Cargando…
Application of failure mode and effects analysis to validate a novel hybrid Linac QC program that integrates automated and conventional QC testing
A hybrid quality control (QC) program was developed that integrates automated and conventional Linac QC, realizing the benefits of both automated and conventional QC, increasing efficiency and maintaining independent measurement methods. Failure mode and effects analysis (FMEA) was then applied in o...
Autores principales: | Pearson, Michael, Butterworth, Victoria, Misson‐Yates, Sarah, Naeem, Marium, Gonzalez Vaz, Regina, Eaton, David, Greener, Tony |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9797170/ https://www.ncbi.nlm.nih.gov/pubmed/36453139 http://dx.doi.org/10.1002/acm2.13798 |