Optimisation of Volume Flow Rates when Using Endovascular Shunting Techniques: An Experimental Study in Different Bench Flow Circuits

OBJECTIVE: Acute tissue ischaemia may arise due to arterial emergencies or during more complex vascular procedures and may be mitigated by temporary shunting techniques.
Endovascular shunting (ES) techniques enable percutaneous access and shunting from the donor artery without the need to completely...

Descripción completa

Detalles Bibliográficos
Autores principales: Millinger, Johan, Bengtsson, Daniel, Langenskiöld, Marcus, Nygren, Andreas, Österberg, Klas, Nordanstig, Joakim
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9772544/
https://www.ncbi.nlm.nih.gov/pubmed/36569522
http://dx.doi.org/10.1016/j.ejvsvf.2022.11.002
_version_ 1784855001406898176
author Millinger, Johan
Bengtsson, Daniel
Langenskiöld, Marcus
Nygren, Andreas
Österberg, Klas
Nordanstig, Joakim
author_facet Millinger, Johan
Bengtsson, Daniel
Langenskiöld, Marcus
Nygren, Andreas
Österberg, Klas
Nordanstig, Joakim
author_sort Millinger, Johan
collection PubMed
description OBJECTIVE: Acute tissue ischaemia may arise due to arterial emergencies or during more complex vascular procedures and may be mitigated by temporary shunting techniques.
Endovascular shunting (ES) techniques enable percutaneous access and shunting from the donor artery without the need to completely interrupt the arterial flow in the donor artery. An endoshunt system may also cover longer distances than most conventional shunts. The aim was to investigate and optimise the flow rates in different endovascular shunt systems. METHODS: Step 1: The flow capacity of different ES configurations was compared with the flow capacity of a 9 Fr Pruitt–Inahara shunt (PIS). An intravenous bag with 0.9% NaCl, pressurised to 90 mmHg, was connected simultaneously to a PIS and to one of the tested ES configurations. The two shunt systems were then opened at the same time.
The delivered fluid volumes from the shunt systems were collected and measured. The volume flow rate was subsequently calculated.
Steps 2 and 3: Within a heart lung machine circuit, pressure–flow charts were constructed for the individual ES components and for the fully connected optimised endoshunt systems. The flow rate was increased in steps of 40–50 mL/min while monitoring the driving pressure, enabling the creation and comparison of the pressure–flow charts for the individually tested components. In total, seven individual inflow and outflow potential ES components were investigated with inflow and outflow diameters ranging from 6 to 15 Fr. RESULTS: ES systems based on standard donor introducers led to substantially lower volume flow than the corresponding PIS volume flow, whereas ES systems based on dedicated 6 or 8 Fr dialysis access introducers (Prelude Short Sheet, Merit Medical) matched PIS flow rates. The introduction of 30 cm long ¼′′ perfusion tubing within the ES system did not affect volume flow for any of the tested ES configurations. CONCLUSION: Endoshunting techniques can match PIS volume flow rates over short and long distances. The achieved ES flow rate is highly dependent on the components used within the ES system.
format Online
Article
Text
id pubmed-9772544
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-97725442022-12-23 Optimisation of Volume Flow Rates when Using Endovascular Shunting Techniques: An Experimental Study in Different Bench Flow Circuits Millinger, Johan Bengtsson, Daniel Langenskiöld, Marcus Nygren, Andreas Österberg, Klas Nordanstig, Joakim EJVES Vasc Forum Original Research OBJECTIVE: Acute tissue ischaemia may arise due to arterial emergencies or during more complex vascular procedures and may be mitigated by temporary shunting techniques.
