Continuous Glucose Monitoring for Patients with COVID-19 Pneumonia: Initial Experience at a Tertiary Care Center

OBJECTIVE: Patients hospitalized with COVID-19 and hyperglycemia require frequent glucose monitoring, usually performed with glucometers. Continuous glucose monitors (CGMs) are common in the outpatient setting but not yet approved for hospital use. We evaluated CGM accuracy, safety for insulin dosin...

Descripción completa

Detalles Bibliográficos
Autores principales: Dumitrascu, Adrian G., Perry, Michelle F., Boone, Rebecca J., Guzman, Maria P., Chirila, Razvan M., McNally, Allyson W., Colibaseanu, Dorin T., Meek, Shon E., Ball, Colleen T., White, Launia J., Chindris, Ana-Maria
Formato: Online Artículo Texto
Lenguaje:English
Publicado: AACE. Published by Elsevier Inc. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9773786/
https://www.ncbi.nlm.nih.gov/pubmed/36566985
http://dx.doi.org/10.1016/j.eprac.2022.12.012
Descripción
Sumario:OBJECTIVE: Patients hospitalized with COVID-19 and hyperglycemia require frequent glucose monitoring, usually performed with glucometers. Continuous glucose monitors (CGMs) are common in the outpatient setting but not yet approved for hospital use. We evaluated CGM accuracy, safety for insulin dosing, and CGM clinical reliability in 20 adult patients hospitalized with COVID-19 and hyperglycemia. METHODS: Study patients were fitted with a remotely monitored CGM. CGM values were evaluated against glucometer readings. The CGM sensor calibration was performed if necessary. CGM values were used to dose insulin, without glucometer confirmation. RESULTS: CGM accuracy against glucometer, expressed as mean absolute relative difference (MARD), was calculated using 812 paired glucometer-CGM values. The aggregate MARD was 10.4%. For time in range and grades 1 and 2 hyperglycemia, MARD was 11.4%, 9.4%, and 9.1%, respectively, with a small variation between medical floors and intensive care units. There was no MARD correlation with mean arterial blood pressure levels, oxygen saturation, daily hemoglobin levels, and glomerular filtration rates. CGM clinical reliability was high, with 99.7% of the CGM values falling within the “safe” zones of Clarke error grid. After CGM placement, the frequency of glucometer measurements decreased from 5 to 3 and then 2 per day, reducing nurse presence in patient rooms and limiting viral exposure. CONCLUSION: With twice daily, on-demand calibration, the inpatient CGM use was safe for insulin dosing, decreasing the frequency of glucometer fingersticks. For glucose levels >70 mg/dL, CGMs showed adequate accuracy, without interference from vital and laboratory values.