Time course of oversensing and impedance changes in developing implantable cardioverter-defibrillator lead fracture
BACKGROUND: Pace–sense conductors comprise a pacing coil to the tip electrode and cable to the ring-electrode. Implantable cardioverter-defibrillator (ICD) lead-monitoring diagnostics include pacing impedance (direct current resistance [DCR]) and measures of oversensing. How they change as fractures...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9795312/ https://www.ncbi.nlm.nih.gov/pubmed/36589919 http://dx.doi.org/10.1016/j.hroo.2022.09.014 |
_version_ | 1784860232478883840 |
---|---|
author | Himes, Adam Hoium, Scott Swerdlow, Charles D. |
author_facet | Himes, Adam Hoium, Scott Swerdlow, Charles D. |
author_sort | Himes, Adam |
collection | PubMed |
description | BACKGROUND: Pace–sense conductors comprise a pacing coil to the tip electrode and cable to the ring-electrode. Implantable cardioverter-defibrillator (ICD) lead-monitoring diagnostics include pacing impedance (direct current resistance [DCR]) and measures of oversensing. How they change as fractures progress is unknown. OBJECTIVES: To characterize the relationship between oversensing, impedance, and structural changes in ICD leads developing pace–sense conductor fractures. METHODS: We performed bending tests on 39 leads connected to ICD generators in an electrolyte bath with simulated electrograms. DCR was recorded every 3 minutes; electrograms were telemetered continuously. Twenty-two leads were tested to develop partial or complete fracture criteria confirmed by imaging, using DCR or DCR variability measured by standard deviation (σ(DCR)). Results are reported for 17 other test leads. RESULTS: Initial oversensing occurred with partial pacing coil fracture vs complete ring cable fracture and correlated with bending-induced DCR peaks. These peaks were too small to be detected by clinical impedance measurements and were characterized by small increases in σ(DCR) (≥0.5 Ω). Impedance threshold alerts occurred at complete pacing coil fracture but only later for ring cable fractures. The oversensing alert triggered before device-detected ventricular fibrillation more frequently than impedance alerts (94% vs 17%; P = .00002). CONCLUSIONS: In conductor fracture, early oversensing corresponds to partial pacing coil fracture or complete ring cable fracture and correlates with transient bending-induced impedance increases, which are detected by impedance variability but too small to trigger clinical impedance alerts. This explains why clinical oversensing alerts provide more warning for device-detected ventricular fibrillation than impedance alerts and suggests how to improve impedance diagnostics based on short-term variability. |
format | Online Article Text |
id | pubmed-9795312 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-97953122022-12-29 Time course of oversensing and impedance changes in developing implantable cardioverter-defibrillator lead fracture Himes, Adam Hoium, Scott Swerdlow, Charles D. Heart Rhythm O2 Experimental BACKGROUND: Pace–sense conductors comprise a pacing coil to the tip electrode and cable to the ring-electrode. Implantable cardioverter-defibrillator (ICD) lead-monitoring diagnostics include pacing impedance (direct current resistance [DCR]) and measures of oversensing. How they change as fractures progress is unknown. OBJECTIVES: To characterize the relationship between oversensing, impedance, and structural changes in ICD leads developing pace–sense conductor fractures. METHODS: We performed bending tests on 39 leads connected to ICD generators in an electrolyte bath with simulated electrograms. DCR was recorded every 3 minutes; electrograms were telemetered continuously. Twenty-two leads were tested to develop partial or complete fracture criteria confirmed by imaging, using DCR or DCR variability measured by standard deviation (σ(DCR)). Results are reported for 17 other test leads. RESULTS: Initial oversensing occurred with partial pacing coil fracture vs complete ring cable fracture and correlated with bending-induced DCR peaks. These peaks were too small to be detected by clinical impedance measurements and were characterized by small increases in σ(DCR) (≥0.5 Ω). Impedance threshold alerts occurred at complete pacing coil fracture but only later for ring cable fractures. The oversensing alert triggered before device-detected ventricular fibrillation more frequently than impedance alerts (94% vs 17%; P = .00002). CONCLUSIONS: In conductor fracture, early oversensing corresponds to partial pacing coil fracture or complete ring cable fracture and correlates with transient bending-induced impedance increases, which are detected by impedance variability but too small to trigger clinical impedance alerts. This explains why clinical oversensing alerts provide more warning for device-detected ventricular fibrillation than impedance alerts and suggests how to improve impedance diagnostics based on short-term variability. Elsevier 2022-09-21 /pmc/articles/PMC9795312/ /pubmed/36589919 http://dx.doi.org/10.1016/j.hroo.2022.09.014 Text en © 2022 Heart Rhythm Society. Published by Elsevier Inc. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Experimental Himes, Adam Hoium, Scott Swerdlow, Charles D. Time course of oversensing and impedance changes in developing implantable cardioverter-defibrillator lead fracture |
title | Time course of oversensing and impedance changes in developing implantable cardioverter-defibrillator lead fracture |
title_full | Time course of oversensing and impedance changes in developing implantable cardioverter-defibrillator lead fracture |
title_fullStr | Time course of oversensing and impedance changes in developing implantable cardioverter-defibrillator lead fracture |
title_full_unstemmed | Time course of oversensing and impedance changes in developing implantable cardioverter-defibrillator lead fracture |
title_short | Time course of oversensing and impedance changes in developing implantable cardioverter-defibrillator lead fracture |
title_sort | time course of oversensing and impedance changes in developing implantable cardioverter-defibrillator lead fracture |
topic | Experimental |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9795312/ https://www.ncbi.nlm.nih.gov/pubmed/36589919 http://dx.doi.org/10.1016/j.hroo.2022.09.014 |
work_keys_str_mv | AT himesadam timecourseofoversensingandimpedancechangesindevelopingimplantablecardioverterdefibrillatorleadfracture AT hoiumscott timecourseofoversensingandimpedancechangesindevelopingimplantablecardioverterdefibrillatorleadfracture AT swerdlowcharlesd timecourseofoversensingandimpedancechangesindevelopingimplantablecardioverterdefibrillatorleadfracture |