Endovascular shunting (ES) techniques enable percutaneous access and shunting from the donor artery without the need to completely interrupt the arterial flow in the donor artery. An endoshunt system may also cover longer distances than most conventional shunts. The aim was to investigate and optimise the flow rates in different endovascular shunt systems. METHODS: Step 1: The flow capacity of different ES configurations was compared with the flow capacity of a 9 Fr Pruitt–Inahara shunt (PIS). An intravenous bag with 0.9% NaCl, pressurised to 90 mmHg, was connected simultaneously to a PIS and to one of the tested ES configurations. The two shunt systems were then opened at the same time.
The delivered fluid volumes from the shunt systems were collected and measured. The volume flow rate was subsequently calculated.
Steps 2 and 3: Within a heart lung machine circuit, pressure–flow charts were constructed for the individual ES components and for the fully connected optimised endoshunt systems. The flow rate was increased in steps of 40–50 mL/min while monitoring the driving pressure, enabling the creation and comparison of the pressure–flow charts for the individually tested components. In total, seven individual inflow and outflow potential ES components were investigated with inflow and outflow diameters ranging from 6 to 15 Fr. RESULTS: ES systems based on standard donor introducers led to substantially lower volume flow than the corresponding PIS volume flow, whereas ES systems based on dedicated 6 or 8 Fr dialysis access introducers (Prelude Short Sheet, Merit Medical) matched PIS flow rates. The introduction of 30 cm long ¼′′ perfusion tubing within the ES system did not affect volume flow for any of the tested ES configurations. CONCLUSION: Endoshunting techniques can match PIS volume flow rates over short and long distances. The achieved ES flow rate is highly dependent on the components used within the ES system. Elsevier 2022-11-19 /pmc/articles/PMC9772544/ /pubmed/36569522 http://dx.doi.org/10.1016/j.ejvsvf.2022.11.002 Text en © 2022 The Authors https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Original Research
Millinger, Johan
Bengtsson, Daniel
Langenskiöld, Marcus
Nygren, Andreas
Österberg, Klas
Nordanstig, Joakim
Optimisation of Volume Flow Rates when Using Endovascular Shunting Techniques: An Experimental Study in Different Bench Flow Circuits
title Optimisation of Volume Flow Rates when Using Endovascular Shunting Techniques: An Experimental Study in Different Bench Flow Circuits
title_full Optimisation of Volume Flow Rates when Using Endovascular Shunting Techniques: An Experimental Study in Different Bench Flow Circuits
title_fullStr Optimisation of Volume Flow Rates when Using Endovascular Shunting Techniques: An Experimental Study in Different Bench Flow Circuits
title_full_unstemmed Optimisation of Volume Flow Rates when Using Endovascular Shunting Techniques: An Experimental Study in Different Bench Flow Circuits
title_short Optimisation of Volume Flow Rates when Using Endovascular Shunting Techniques: An Experimental Study in Different Bench Flow Circuits
title_sort optimisation of volume flow rates when using endovascular shunting techniques: an experimental study in different bench flow circuits
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9772544/
https://www.ncbi.nlm.nih.gov/pubmed/36569522
http://dx.doi.org/10.1016/j.ejvsvf.2022.11.002
work_keys_str_mv AT millingerjohan optimisationofvolumeflowrateswhenusingendovascularshuntingtechniquesanexperimentalstudyindifferentbenchflowcircuits
AT bengtssondaniel optimisationofvolumeflowrateswhenusingendovascularshuntingtechniquesanexperimentalstudyindifferentbenchflowcircuits
AT langenskioldmarcus optimisationofvolumeflowrateswhenusingendovascularshuntingtechniquesanexperimentalstudyindifferentbenchflowcircuits
AT nygrenandreas optimisationofvolumeflowrateswhenusingendovascularshuntingtechniquesanexperimentalstudyindifferentbenchflowcircuits
AT osterbergklas optimisationofvolumeflowrateswhenusingendovascularshuntingtechniquesanexperimentalstudyindifferentbenchflowcircuits
AT nordanstigjoakim optimisationofvolumeflowrateswhenusingendovascularshuntingtechniquesanexperimentalstudyindifferentbenchflowcircuits