epidemiology, to investigate the origin of M. chimaera contamination of HCUs. From November 2018 to May 2021, 417 water samples from 52 HCUs (Stockert 3T, n = 41 and HCU40, n = 11) and 23 hospital taps (used to fill the HCU tanks) were concentrated, decontaminated, and cultured for M. chimaera. Positive cultures (n = 53) were purified by agar plate subcultures and analyzed by IR-Biotyper platform and Ion Torrent sequencing system. IR-Biotyper spectra results were analyzed using a statistical approach of dimensionality reduction by linear discriminant analysis (LDA), generating three separate clusters of M. chimaera, ascribable to each hospital. Furthermore, the only M. chimaera-positive sample from tap water clustered with the isolates from the HCUs of the same hospital, confirming that the plumbing system could represent the source of HCU contamination and, potentially, of patient infection. According to the genome-based phylogenies and following the classification proposed by van Ingen and collaborators in 2017, three distinct M. chimaera groups appear to have contaminated the HCU water systems: subgroups 1.1, 2.1, and branch 2. Most of the strains isolated from HCUs at the same hospital share a highly similar genetic profile. The nonrandom distribution obtained with WGS and IR-Biotyper leads to the hypothesis that M. chimaera subtypes circulating in the local plumbing colonize HCUs through the absolute filter, in addition with the current hypothesis that contamination occurs at the HCU production site. This opens the possibility that other medical equipment, such as endoscope reprocessing device or hemodialysis systems, could be contaminated by M. chimaera. IMPORTANCE Our manuscript focuses on interventions to reduce waterborne disease transmission, improve sanitation, and control infection. Sanitary water can be contaminated by nontuberculous Mycobacteria, including M. chimaera, a causative agent of invasive infections in immunocompromised patients. We found highly similar genetic and phenotypic profiles of M. chimaera isolated from heater-cooler units (HCU) used during surgery to thermo-regulate patients' body temperature, and from the same hospital tap water. These results lead to the hypothesis that M. chimaera subtypes circulating in the local plumbing colonize HCUs through the absolute filter, adding to the current hypothesis that contamination occurs at the HCU production site. In addition, this opens the possibility that other medical equipment using sanitized water, such as endoscope reprocessing devices or hemodialysis systems, could be contaminated by nontuberculous Mycobacteria, suggesting the need for environmental surveillance and associated control measures.
format Online
Article
Text
id pubmed-9769643
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher American Society for Microbiology
record_format MEDLINE/PubMed
spelling pubmed-97696432022-12-22 Investigating the Origin of Mycobacterium chimaera Contamination in Heater-Cooler Units: Integrated Analysis with Fourier Transform Infrared Spectroscopy and Whole-Genome Sequencing Bisognin, F. Messina, F. Butera, O. Nisii, C. Mazzarelli, A. Cristino, S. Pascale, M. R. Lombardi, G. Cannas, A. Dal Monte, P. Microbiol Spectr Research Article Mycobacterium chimaera is ubiquitously spread in the environment, including factory and hospital water systems. Invasive cases of M. chimaera infection have been associated with aerosols produced by the use of heater-cooler units (HCU) during cardiac surgery. The aim of this study was to evaluate for the first time the performance of IR-Biotyper system on a large number of M. chimaera isolates collected from longitudinal environmental HCUs samples and water sources from hospitals located in three Italian provinces. In addition, IR-Biotyper results were compared with whole-genome sequencing (WGS) analysis, the reference method for molecular epidemiology, to investigate the origin of M. chimaera contamination of HCUs. From November 2018 to May 2021, 417 water samples from 52 HCUs (Stockert 3T, n = 41 and HCU40, n = 11) and 23 hospital taps (used to fill the HCU tanks) were concentrated, decontaminated, and cultured for M. chimaera. Positive cultures (n = 53) were purified by agar plate subcultures and analyzed by IR-Biotyper platform and Ion Torrent sequencing system. IR-Biotyper spectra results were analyzed using a statistical approach of dimensionality reduction by linear discriminant analysis (LDA), generating three separate clusters of M. chimaera, ascribable to each hospital. Furthermore, the only M. chimaera-positive sample from tap water clustered with the isolates from the HCUs of the same hospital, confirming that the plumbing system could represent the source of HCU contamination and, potentially, of patient infection. According to the genome-based phylogenies and following the classification proposed by van Ingen and collaborators in 2017, three distinct M. chimaera groups appear to have contaminated the HCU water systems: subgroups 1.1, 2.1, and branch 2. Most of the strains isolated from HCUs at the same hospital share a highly similar genetic profile. The nonrandom distribution obtained with WGS and IR-Biotyper leads to the hypothesis that M. chimaera subtypes circulating in the local plumbing colonize HCUs through the absolute filter, in addition with the current hypothesis that contamination occurs at the HCU production site. This opens the possibility that other medical equipment, such as endoscope reprocessing device or hemodialysis systems, could be contaminated by M. chimaera. IMPORTANCE Our manuscript focuses on interventions to reduce waterborne disease transmission, improve sanitation, and control infection. Sanitary water can be contaminated by nontuberculous Mycobacteria, including M. chimaera, a causative agent of invasive infections in immunocompromised patients. We found highly similar genetic and phenotypic profiles of M. chimaera isolated from heater-cooler units (HCU) used during surgery to thermo-regulate patients' body temperature, and from the same hospital tap water. These results lead to the hypothesis that M. chimaera subtypes circulating in the local plumbing colonize HCUs through the absolute filter, adding to the current hypothesis that contamination occurs at the HCU production site. In addition, this opens the possibility that other medical equipment using sanitized water, such as endoscope reprocessing devices or hemodialysis systems, could be contaminated by nontuberculous Mycobacteria, suggesting the need for environmental surveillance and associated control measures. American Society for Microbiology 2022-10-12 /pmc/articles/PMC9769643/ /pubmed/36222693 http://dx.doi.org/10.1128/spectrum.02893-22 Text en Copyright © 2022 Bisognin et al. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 International license (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Research Article
Bisognin, F.
