Solid Sampling Pyrolysis Adsorption-Desorption Thermal Conductivity Method for Rapid and Simultaneous Detection of N and S in Seafood
In this work, a rapid method for the simultaneous determination of N and S in seafood was established based on a solid sampling absorption-desorption system coupled with a thermal conductivity detector. This setup mainly includes a solid sampling system, a gas line unit for controlling high-purity o...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9786581/ https://www.ncbi.nlm.nih.gov/pubmed/36558042 http://dx.doi.org/10.3390/molecules27248909 |
Sumario: | In this work, a rapid method for the simultaneous determination of N and S in seafood was established based on a solid sampling absorption-desorption system coupled with a thermal conductivity detector. This setup mainly includes a solid sampling system, a gas line unit for controlling high-purity oxygen and helium, a combustion and reduction furnace, a purification column system for moisture, halogen, SO(2), and CO(2), and a thermal conductivity detector. After two stages of purging with 20 s of He sweeping (250 mL/min), N(2) residue in the sample-containing chamber can be reduced to <0.01% to improve the device’s analytical sensitivity and precision. Herein, 100 s of heating at 900 °C was chosen as the optimized decomposition condition. After the generated SO(2), H(2)O, and CO(2) were absorbed by the adsorption column in turn, the purification process executed the vaporization of the N-containing analyte, and then N(2) was detected by the thermal conductivity cell for the quantification of N. Subsequently, the adsorbed SO(2) was released after heating the SO(2) adsorption column and then transported to the thermal conductivity cell for the detection and quantification of S. After the instrumental optimization, the linear range was 2.0–100 mg and the correlation coefficient (R) was more than 0.999. The limit of detection (LOD) for N was 0.66 μg and the R was less than 4.0%, while the recovery rate ranged from 95.33 to 102.8%. At the same time, the LOD for S was 2.29 μg and the R was less than 6.0%, while the recovery rate ranged from 92.26 to 105.5%. The method was validated using certified reference materials (CRMs) and the measured N and S concentrations agreed with the certified values. The method indicated good accuracy and precision for the simultaneous detection of N and S in seafood samples. The total time of analysis was less than 6 min without the sample preparation process, fulfilling the fast detection of N and S in seafood. The establishment of this method filled the blank space in the area of the simultaneous and rapid determination of N and S in aquatic product solids. Thus, it provided technical support effectively to the requirements of risk assessment and detection in cases where supervision inspection was time-dependent. |
---|