Cargando…

The predictability of graft thickness for Descemet’s stripping automated endothelial keratoplasty using a mechanical microkeratome system

Descemet's stripping automated endothelial keratoplasty (DSAEK) is used for treating corneal endothelial dysfunction, and the postoperative visual acuity outcome depends on the thickness of the graft. We created a simple nomogram using factors affecting the cutting thickness during graft prepar...

Descripción completa

Detalles Bibliográficos
Autores principales: Nishisako, Sota, Yamaguchi, Takefumi, Kusano, Yuki, Higa, Kazunari, Aoki, Dai, Sasaki, Chiaki, Shimazaki, Jun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9789079/
https://www.ncbi.nlm.nih.gov/pubmed/36564442
http://dx.doi.org/10.1038/s41598-022-26679-w
Descripción
Sumario:Descemet's stripping automated endothelial keratoplasty (DSAEK) is used for treating corneal endothelial dysfunction, and the postoperative visual acuity outcome depends on the thickness of the graft. We created a simple nomogram using factors affecting the cutting thickness during graft preparation via a mechanical microkeratome system for DSAEK. This retrospective study was conducted from May 2018 through October 2022 and included donor eyes cut by automatic methods. We measured the graft thickness, cutting accuracy, and assessed ten variables with donor/cornea-related factors potentially affecting the cutting thickness. Subsequently, we created a simple nomogram. We analyzed 81 donor tissues, and the donor median age was 76 years. The mean central graft thickness was 122.2 μm, with 62% of the grafts that could be cut within the target central graft thickness range. Comparatively, donor corneas from those with cardiac diseases were cut deeper (P = 0.007). The developed nomogram provided a 83% probability of estimating the post-cutting graft thickness within 25 µm. Our nomogram, which considers cause of death, enables reproducible production of graft of a desired thickness. A detailed analysis of donor tissues, including the cause of donor death and the characteristics from pressurization to cutting, will enable more precise DSAEK graft preparation.