Messina, F.
Butera, O.
Nisii, C.
Mazzarelli, A.
Cristino, S.
Pascale, M. R.
Lombardi, G.
Cannas, A.
Dal Monte, P.
Investigating the Origin of Mycobacterium chimaera Contamination in Heater-Cooler Units: Integrated Analysis with Fourier Transform Infrared Spectroscopy and Whole-Genome Sequencing
title Investigating the Origin of Mycobacterium chimaera Contamination in Heater-Cooler Units: Integrated Analysis with Fourier Transform Infrared Spectroscopy and Whole-Genome Sequencing
title_full Investigating the Origin of Mycobacterium chimaera Contamination in Heater-Cooler Units: Integrated Analysis with Fourier Transform Infrared Spectroscopy and Whole-Genome Sequencing
title_fullStr Investigating the Origin of Mycobacterium chimaera Contamination in Heater-Cooler Units: Integrated Analysis with Fourier Transform Infrared Spectroscopy and Whole-Genome Sequencing
title_full_unstemmed Investigating the Origin of Mycobacterium chimaera Contamination in Heater-Cooler Units: Integrated Analysis with Fourier Transform Infrared Spectroscopy and Whole-Genome Sequencing
title_short Investigating the Origin of Mycobacterium chimaera Contamination in Heater-Cooler Units: Integrated Analysis with Fourier Transform Infrared Spectroscopy and Whole-Genome Sequencing
title_sort investigating the origin of mycobacterium chimaera contamination in heater-cooler units: integrated analysis with fourier transform infrared spectroscopy and whole-genome sequencing
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9769643/
https://www.ncbi.nlm.nih.gov/pubmed/36222693
http://dx.doi.org/10.1128/spectrum.02893-22
work_keys_str_mv AT bisogninf investigatingtheoriginofmycobacteriumchimaeracontaminationinheatercoolerunitsintegratedanalysiswithfouriertransforminfraredspectroscopyandwholegenomesequencing
AT messinaf investigatingtheoriginofmycobacteriumchimaeracontaminationinheatercoolerunitsintegratedanalysiswithfouriertransforminfraredspectroscopyandwholegenomesequencing
AT buterao investigatingtheoriginofmycobacteriumchimaeracontaminationinheatercoolerunitsintegratedanalysiswithfouriertransforminfraredspectroscopyandwholegenomesequencing
AT nisiic investigatingtheoriginofmycobacteriumchimaeracontaminationinheatercoolerunitsintegratedanalysiswithfouriertransforminfraredspectroscopyandwholegenomesequencing
AT mazzarellia investigatingtheoriginofmycobacteriumchimaeracontaminationinheatercoolerunitsintegratedanalysiswithfouriertransforminfraredspectroscopyandwholegenomesequencing
AT cristinos investigatingtheoriginofmycobacteriumchimaeracontaminationinheatercoolerunitsintegratedanalysiswithfouriertransforminfraredspectroscopyandwholegenomesequencing
AT pascalemr investigatingtheoriginofmycobacteriumchimaeracontaminationinheatercoolerunitsintegratedanalysiswithfouriertransforminfraredspectroscopyandwholegenomesequencing
AT lombardig investigatingtheoriginofmycobacteriumchimaeracontaminationinheatercoolerunitsintegratedanalysiswithfouriertransforminfraredspectroscopyandwholegenomesequencing
AT cannasa investigatingtheoriginofmycobacteriumchimaeracontaminationinheatercoolerunitsintegratedanalysiswithfouriertransforminfraredspectroscopyandwholegenomesequencing
AT dalmontep investigatingtheoriginofmycobacteriumchimaeracontaminationinheatercoolerunitsintegratedanalysiswithfouriertransforminfraredspectroscopyandwholegenomesequencing